TC50E Immersion Circulator

Contents

  1. Overview p. 1
    Product name and general description of the TC50E unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-50...50 °C
Temperature stability at -10°C0,5 K
temperature set point / displaydigital
Temperature indicationdigital
Sensor external connectionPt100
Cooling power
at 0°C0,3 kW
at -20°C0,26 kW
at -30°C0,2 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,06 kg
CO2 equivalent0,0 t
Nominal diameter probe48 mm
Length of probe155 mm
Length flexible connection1150 mm
Overall dimensions WxDxH **260x330x415 mm
Net weight25 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current3 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S394476

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit covers a process temperature range from -50 to 50 °C, so it suits low-temperature cooling tasks but not heating above 50 °C.

Defines the application window and excludes processes requiring higher temperatures.

Source
“Operating temperature range -50...50 °C”
Performance

Cooling capacity drops from 0.3 kW at 0 °C to 0.2 kW at -30 °C, so capacity must be sized at the actual operating temperature.

Prevents undersizing when the load is specified at a warmer temperature than the real process.

Source
“at 0°C 0,3 kW at -20°C 0,26 kW at -30°C 0,2 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation, ventilation and ignition-source control.

A3 flammability classification drives room ventilation, electrical zoning and service procedures.

Source
“R-290 (A3, H220)”
Installation

The refrigerant charge is only 0.06 kg with a GWP of 0.02, so leak reporting and F-gas obligations are minimal.

Small charge simplifies compliance and reduces refrigerant-related maintenance burden.

Source
“Refrigerant quantity 0,06 kg”
Installation

The unit requires a 208-240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse, so the site electrical supply must be verified.

Determines cabling, breaker sizing and whether the available site supply is compatible.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency combination is not allowed.

Prevents installation on unstable or generator supplies that fall outside the permitted combined tolerance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop if ambient rises.

Room conditions must be checked at the installation site to avoid reduced performance or shutdown.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The probe is 48 mm in diameter and 155 mm long with a 1150 mm flexible connection, so vessel geometry and routing must accommodate these dimensions.

Ensures the immersion probe fits the vessel port and the flexible line reaches the controller.

Source
“Nominal diameter probe 48 mm Length of probe 155 mm Length flexible connection 1150 mm”
Operation

The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit, so the compressor is not cycled for control.

Explains continuous compressor operation, expected energy use and valve-related maintenance.

Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Operation

Temperature stability at -10 °C is 0.5 K, which sets the achievable process control accuracy.

Allows the engineer to confirm whether the process tolerance can be met.

Source
“Temperature stability at -10°C 0,5 K”
Compliance

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for industrial environments.

Class A equipment may not be suitable for residential or light-commercial EMC environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Avoids delivery of the wrong plug type and rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment complies with US-SNAP and applicable EU laws, with the US-SNAP end-use being industrial process refrigeration.

Confirms the refrigerant is acceptable for the intended industrial use in regulated markets.

Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”