Peter Huber Kältemaschinenbau TC50E
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC50E

Peter Huber Kältemaschinenbau TC50E — Immersion coolers.

The TC50E is an air-cooled immersion cooler with a 48 mm × 155 mm probe on a 1150 mm flexible line, covering -50 to 50 °C with 0.3 kW at 0 °C. It is the smallest single-stage unit in the series, distinguished by its R-290 charge of only 0.06 kg. It serves low-temperature cooling of baths and vessels where no pump is required.

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC50E Current —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Pump speed adjustable no
Pump type no pump
External sensor connection Pt100
Temperature range -50...50 °C
Temperature stability mfr ±0, 5 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 260 x 330 x 415 mm
Refrigeration machine mfr air-cooled, natural refrigerant
Dimensions, Weight & Mechanical Design
Weight 25 kg
Electrical, Power & Drive Specifications
Current consumption 3 A (max.)
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller MPC
Temperature display LED Display
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 06 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity mfr 0: 0, 3; -20: 0, 26; -30: 0, 2 kW (at °C)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The TC50E is a direct-immersion refrigeration unit: an air-cooled condensing unit connected by a 1150 mm flexible line to a 48 mm diameter, 155 mm long probe that is placed directly into the process bath or vessel. There is no pump and no pump-speed control, so the unit cools a bath that is mixed by other means. Within the series it is the smallest single-stage model; the TC100/TC100E/TC100E-F siblings are larger, and the CT50 Single OLÉ is a different construction. The refrigerant is R-290 (A3, H220) with a first-stage charge of only 0.06 kg.

Behavior under load

Cooling capacity is temperature-dependent and must be sized at the actual process temperature: 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C. The working range is -50 to 50 °C, so the unit is for cooling only and cannot heat above 50 °C. Temperature stability is 0.5 K at -10 °C. The refrigeration unit runs continuously and control is performed by a solenoid valve in the cooling circuit, so the compressor is not cycled for temperature control.

Control, installation and operation

Control is by an MPC controller with LED display and a Pt100 external sensor connection; alarms are optical and acoustic. The unit requires a 208-240 V, 1~ or 2~, 50/60 Hz supply with a minimum 10 A fuse and a maximum current of 3 A. Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. Ambient temperature must remain between 5 and 40 °C, and cooling capacity may drop as ambient rises.

Integration & Operating Notes

Electrical
Specify the country-specific plug at ordering

Single- and two-phase devices in the 100–240 V range are delivered with a power cable and a country-specific plug, which must be stated when ordering. Confirm the plug type and the 10 A minimum fuse rating of the site circuit before delivery to avoid rework at installation.

Datasheet TC50E Immersion Circulator, Specifications
Process Limits
Size cooling capacity at the real operating temperature

Capacity falls from 0.3 kW at 0 °C to 0.26 kW at -20 °C and 0.2 kW at -30 °C. A load specified at a warmer temperature will be undersized if the process actually runs colder, so the heat load must be evaluated at the lowest intended bath temperature.

Datasheet TC50E Immersion Circulator, Specifications
Process Limits
Do not use for heating above 50 °C

The operating range ends at 50 °C, so the unit cannot serve processes that require heating above this value. It is a cooling device for low-temperature tasks; any process step above 50 °C must be handled by separate equipment.

Datasheet TC50E Immersion Circulator, Specifications
Cooling & Media
Treat the R-290 charge as a flammable A3 refrigerant

R-290 is classified A3 (H220), so installation requires attention to ventilation and ignition-source control in the vicinity of the refrigeration circuit. The charge is only 0.06 kg, which keeps F-gas reporting minimal, but the flammability classification still governs site zoning and service procedures.

Datasheet TC50E Immersion Circulator, Specifications
Environment
Keep ambient between 5 and 40 °C to preserve capacity

The unit is air-cooled, so its heat rejection depends on room conditions. Below 5 °C or above 40 °C ambient the unit is outside its permitted environment, and as ambient rises the available cooling capacity may drop, which must be accounted for in the heat balance.

Datasheet TC50E Immersion Circulator, Specifications
Control & I/O
External Pt100 sensor connection is available for process control

The MPC controller accepts an external Pt100 sensor and displays temperature on an LED display, with optical and acoustic alarms. This allows control at the process measurement point rather than at the unit, but the sensor and its cabling are not part of the base delivery and must be provided separately.

Datasheet TC50E Immersion Circulator, Specifications
Electrical
Class A equipment is intended for industrial EMC environments

The unit is classified Class A, Group 1 to EN55011, so it is intended for industrial environments and may not be suitable for residential or light-commercial locations without additional EMC measures.

Datasheet TC50E Immersion Circulator, Specifications

Selection & Integration Q&A

When should I choose the TC50E rather than the larger TC100/TC100E siblings?

Choose the TC50E when the heat load fits its capacity curve (0.3 kW at 0 °C down to 0.2 kW at -30 °C) and the process stays within -50 to 50 °C. The TC100/TC100E siblings are larger units for higher loads; the TC50E is the smallest single-stage immersion cooler in the series.

What supply and protection does the TC50E need from the site?

A 208-240 V, 1~ or 2~, 50/60 Hz supply with a minimum 10 A fuse and maximum current of 3 A. Voltage tolerance is ±5% only with ±2% frequency, and -5% voltage with +2% frequency is not allowed. Specify the country-specific plug at ordering.

What ambient and installation limits apply to the TC50E?

Ambient must stay between 5 and 40 °C; above that, cooling capacity may drop. The probe is 48 mm diameter and 155 mm long with a 1150 mm flexible line, so the vessel port and routing must accommodate these. R-290 (A3) requires ventilation and ignition-source control.

What is not included in the base delivery and must be provided separately?

The external Pt100 sensor and its cabling for the sensor connection are not part of the base unit and must be supplied separately. The power cable and country-specific plug are included only if specified at ordering. No pump is fitted, so bath mixing must be provided by other means.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TC50E Immersion Circulator
Datasheet ·83,0 КБ ·2 pp.
View PDF