Product name and general description of the chiller unit
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 35 rows
Technical data according to DIN 12876 — Unichiller 025 OLÉ (Order-No.: 3052.0033.98)
Parameter
Value
Operating temperature range
-10...40 °C
temperature set point / display
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Interface digital
USB (Device), RS232
Interface
Temperature stability at -10°C
0,5 K
Alarm message
optic, acoustic
Safety classification
I / NFL
Cooling power
at 15°C
2,5 kW
at 0°C
2 kW
at -10°C
1,2 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,55 kg
CO2 equivalent
0,8 t
Circulation pump
Immersion pump
at 0,5 bar
21 l/min
max. delivery
29 l/min
max. delivery pressure
1 bar
Pump connection
G3/4 male
min. filling capacity
3,8 l
expansion tank
1,7 l
Overall dimensions WxDxH **
460x590x743 mm
Net weight
78 kg
sound pressure level +/- 4 dB(A)
64 dB(A)
Power supply requirement
208-230V 1~/2~ 60Hz
max. current
12 A
Fuse
16 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S420207
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Selection
The chiller's cooling capacity is rated at 20°C room temperature and will drop if ambient temperature rises, so selection must account for actual site ambient conditions.
Cooling capacity at 15/0/-10°C is only valid at 20°C ambient; higher ambient reduces available capacity and may require a larger unit.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 208-230V single or two-phase 60Hz supply with a 16A fuse and draws up to 12A, so the electrical supply must be sized accordingly.
Determines breaker/fuse rating, cable sizing and whether a country-specific plug must be specified at order.
Source
“Power supply requirement 208-230V 1~/2~ 60Hz”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug only if specified at ordering, so the plug type must be stated on the order.
Without specifying the plug at order, the unit may arrive without the correct local plug, delaying commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively but must not be combined in opposite directions, so supply quality must be checked before installation.
A supply outside these combined tolerances (e.g. -5% voltage with +2% frequency) is not allowed and can damage the compressor.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as heat transfer media without a more resistant pump option.
Prevents chemical attack and pump failure; if acetone or polyglycol is required, a more resistant plastic pump must be ordered at extra cost.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water, to avoid freezing and damage.
Guides selection of heat transfer fluid for operation down to -10°C and prevents freezing-related damage.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient, so it is not suitable for wet or dusty locations.
IP20 offers no water protection; installation in washdown or outdoor areas is not permitted.
Source
“Degree of Protection IP20”
Installation
Space requirements around the unit must be respected per the operating conditions, so clearance for air-cooled condenser airflow must be planned.
Air-cooled units need adequate clearance; insufficient space reduces cooling capacity and can cause overheating.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurized gas), so handling and disposal must follow pressurized-gas safety rules.
Informs service personnel and site safety about refrigerant hazard class and handling requirements.
Source
“R-449A (A1, H280)”
Compliance
The equipment emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial electromagnetic environments and may cause interference in residential areas.
Class A equipment is not intended for residential use; EMC planning and separation from sensitive electronics may be needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Compliance
The unit complies with US-SNAP for industrial process refrigeration and applicable EU laws, so it is acceptable for industrial process cooling applications in those markets.
Confirms regulatory acceptability for industrial process refrigeration; other end-uses may not be covered.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Operation
The OLÉ controller includes an autostart function after power failure, so the chiller will resume operation automatically when power returns.
Affects unattended operation and restart behavior; may require additional external safety interlocks depending on process risk.
Source
“Autostart function for power failure”
Selection
An optional factory-fitted Pt100 sensor connection allows display (not control) of an external process temperature, so it must be ordered with the unit if needed.
If process temperature monitoring is required, this option must be specified at order since it cannot be retrofitted in the field.
Source
“Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”