Peter Huber Kältemaschinenbau Unichiller 025 OLÉ
Peter Huber Kältemaschinenbau Unichiller 025 OLÉ — Recirculating chillers with Noise level 64 dB.
The Unichiller 025 OLÉ is an air-cooled recirculating chiller with an immersion pump and the OLÉ controller, rated for -10 to 40 °C in externally closed loops. It is the 025-capacity member of the Unichiller series, distinguished from the 015 siblings by higher cooling output and 64 dB(A) noise. It serves industrial process cooling at 208-230 V, 60 Hz.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 025 OLÉ Current | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 64 dB(A) |
| Maximum delivery flow rate | 29 l/min (also 1 bar max pressure) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | Immersion pump |
| Temperature range | -10...40 °C |
| Weight | 78 kg |
| Current consumption | 12 A |
| Fuse rating | 16 A |
| Power supply | 208-230V 1~/2~ 60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280), with an immersion pump (non-adjustable speed) and G3/4 AG male pump connections.
- OLÉ controller with OLED display, 0.1 K display resolution, USB (Device) and RS232 interfaces.
- Rated 64 dB(A); the 015 and 015-H siblings are rated 63 dB(A), so this 025 unit is the larger-capacity, slightly noisier modification in the same series.
- Intended only for externally closed applications; models with "P" in the name are the ones suited to high pressure-drop loops, not this unit.
Working range and behavior
- Temperature range -10 to 40 °C; ambient limit 5 to 40 °C. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises, so site ambient must be used for sizing.
- Maximum delivery flow 29 l/min at a maximum pressure of 1 bar.
- Electrical: 208-230 V 1~/2~ 60 Hz, 12 A current consumption, 16 A fuse, IP20, safety class I / NFL.
- The plastic pump is not resistant to acetone or polyglycols; for freeze protection water should be mixed with glysantine or ethylene glycol.
Installation, control and operation
- Must not be installed in an ATEX zone; indoor use only, illuminance at least 300 lx, up to 2000 m altitude.
- Free-standing models require 50 to 200 cm wall clearance per side for maintenance; condenser airflow clearance must be planned.
- Power-failure response is configurable (OFF/Standby, ON, Power Failure Auto); the OLÉ controller includes an autostart function after power failure.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.
- Class A, Group 1 emissions per EN55011: intended for industrial electromagnetic environments.
Integration & Operating Notes
Voltage tolerance is ±5% and frequency tolerance ±2%, but the two must not be combined in opposite directions. Verify supply quality before installation; an out-of-tolerance supply can damage the compressor. The 16 A fuse and 12 A draw define the branch circuit.
Single/two-phase devices are delivered with a power cable and country-specific plug only if specified when ordering. Without this, the unit may arrive without the correct local plug and commissioning is delayed.
Free-standing models require a wall clearance of 50 to 200 cm on either side. Air-cooled condenser airflow clearance must also be respected per the operating conditions; insufficient space reduces cooling capacity and can cause overheating.
The unit may only be used indoors at an illuminance of at least 300 lx and up to 2000 m altitude. It must not be installed or started up inside an ATEX zone because it is not explosion-protected. IP20 also rules out wet or dusty locations.
The unit is designed for externally closed applications. The plastic pump is not resistant to acetone or polyglycols, so these must not be used as heat transfer media without a more resistant pump option. For freeze protection, mix water with glysantine or ethylene glycol.
Maximum tightening torques for the G3/4 AG male pump connections must not be exceeded, and connections must be counterheld during hose installation to prevent damage and leaks.
The power-failure response can be set to OFF/Standby, ON, or Power Failure Auto via the controller, and the OLÉ controller includes an autostart function after power failure. For unattended processes, decide whether automatic restart is acceptable or external interlocks are needed.
Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles. This is critical for exothermic processes where cutting cooling would cause runaway temperatures.
Cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurized gas), so handling and disposal follow pressurized-gas safety rules.
The parking brakes must be activated at the casters (if any) and/or the leveling feet must be unscrewed/activated before the unit is put into operation, to prevent movement and crushing injuries during operation.
Selection & Integration Q&A
Pick the 025 when your heat load needs the larger capacity of this modification; the 015/015-H siblings are smaller and rated 63 dB(A) versus 64 dB(A) here. If your loop has high pressure drop, note that only models with "P" in the name are suited to that, not this unit.
It requires 208-230 V 1~/2~ 60 Hz with a 16 A fuse and draws up to 12 A. Voltage and frequency tolerances are ±5% and ±2% but must not be combined oppositely. The country-specific plug must be specified at ordering or the unit may arrive without it.
Indoors only, 5-40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, never in an ATEX zone. Free-standing models need 50-200 cm wall clearance per side for maintenance, plus condenser airflow clearance; IP20 rules out wet or dusty sites.
The power cable and country-specific plug are only included if specified at order. The optional Pt100 sensor connection for external process temperature display is factory-fitted only and must be ordered with the unit. Refrigerant circuit service requires approved contractors.