** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 160Tw (Order-No.: 3056.0006.01)
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
16 kW
Cooling power at 0°C
9,5 kW
Cooling power at -10°C
5,5 kW
Cooling power at -20°C
4 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,1 kg
CO2 equivalent
2,9 t
Circulation pump
C3
max. delivery
96 l/min
max. delivery pressure
5,6 bar
Delivery at 0,5 bar
92 l/min
Delivery at 1,0 bar
86 l/min
Delivery at 1,5 bar
81 l/min
Delivery at 2,0 bar
75 l/min
Delivery at 2,5 bar
68 l/min
Delivery at 3,0 bar
62 l/min
Delivery at 3,5 bar
54 l/min
Delivery at 4,0 bar
46 l/min
Delivery at 4,5 bar
37 l/min
Delivery at 5,0 bar
24 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 15°C
600 l/h
Consumption at water 15°C, flow 0°C
320 l/h
Consumption at water 15°C, flow -10°C
133 l/h
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires cooling water inlet temperature between 5°C and 20°C, and pipes must be insulated if water temperature can fall below the condensation point.
Defines the permissible cooling water supply conditions and insulation requirement for correct installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual capacity will differ under other conditions.
Prevents overestimating cooling capacity when site conditions differ from the rating basis.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Performance
Higher cooling water temperature reduces cooling power and may require increased cooling water flow rate.
Affects sizing of the cooling water supply and expected performance at the installation site.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility checks to avoid corrosion or damage.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Cooling water pressure must be between 1 bar minimum differential and 6 bar maximum.
Sets the required cooling water supply pressure window for safe and effective operation.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
Three-phase devices with current consumption less than 63A are supplied with cable but without plug.
Requires the installer to provide a suitable plug and connection on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable power supply quality and prohibited voltage/frequency combinations for compliance.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.
Informs EMC planning and whether additional filtering or separation may be needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has protection degree IP20 and must be installed in an environment with ambient temperature between 5°C and 40°C.
Constrains the installation location regarding dust/water ingress and ambient temperature.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The chiller is water-cooled and uses refrigerant R-449A with GWP 1397 and a charge of 2.1 kg (2.9 t CO2 equivalent).
Relevant for refrigerant regulations, leak detection requirements and environmental compliance.
Source
“Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,1 kg
CO2 equivalent
2,9 t”
Selection
The circulation pump delivers a maximum of 96 l/min and 5.6 bar, with flow dropping as pressure rises.
Used to check whether the pump can overcome the external circuit pressure drop at the required flow.