Performance, temperature, and dimensional data per DIN 12876
p. 2table, 9 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Net weight
477 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
14 A
Fuse (3 phase)
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S309173
Technical data according to DIN 12876
Parameter
Unichiller 230-H
Order-No.
3039.0018.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C
at 100°C
23 kW
at 20°C
23 kW
at 15°C
21 kW
at 0°C
13,5 kW
at -10°C
9 kW
at -20°C
5,5 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,49 kg
CO2 equivalent
2,1 t
Circulation pump:
E1
max. delivery
54 l/min
max. delivery pressure
3,5 bar
Förderleistung bei 0,17 bar
52 l/min
Delivery at 0,5 bar
48 l/min
Delivery at 1,0 bar
42 l/min
Delivery at 1,5 bar
36 l/min
Delivery at 2,0 bar
28 l/min
Delivery at 2,5 bar
21 l/min
Delivery at 3,0 bar
11 l/min
Pump connection
G1 1/4 male
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
938x1288x2003 mm
** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 400 V supply with a 16 A fuse per phase, so the site electrical infrastructure must be sized accordingly.
Determines the electrical connection, cable sizing and breaker rating needed before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.
Affects installation planning and the bill of materials for the electrical hook-up.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts the permissible installation environment and enclosure requirements.
Source
“Degree of Protection IP20”
Operation
The chiller may only be operated between 5 °C and 40 °C ambient, and cooling capacity drops if ambient rises above 20 °C.
Sets the allowable room conditions and warns that published cooling capacities are only valid at 20 °C ambient.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit is air-cooled and needs sufficient free space around it, so the installation layout must respect the manufacturer's space requirements.
Prevents derating or overheating caused by restricted air flow in a cramped installation.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Compliance
Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed together with ±2 % frequency, and certain combinations are prohibited.
Must be checked against the site power quality to avoid invalid supply conditions and warranty issues.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurised gas), and the charge is 1.49 kg with 2.1 t CO2 equivalent.
Drives refrigerant handling, labelling, leak-check and F-gas reporting obligations.
Source
“R-449A (A1, H280)”
Safety
The chiller is safety class III / FL with adjustable overtemperature protection per DIN 12876, which is relevant for protecting the external process.
Confirms the built-in safety concept and the need to set the overtemperature limit for the application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance
Cooling capacity falls steeply at low temperatures, from 23 kW at 20 °C to 5.5 kW at -20 °C, so the load must be checked at the lowest operating point.
Prevents oversizing assumptions based on the 20 °C figure when the process runs cold.
Source
“at -20°C 5,5 kW”
Performance
The circulation pump delivers 54 l/min at zero head but only 11 l/min at 3.0 bar, so the system pressure drop must be matched to the pump curve.
Ensures the external circuit resistance does not starve the required flow rate.
Source
“Delivery at 3,0 bar 11 l/min”
Installation
The pump connection is G1 1/4 male, so hoses and fittings must match this thread size.
Determines the connection hardware needed for the external circuit.
Source
“Pump connection G1 1/4 male”
Operation
The unit has a minimum filling capacity of 18 l and an expansion volume of 48 l, which sets the heat transfer fluid charge and expansion allowance.
Needed to size the fluid fill and confirm the expansion tank is adequate for the loop.
Source
“min. filling capacity 18 l”
Selection
The chiller is intended for externally closed applications, so an open external circuit is not a valid configuration.
Rules out use with open baths or unsealed external loops during selection.
Source
“For externally closed applications.”
Compliance
The equipment is classified EN55011 Class A Group 1, which is industrial EMC classification and may require additional measures in residential areas.
Affects EMC planning and suitability of the installation location.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”