Unichiller 250T-H Technical Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 250T-H chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C25 kW
Cooling power at 0°C18 kW
Cooling power at -10°C11 kW
Cooling power at -20°C6 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so selection must account for actual room conditions.

Cooling capacity is rated at 20°C ambient; higher ambient reduces available cooling and may require a larger unit.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Three-phase units below 63 A are supplied with cable but without a plug, so the installer must provide a suitable plug.

Affects electrical installation planning and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be respected simultaneously; some combinations are not allowed.

Prevents connecting the unit to a supply whose combined voltage/frequency deviation exceeds permitted limits.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit requires 400 V three-phase 50 Hz supply.

Determines the electrical supply and cabling required at the installation site.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit has IP20 protection, so it must be installed in a dry, protected environment.

IP20 is not water- or dust-tight; installation location must avoid moisture and conductive dust.

Source
“Degree of Protection IP20”
Installation

Space requirements around the unit must be respected per the manufacturer's operating conditions.

Air-cooled unit needs adequate clearance for heat rejection and service access.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurized gas).

Affects handling, labeling and safety measures for refrigerant and pressure equipment.

Source
“R-449A (A1, H280)”
Compliance

The unit is EMC Class A Group 1, which may require additional measures in residential environments.

Class A equipment is intended for industrial environments and may cause interference in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The unit is intended for externally closed applications only.

Prevents use in open-loop or vented setups that could damage the pump or system.

Source
“For externally closed applications.”
Operation

Ambient temperature must stay between 5°C and 40°C for proper operation.

Operating outside this range may affect cooling performance and reliability.

Source
“min. ambient temperature 5 °C”
Performance

The circulation pump delivers 210 l/min max at 4.7 bar max, with flow decreasing as pressure rises.

Needed to match pump curve to the external circuit's pressure drop.

Source
“max. delivery 210 l/min”
Maintenance

The expansion tank is atmospheric open and has a 39 l expansion volume.

Determines fluid fill limits and expansion allowance for the loop.

Source
“Volume of expansion 39 l”
Selection

Minimum filling capacity is 14 l, which sets the minimum system volume for proper operation.

Ensures the external loop has sufficient thermal mass and pump stability.

Source
“min. filling capacity 14 l”
Compliance

The equipment complies with US-SNAP and applicable EU laws for industrial process refrigeration.

Confirms regulatory suitability for industrial process refrigeration applications.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws.”