  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 5 rows
  4. p. 1 table, 38 rows

Technical data according to DIN 12876

ParameterValue
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S553460

Technical data according to DIN 12876 — Unichiller P007-H OLÉ (Order-No. 3012.0263.98)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at 15°C0,7 kW
Cooling power at 0°C0,55 kW
Cooling power at -10°C0,4 kW
Cooling power at -20°C0,2 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,3 kg
CO2 equivalent0,4 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight57 kg
sound pressure level +/- 4 dB(A)58 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current14 A
Fuse16 A

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is rated for a -20 to 100 °C operating range, so it suits both sub-ambient cooling and moderate heating duties.

Defines the process window the chiller can cover before you consider a different model.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity falls from 0.7 kW at 15 °C to 0.2 kW at -20 °C, so low-temperature duty needs a much larger unit or reduced heat load.

Sizing must be based on the capacity at the actual working temperature, not the headline figure.

Source
“Cooling power at 15°C 0,7 kW at 0°C 0,55 kW at -10°C 0,4 kW at -20°C 0,2 kW”
Installation

The refrigeration machine is air-cooled, so the installation site must provide adequate air flow and respect the stated space requirements.

Air-cooled condensers reject heat to the room and will derate or trip if ventilation is insufficient.

Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Installation

The electrical supply must be 208-240 V single or two-phase at 50/60 Hz with a 16 A fuse and 14 A max current.

Determines the site power provision and breaker sizing before delivery.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction, so supply quality must be checked.

A supply outside these combined limits can damage the compressor and voids compliant operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

Cooling capacity is quoted at 20 °C room temperature and drops as ambient rises, so hot plant rooms reduce available duty.

Ambient conditions at the installation site directly affect achievable cooling performance.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The unit may only be operated between 5 °C and 40 °C ambient temperature with IP20 protection.

Operating outside this range risks condensation, overheating or loss of protection.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant is R-449A with GHS classification H280 (pressurised gas) and GWP 1397, so leak handling and CO2 reporting apply.

Affects refrigerant handling procedures, room ventilation and environmental reporting obligations.

Source
“R-449A (A1, H280)”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used as heat transfer fluid.

Prevents premature pump failure and leakage when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used neat.

Protects the evaporator and circuit from freezing when operating below ambient.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single and two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.

Missing plug specification at order leads to a unit that cannot be connected on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Restricts where the unit may be installed and what EMC measures may be needed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The pump connection is G3/4 male and the minimum filling capacity is 3.8 l with a 1.7 l expansion tank.

Needed to plan hose adapters and to charge the correct volume of heat transfer fluid.

Source
“Pump connection G3/4 male min. filling capacity 3,8 l expansion tank 1,7 l”
Selection

The pump delivers 25 l/min max at 2.5 bar max, with flow falling to 6 l/min at 2.0 bar, so high-pressure-drop loops need checking.

Confirms whether the pump can overcome the pressure drop of the customer's circuit.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”