Selection
The unit is rated for a -20 to 100 °C operating range, so it suits both sub-ambient cooling and moderate heating duties.
Defines the process window the chiller can cover before you consider a different model.
Source
“Operating temperature range -20...100 °C”
Performance
Cooling capacity falls from 0.7 kW at 15 °C to 0.2 kW at -20 °C, so low-temperature duty needs a much larger unit or reduced heat load.
Sizing must be based on the capacity at the actual working temperature, not the headline figure.
Source
“Cooling power at 15°C 0,7 kW at 0°C 0,55 kW at -10°C 0,4 kW at -20°C 0,2 kW”
Installation
The refrigeration machine is air-cooled, so the installation site must provide adequate air flow and respect the stated space requirements.
Air-cooled condensers reject heat to the room and will derate or trip if ventilation is insufficient.
Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Installation
The electrical supply must be 208-240 V single or two-phase at 50/60 Hz with a 16 A fuse and 14 A max current.
Determines the site power provision and breaker sizing before delivery.
Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation
Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction, so supply quality must be checked.
A supply outside these combined limits can damage the compressor and voids compliant operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
Cooling capacity is quoted at 20 °C room temperature and drops as ambient rises, so hot plant rooms reduce available duty.
Ambient conditions at the installation site directly affect achievable cooling performance.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The unit may only be operated between 5 °C and 40 °C ambient temperature with IP20 protection.
Operating outside this range risks condensation, overheating or loss of protection.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The refrigerant is R-449A with GHS classification H280 (pressurised gas) and GWP 1397, so leak handling and CO2 reporting apply.
Affects refrigerant handling procedures, room ventilation and environmental reporting obligations.
Source
“R-449A (A1, H280)”
Safety
The plastic pump is not resistant to acetone or polyglycols, so these media must not be used as heat transfer fluid.
Prevents premature pump failure and leakage when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used neat.
Protects the evaporator and circuit from freezing when operating below ambient.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single and two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.
Missing plug specification at order leads to a unit that cannot be connected on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance
The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Restricts where the unit may be installed and what EMC measures may be needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The pump connection is G3/4 male and the minimum filling capacity is 3.8 l with a 1.7 l expansion tank.
Needed to plan hose adapters and to charge the correct volume of heat transfer fluid.
Source
“Pump connection G3/4 male min. filling capacity 3,8 l expansion tank 1,7 l”
Selection
The pump delivers 25 l/min max at 2.5 bar max, with flow falling to 6 l/min at 2.0 bar, so high-pressure-drop loops need checking.
Confirms whether the pump can overcome the pressure drop of the customer's circuit.
Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”