Unichiller P010w Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller P010w chiller unit
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Net weight57 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current6,5 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

Technical data according to DIN 12876 — Unichiller P010w (Order-No.: 3050.0018.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C1 kW
at 0°C0,8 kW
at -10°C0,5 kW
at -20°C0,15 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,2 kg
CO2 equivalent0,3 t
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm

** сноска в исходном тексте не раскрыта

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar, so the site water supply must be verified before installation.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The cooling water intake temperature must be between 5°C and 20°C, and pipes must be insulated if the water can fall below the dew point.

Out-of-range cooling water temperature reduces cooling capacity and can cause condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly as the required temperature decreases, from 1 kW at +15°C to only 0.15 kW at -20°C.

The engineer must confirm that the available cooling capacity at the target temperature meets the process load.

Source
“Cooling power at ambient temperature +20°C”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used without a special upgraded pump.

Using incompatible heat transfer fluids will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Proper antifreeze selection prevents freezing and damage in the cooling circuit at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance

Voltage and frequency tolerances must be applied simultaneously: ±5% voltage with ±2% frequency, but opposite-sign deviations are not allowed.

The site power supply must be checked against this combined tolerance to avoid motor and electronics stress.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is rated IP20, so it must be installed in a dry indoor location protected from water ingress.

IP20 offers no water protection, so the installation environment must be dry and clean.

Source
“Degree of Protection IP20”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, meaning it is non-flammable but stored under pressure.

Personnel must be aware of the pressurized refrigerant hazard during handling and service.

Source
“R-449A (A1, H280)”
Selection

The unit has a minimum filling capacity of 3.8 liters and an expansion volume of 1.7 liters, which determines the required heat transfer fluid volume.

The engineer must plan for at least this fluid volume and account for thermal expansion.

Source
“min. filling capacity 3,8 l”
Installation

The pump connections are G3/4 male and cooling water connections are G1/2 male, so matching hose couplings must be provided.

Correct thread matching avoids leaks and ensures proper connection during installation.

Source
“Pump connection G3/4 male”
Operation

The maximum ambient temperature for operation is 40°C and the minimum is 5°C, so the installation room must stay within this range.

Operating outside the ambient range can reduce performance or cause unit faults.

Source
“max. ambient temperature 40 °C”
Compliance

The equipment emits electromagnetic disturbance classified as Class A, Group 1 per EN55011, which may require separation from sensitive equipment.

Class A equipment may cause interference in residential environments and needs appropriate EMC planning.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”