Unichiller P015-H Recirculating Chiller Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller P015-H recirculating chiller unit
  2. Spec p. 2
    Performance and technical data according to DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **420x487x579 mm
Net weight62 kg
sound pressure level +/- 4 dB(A)63 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S612715

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller P015-H (Order-No.: 3051.0048.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C1,5 kW
at 0°C1 kW
at -10°C0,7 kW
at -20°C0,3 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,17 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l

Строка «Cooling power at ambient temperature +20°C» напечатана с точкой-заполнителем без значения; в таблице оставлена пустой.

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.

Defines the application window and excludes processes outside this range.

Source
“Operating temperature range -20...100 °C”
Selection

Cooling capacity drops sharply at low temperatures, from 1.5 kW at 15 °C to 0.3 kW at -20 °C, so low-temperature duty needs careful sizing.

Prevents undersizing the unit for cold setpoints.

Source
“at 15°C 1,5 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,3 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source control.

Flammable refrigerant classification drives installation safety requirements.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The unit requires a 220-240 V single-phase 50 Hz supply with a maximum current of 15.5 A and a 16 A fuse.

Determines electrical supply, cabling and protection sizing.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 15,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are coupled: ±5% voltage is allowed only with simultaneous ±2% frequency tolerance, and some combinations are not allowed.

Avoids invalid supply conditions that could damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20 and must be installed in an environment between 5 °C and 40 °C ambient temperature.

Defines the permissible installation environment and protection against ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is valid at 20 °C room temperature and may drop if ambient temperature rises.

Requires derating when installed in warm rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The pump connection is G3/4 male, so hoses and fittings must match this thread.

Ensures correct plumbing connection at installation.

Source
“Pump connection G3/4 male”
Operation

The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow falling as pressure rises.

Needed to check whether the pump can overcome the process pressure drop.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Operation

The minimum filling capacity is 3.8 l and expansion volume is 1.7 l, which sets the heat transfer fluid charge.

Determines fluid volume and expansion vessel sizing.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Compliance

The equipment is classified as EN55011 Class A, Group 1, which is industrial EMC classification.

Class A equipment may not be suitable for residential environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

Prevents delivery with the wrong plug for the installation country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Overtemperature protection is a key safety function for the process fluid and equipment.

Source
“With adjustable overtemperature protection according to DIN 12876.”