Unichiller P025-H Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller P025-H chiller unit and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data according to DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Net weight81 kg
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S597891

Technical data according to DIN 12876 — Unichiller P025-H (Order-No.: 3052.0047.01)

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C2,5 kW
at 0°C2 kW
at -10°C1,2 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,25 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **460x590x743 mm

** Overall dimensions WxDxH — см. сноску в оригинале; строка 'Cooling power at ambient temperature +20°C' напечатана без значения (значения приведены для 15°C, 0°C, -10°C).

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller uses R-290 (propane, A3 flammable) refrigerant, so the installation location must be assessed for flammable-refrigerant requirements.

A3 flammability classification affects room ventilation, ignition-source control and local safety compliance at the installation site.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Selection

Cooling capacity is only 1.2 kW at -10 °C versus 2.5 kW at +15 °C, so the load must be checked at the actual working temperature.

Sizing on the +15 °C figure would overspecify the achievable duty at low temperature and risk insufficient cooling.

Source
“at -10°C 1,2 kW”
Performance

Cooling capacity figures are valid only at 20 °C room temperature and may drop at higher ambient temperatures.

The machine room temperature must be kept within the stated ambient range or the rated capacity cannot be relied on.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 220-240 V single-phase 50 Hz supply with a 16 A fuse and draws up to 15 A.

The electrical supply, cable and protective device must be sized for 15 A on a dedicated single-phase circuit.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 15 A Fuse 16 A”
Installation

Voltage and frequency tolerances must be met simultaneously, and a -5% voltage with +2% frequency combination is explicitly not allowed.

Supply quality at the site must be verified against the combined tolerance rule before commissioning.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used as heat transfer fluid.

Fluid selection must avoid these media or a more resistant pump option must be ordered at extra cost.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

The heat transfer fluid must be specified with antifreeze to operate safely down to -10 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The pump connections are G3/4 male, and hose couplings for G3/4 male are included in the scope of delivery.

External piping and hoses must be ordered to match the G3/4 male thread.

Source
“Pump connection G3/4 male”
Performance

Pump delivery falls from 25 l/min at zero head to 6 l/min at 2.0 bar, so the system pressure drop determines the usable flow.

The external circuit pressure drop must be checked against the pump curve to confirm adequate flow.

Source
“Delivery at 2,0 bar 6 l/min”
Installation

The minimum filling capacity is 3.8 l with 1.7 l expansion volume, which sets the minimum fluid charge and expansion allowance.

The fluid volume and expansion vessel capacity must be planned for the external loop.

Source
“min. filling capacity 3,8 l”
Installation

The unit is rated IP20, so it must be installed in a dry, protected location free from water ingress.

IP20 limits the installation environment and rules out washdown or wet areas.

Source
“Degree of Protection IP20”
Installation

The permitted ambient temperature range is 5 °C to 40 °C, which constrains the room where the chiller is placed.

The installation room must stay within this range for reliable operation and rated performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

The overtemperature cut-out must be set and verified during commissioning as part of the safety concept.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment is intended for industrial environments and may require EMC measures in residential or sensitive areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”