Peter Huber Kältemaschinenbau Unichiller P025-H
Peter Huber Kältemaschinenbau Unichiller P025-H — Recirculating chillers with Noise level 64 dB.
The Unichiller P025-H is a single-phase 220–240 V/50 Hz recirculating chiller with a 'P' pump variant built for externally closed loops with high pressure drop, covering -10 to 100 °C. Its R-290 (A3) refrigerant and 1.2 kW at -10 °C capacity distinguish it from the 015 siblings, which share the same 63 dB(A) noise class.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P025-H Current | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 64 dB(A) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Weight | 81 kg |
| Current consumption | 15 A (max) |
| Fuse rating | 16 A |
| Power supply | 220-240V 1~ 50Hz V, Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P025-H is the high-pressure-drop modification of the Unichiller series: the 'P' in the model name designates a pump configuration intended for externally closed circuits with elevated pressure loss, per the operation manual.
- It is a free-standing, castor-mounted unit (leveling feet/parking brakes must be activated before start-up) with a G3/4 male pump connection; hose couplings for G3/4 male are included in delivery.
- Sibling Unichiller 015 variants are rated at 63 dB(A); this model is listed at 64 dB(A). The 015 family does not carry the 'P' high-pressure-drop designation.
- The controller is the Pilot ONE; interfaces are Ethernet, USB (host and device) and RS232. Protection rating is IP20, safety class III/FL per DIN 12876 with adjustable overtemperature protection.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 2.5 kW at +15 °C but only 1.2 kW at -10 °C, so the load must be checked at the actual working temperature, not at a nominal point.
- Rated output figures are valid at 20 °C room temperature; capacity may drop as ambient rises within the permitted 5–40 °C range.
- Pump delivery falls from 25 l/min at zero head to 6 l/min at 2.0 bar; the external circuit pressure drop therefore sets the usable flow. Pump speed is not adjustable.
- The plastic pump is not resistant to acetone or polyglycols; for operation down to -10 °C, water must be mixed with glysantine or ethylene glycol for freeze protection.
- Minimum filling capacity is 3.8 l with 1.7 l expansion volume, setting the minimum charge and expansion allowance for the external loop.
Installation, control and operation
- Electrical: 220–240 V 1~ 50 Hz, max. current 15 A, 16 A fuse. Voltage and frequency tolerances must be met simultaneously; the combination -5 % voltage with +2 % frequency is explicitly not allowed.
- The unit is Class A, Group 1 to EN55011 and must be installed indoors at ≥300 lx, up to 2000 m altitude, outside any ATEX zone (not explosion-protected).
- R-290 (A3, H220) refrigerant requires assessment of flammable-refrigerant requirements and checking refrigerant charge against room volume at the installation site.
- Wall clearance of 50–200 cm per side is required for maintenance on free-standing models; electrical lines directly below a water-cooled unit must be protected against condensate ingress.
- Power-failure response is configurable via Pilot ONE as OFF/Standby, ON, or Power Failure Auto. In Process Safety mode, maximum cooling capacity is released when the OT trigger value is reached instead of switching off all poles.
Integration & Operating Notes
Capacity falls from 2.5 kW at +15 °C to 1.2 kW at -10 °C, and the published figures assume 20 °C room temperature. A load sized on the +15 °C value will be undercooled at low setpoints or in a warmer room.
The supply must satisfy voltage and frequency tolerances simultaneously; the datasheet explicitly forbids -5 % voltage combined with +2 % frequency. Provide a dedicated single-phase circuit with a 16 A fuse, since the unit draws up to 15 A.
The pump is not resistant to acetone or polyglycols (depending on manufacturer), so these cannot serve as heat transfer fluid. For setpoints down to -10 °C, mix water with glysantine or ethylene glycol for freeze protection.
The unit uses R-290 (A3, H220) flammable refrigerant. The installation location must be evaluated for flammable-refrigerant requirements, and the charge must be checked against the room size so the refrigerant limit per m³ of ambient air is not exceeded.
Free-standing models require 50 to 200 cm wall clearance per side for maintenance. The unit is IP20 and may only be used indoors at ≥300 lx and up to 2000 m altitude; it must not be installed or started inside an ATEX zone.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. In Process Safety mode, maximum cooling capacity is released when the OT trigger value is reached instead of switching off all poles — relevant for exothermic loads.
Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Plan service access accordingly, and shut off cooling water supply even for short downtimes such as overnight.
Selection & Integration Q&A
Choose the P025-H when the external loop has high pressure drop — the 'P' designates a pump suited to such applications — and when capacity near -10 °C (1.2 kW) is needed. The 015 variants are rated 63 dB(A) and lack the high-pressure-drop designation.
An externally closed loop with G3/4 male mating connections (hose couplings included). Check loop pressure drop against the pump curve (25 l/min at 0 bar down to 6 l/min at 2.0 bar), and allow at least 3.8 l filling capacity plus 1.7 l expansion volume.
Indoors only, 5–40 °C ambient, ≥300 lx, up to 2000 m altitude, IP20 dry location, outside ATEX zones. Because of the R-290 (A3) refrigerant, check charge against room volume and keep 50–200 cm side clearance for maintenance.
Hose couplings for G3/4 male are included, but heat transfer fluid (water with glysantine or ethylene glycol for freeze protection) is not. Refrigerant-circuit work must be done by approved refrigeration contractors; shut off cooling water even overnight.