Product introduction and main features of the chiller
Specpp. 1-2
Technical data and performance parameters per DIN 12876
p. 2table, 20 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
348 l/h
Consumption at water 15°C, flow 0°C
330 l/h
Consumption at water 15°C, flow -10°C
270 l/h
Consumption at water 15°C, flow -20°C
240 l/h
min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
740x1160x1050 mm
Net weight
321 kg
sound pressure level +/- 4 dB(A)
58 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
15,5 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S432091
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P100w-H — Technical data according to DIN 12876
Parameter
Value
Order-No.
3059.0012.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C at 100°C
10 kW
Cooling power at ambient temperature +20°C at 20°C
10 kW
Cooling power at ambient temperature +20°C at 0°C
7,8 kW
Cooling power at ambient temperature +20°C at -10°C
5,3 kW
Cooling power at ambient temperature +20°C at -20°C
3,1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1 kg
CO2 equivalent
1,4 t
Circulation pump
E2
max. delivery
130 l/min
max. delivery pressure
5,7 bar
Delivery at 2,4 bar
116 l/min
Delivery at 2,5 bar
115 l/min
Delivery at 3,0 bar
108 l/min
Delivery at 3,5 bar
100 l/min
Delivery at 4,0 bar
92 l/min
Delivery at 4,5 bar
80 l/min
Delivery at 5,0 bar
59 l/min
Delivery at 5,5 bar
20 l/min
Pump connection
G1 1/4 male
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires a three-phase 400 V supply with a 3x16 A fuse and draws up to 15.5 A, so the site electrical supply must be sized accordingly.
Determines the electrical installation, cable and protection device rating before the unit can be commissioned.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is water-cooled and needs a cooling water connection of G1/2 male with a minimum differential pressure of 0.5 bar and maximum pressure of 6 bar.
The cooling water circuit must be designed to deliver adequate flow and pressure without exceeding the unit's limits.
Source
“min. cooling water differential pressure 0,5 bar”
Operation
Cooling water inlet temperature must stay between 5 °C and 20 °C; below the condensation point the cooling water pipes inside the unit must be insulated.
Operating outside this range risks condensation, reduced cooling power or equipment damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so performance must be derated for warmer water.
Selection must account for actual site cooling water temperature to avoid insufficient cooling capacity.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water quality and treatment must be compatible with these materials to prevent corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
The site power quality must be verified against these combined tolerances to avoid invalid operating conditions.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
The installer must plan for a matching plug and connection point before commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, with 1 kg charge equal to 1.4 t CO2 equivalent.
Affects refrigerant handling, leak detection and environmental reporting obligations.
Source
“R-449A (A1, H280)”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C and has IP20 protection, so it must be installed in a dry indoor environment.
Site ambient conditions and ingress protection constrain where the chiller can be located.
Source
“max. ambient temperature 40 °C”
Installation
The unit weighs 321 kg and measures 740x1160x1050 mm, and space requirements must be respected.
Floor loading, access routes and clearance for installation and maintenance must be planned.
Source
“Please respect space requirements.”
Operation
The pump delivers up to 130 l/min at 5.7 bar but only 20 l/min at 5.5 bar, so system pressure drop must be checked against the pump curve.
High-pressure-drop applications must be evaluated to ensure adequate flow is available.
Source
“max. delivery pressure 5,7 bar”
Maintenance
The minimum filling capacity is 18 l and expansion volume is 48 l, which must be considered when preparing the heat transfer circuit.
Determines the required volume of heat transfer fluid and expansion tank sizing.