Unimotive 20w Immersion Circulator

Contents

  1. Overview p. 1
    Product introduction and main features of the Unimotive 20w
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits
  3. Spec p. 2
    DIN 12876 compliant technical parameters table
  4. p. 2 table, 25 rows
  5. p. 1 table, 26 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C920 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,2 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight446 kg
Power supply requirement400V 3~ 50Hz
max. current31,5 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S548098

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...95 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Water/ Glysantin 40/60 at 20°C21 kW
Cooling power with Water/ Glysantin 40/60 at 0°C17,5 kW
at -20°C9,5 kW
at -30°C5,5 kW
at -40°C3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,9 kg
CO2 equivalent6,2 t
Circulation pump:MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so the heat transfer fluid must be selected accordingly.

Determines the permissible heat transfer fluid and prevents unsuitable fluids from being used.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

The cooling water intake temperature must be between 5°C and 20°C, which constrains the site cooling water supply.

Needed to verify the available cooling water temperature at the installation site before commissioning.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water falls below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion damage in the cooling water circuit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops and cooling water flow rate may increase as cooling water temperature rises, so performance must be derated accordingly.

Affects sizing of the cooling water supply and expected performance at the site.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or clogging.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Must be checked against the site power supply to ensure compliant operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical installation and protective device selection.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.

Needed for planning the electrical connection and ordering the correct plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Space requirements must be respected, so the installation layout must account for the specified clearances.

Ensures adequate access and ventilation at the installation site.

Source
“Please respect space requirements. See operating conditions at www.huber-online”
Operation

The unit has an adjustable overtemperature protection according to DIN 12876, which must be set for safe operation.

Safety setting that must be configured before operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The unit is intended for externally closed applications, which limits the hydraulic circuit design.

Determines whether the unit suits the intended process loop.

Source
“For externally closed applications.”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so matching fittings are required.

Ensures correct mating fittings and hoses are procured for installation.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”