Induztr
Engineering Digest Document Viewer Related Models

Unistat 425w Technical Data

Contents

  1. Overview p. 1
    Introduces the Unistat 425w unit and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 18 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C150 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,6 l
Filling capacity expansion tank3,3 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight177 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
max. current13 A
Fuse3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S508110

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 425w (Order-No.: 1050.0011.01)

ParameterValue
Operating temperature range-40...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power2 kW
Cooling power with Thermooil at 250°C2,8 kW
Cooling power with Thermooil at 200°C2,8 kW
Cooling power with Thermooil at 100°C2,8 kW
Cooling power with Thermooil at 20°C2,8 kW
Cooling power with Ethanol at 0°C2,5 kW
Cooling power with Ethanol at -20°C1,9 kW
Cooling power with Ethanol at -40°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,97 kg
CO2 equivalent2,1 t
Circulation pump:MK pump
max. delivery91 l/min
max. delivery pressure1,5 bar
Delivery at 0,25 bar85 l/min
Delivery at 0,5 bar78 l/min
Delivery at 0,75 bar71 l/min
Delivery at 1,0 bar60 l/min
Delivery at 1,25 bar40 l/min
Pump connectionM30x1,5 male

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 13 A Fuse 3x16A”
Installation

The cooling water circuit must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with G1/2 male connections.

Defines the required cooling water supply pressure and connection size for proper operation.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.

Specifies the allowable cooling water temperature range and a condensation-prevention measure.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects sizing of the cooling water supply and expected performance at the installation site.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Compliance

Voltage and frequency tolerances must be applied together: ±5% voltage with ±2% frequency, but opposite-sign deviations are not allowed.

Prevents connecting the unit to a supply whose combined voltage/frequency deviation exceeds the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The equipment has a leakage current greater than 3.5 mA, which must be considered for protective earthing and RCD selection.

Impacts electrical safety design and the choice of residual current protective devices.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Clarifies what electrical termination is included and what the installer must supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is rated IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.

Sets environmental limits and protection level for the installation location.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The pump delivers 91 l/min maximum and 1.5 bar maximum pressure, with flow falling as pressure rises.

Allows checking whether the pump curve suits the external circuit's flow and pressure drop.

Source
“max. delivery 91 l/min max. delivery pressure 1,5 bar”
Selection

The refrigerant R-452A has a GWP of 2141 and a charge of 0.97 kg, giving 2.1 t CO2 equivalent.

Relevant for refrigerant regulations, leak monitoring and environmental compliance.

Source
“Global Warming Potential (GWP) 2141 Refrigerant quantity 0,97 kg CO2 equivalent 2,1 t”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The unit has a minimum filling capacity of 3.6 l and an expansion tank capacity of 3.3 l.

Determines the minimum heat transfer fluid volume needed for safe operation.

Source
“min. filling capacity 3,6 l Filling capacity expansion tank 3,3 l”