Peter Huber Kältemaschinenbau Unistat 425w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 425w Dynamisches

Peter Huber Kältemaschinenbau Unistat 425w Dynamisches — Temperature control units with Noise level 61 dB.

The Unistat 425w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C with 2 kW heating. It is the dynamic (high-flow) modification of the 4xx series, delivering up to 91 l/min at 1.5 bar for externally closed circuits. It serves precise temperature control of reactors, distillation systems and other jacketed process equipment.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Unistat 425w Dynamisches Current 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Control output POKO ONE
Heating capacity 2 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -40...250 °C
Dimensions, Weight & Mechanical Design
Weight 177 kg
Electrical, Power & Drive Specifications
Current consumption 13 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 150 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Leakage current exceeds 3.5 mA — RCD and earthing design required

The equipment has a leakage current greater than 3.5 mA. Protective earthing and the selection of residual current protective devices must account for this; a standard 30 mA RCD may not be suitable without engineering review.

Datasheet, Technical Specifications
Mounting
Do not install in an ATEX zone; IP20 and 5–40 °C ambient apply

The unit must not be installed or started up within an ATEX zone; explosion protection requires a separate Ex px cabinet. The IP20 rating and ambient temperature range of 5–40 °C define the allowable installation location. A condensation water / heat transfer fluid drip tray must be installed below the unit; national legislation may prescribe it.

Manual, Section 1.3.3 and 1.4
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power-interruption response can be configured as Off/Standby (default) or ON/Temperature control active; choose according to whether unattended restart is acceptable.

Manual, Section 1.7.2 and 1.8.1

Selection & Integration Q&A

When should I choose the Unistat 425w Dynamisches instead of a Grande Fleur variant in the same series?

Choose the 425w Dynamisches for high-flow, low-pressure-drop externally closed circuits needing up to 91 l/min at 1.5 bar and -40 to 250 °C. Grande Fleur variants differ in configuration; confirm pump curve and cooling capacity against your circuit's pressure drop before selecting.

What power supply and protective devices must be provided for this unit?

A 400 V 3~ 50 Hz supply with 3x16 A fusing; max current 13 A. The unit comes with cable but no plug, so a suitable plug must be provided. Leakage current >3.5 mA requires RCD and earthing design review. Voltage/frequency tolerances apply together (±5% / ±2%).

What are the installation environment and duty-cycle limits from the manual?

Install indoors, IP20, ambient 5–40 °C, not in an ATEX zone. Transport and store upright only; allow ~2 h acclimation after temperature changes before power-on. Leveling feet must be activated. A drip tray is required below the unit. Cooling water intake 5–20 °C; shut off cooling water during overnight downtimes.

What maintenance items and accessories are not included in base delivery?

Base delivery excludes the power plug, cooling water hoses, heat transfer fluid, and drip tray. Periodic tasks include cleaning the hat-type strainer, inspecting hoses and mechanical seal, and checking the emergency stop switch. Refrigerant work requires approved contractors. A gas detection sensor and processing unit are optional and installed by the operating company.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 425w Technical Data
Datasheet ·77,5 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF