Introduces the Unistat 825w unit and its main features
Specpp. 1-2
Lists performance, temperature, and physical specifications per DIN 12876
p. 2table, 23 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Circulation pump: max. delivery
40 l/min
max. delivery pressure
0,9 bar
Delivery at 0,1 bar
30 l/min
Delivery at 0,3 bar
25 l/min
Delivery at 0,5 bar
18 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
231 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3 l
Filling capacity expansion tank
3,6 l
Overall dimensions WxDxH **
460x604x1465 mm
Net weight
223 kg
Power supply requirement
400V 3~ 50Hz
max. current
15 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S466954
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat 825w
Parameter
Value
Operating temperature range
-85...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power with Thermooil at 250°C
2,3 kW
Cooling power with Thermooil at 200°C
2,3 kW
Cooling power with Thermooil at 100°C
2,3 kW
Cooling power with Ethanol at 0°C
2,4 kW
Cooling power with Ethanol at -20°C
2,4 kW
Cooling power with Ethanol at -40°C
2,4 kW
Cooling power with Ethanol at -60°C
1,5 kW
Cooling power with Ethanol at -80°C
0,3 kW
Cooling power with Ethanol at -85°C
0,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,95 kg
CO2 equivalent
2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,39 kg
CO2 equivalent 2nd stage
5,8 t
UN-number
UN 2857
Order-No.: 1079.0002.01
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V 3-phase 50 Hz supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and breaker sizing before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 15 A Fuse 3x16A”
Installation
The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, with G1/2 male connections.
Defines the required cooling water supply pressure and connection type for proper operation.
Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Operation
Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if it can drop below the dew point.
Operating outside this range reduces cooling power or causes condensation, affecting reliability.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 2.4 kW at -40 °C to only 0.2 kW at -85 °C.
Critical for selecting the unit when deep cooling loads are expected near the lower limit.
Source
“at -80°C 0,3 kW at -85°C 0,2 kW”
Safety
The refrigerant R-452A has a GWP of 2141 and 0.95 kg charge, giving 2.0 t CO2 equivalent, which matters for F-gas compliance.
Affects environmental compliance, leak monitoring obligations and refrigerant handling.
Source
“Global Warming Potential (GWP) 2141 Refrigerant quantity 0,95 kg CO2 equivalent 2,0 t”
Safety
The second-stage refrigerant R-23 has a very high GWP of 14800 with 0.39 kg charge (5.8 t CO2e).
High-GWP refrigerant drives regulatory reporting and service requirements.
Source
“Global Warming Potential (GWP) 14800 Refrigerant quantity 2nd stage 0,39 kg CO2 equivalent 2nd stage 5,8 t”
Installation
Voltage and frequency tolerances are ±5% and ±2% but must not combine in the unfavorable direction (e.g., -5% voltage with +2% frequency is not allowed).
Prevents motor and electronics damage from out-of-tolerance supply conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices under 63 A are delivered with cable but without a plug, so a plug must be sourced and fitted on site.
Affects installation planning and electrical connection work.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The pump delivers 40 l/min max and 0.9 bar max pressure, with flow falling to 18 l/min at 0.5 bar.
Needed to verify the external circuit pressure drop and achievable flow.