Introduces the Unistat 950w unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 20 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,0 bar
167 l/min
Delivery at 3,0 bar
115 l/min
Pump connection
flange DN40 PN10/40 EN1092-1
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G1 1/4 male
Consumption at water 15°C, flow 0°C
5000 l/h
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
30 l
Filling capacity expansion tank
90 l
Overall dimensions WxDxH **
2630x1300x1980 mm
Net weight
2147 kg
Power supply requirement
400V 3~ 50Hz
max. current
120 A
Fuse
3x125A
Pressure equipment category
IV
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S513334
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-90...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
36 kW
Cooling power with Thermooil at 200°C
36 kW
Cooling power with Thermooil at 100°C
36 kW
Cooling power with Thermooil at 20°C
36 kW
Cooling power with Ethanol at 0°C
36 kW
Cooling power with Ethanol at -20°C
36 kW
Cooling power with Ethanol at -40°C
36 kW
Cooling power with Ethanol at -60°C
25 kW
Cooling power with Ethanol at -80°C
10 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-507 (A1, H280)
Global Warming Potential (GWP)
3985
Refrigerant quantity
19,5 kg
CO2 equivalent
77,7 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
7 kg
CO2 equivalent 2nd stage
103,6 t
Circulation pump: MK pump max. delivery
240 l/min
max. delivery pressure
4 bar
Delivery at 1,0 bar
207 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V 3-phase 50 Hz supply with a maximum current of 120 A and 3x125 A fusing, so the electrical infrastructure must be sized accordingly.
Determines the site power distribution, cable sizing and breaker rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 120 A Fuse 3x125A”
Installation
The cooling water connection is G1 1/4 male and requires 5000 l/h at 15 °C with a minimum differential pressure of 3 bar and maximum pressure of 6 bar.
Defines the cooling water supply capacity and pressure limits that must be available at the installation site.
Source
“Consumption at water 15°C, flow 0°C 5000 l/h min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Operation
The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling power drops as the cooling water temperature rises.
Cooling water temperature directly affects achievable cooling capacity and must be kept within this range.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and corrosion on the cooling water circuit at low intake temperatures.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to avoid corrosion or damage to the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable supply quality; out-of-tolerance combinations can damage the unit or void compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices drawing more than 63 A are supplied without cable and without plug, so the customer must provide the power connection.
Affects installation planning and cost since the power cable and plug are not included.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Safety
The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.
Indicates the safety category for flammable or non-flammable fluids and the required protective measures.
Source
“Safety classification III / FL”
Safety
The refrigerant R-507 has a GWP of 3985 with 19.5 kg charge (77.7 t CO2 equivalent), and the second stage R-23 has a GWP of 14800 with 7 kg charge (103.6 t CO2 equivalent).
Relevant for refrigerant regulations, leak monitoring and environmental compliance obligations.
Source
“Global Warming Potential (GWP) 3985 Refrigerant quantity 19,5 kg CO2 equivalent 77,7 t”
Installation
The unit has overall dimensions of 2630x1300x1980 mm and a net weight of 2147 kg, requiring adequate floor space and structural support.
Needed for site layout, transport route planning and floor loading assessment.
Source
“Overall dimensions WxDxH ** 2630x1300x1980 mm Net weight 2147 kg”
Operation
The ambient temperature must be between 5 °C and 40 °C, and the degree of protection is IP20.
Defines the permissible operating environment and limits on dust and water exposure.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The pump connection is a flange DN40 PN10/40 according to EN1092-1, and pipework must use ball socket per DIN 3863 and sleeve nut per DIN 3870.
Determines the mating flange and hose fittings required for connecting the external circuit.
Source
“Pump connection flange DN40 PN10/40 EN1092-1”
Performance
The circulation pump delivers a maximum of 240 l/min at 4 bar, with 207 l/min at 1.0 bar, 167 l/min at 2.0 bar and 115 l/min at 3.0 bar.
Used to verify that the external circuit pressure drop is compatible with the available pump curve.