Peter Huber Kältemaschinenbau Unistat 950w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 950w Dynamisches

Peter Huber Kältemaschinenbau Unistat 950w Dynamisches — Temperature control units.

The Unistat 950w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and a -90…200 °C working range. It is the largest modification in the Unistat w Dynamisches series, distinguished by 36 kW heating and a 240 l/min pump. It serves as a high-capacity circulation source for externally closed industrial processes requiring precise temperature control.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 950w Dynamisches Current —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 90 l
Filling volume 30 l
Heating capacity 36 kW
Pump connection flange DN40 PN10/40 EN1092-1
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...200 °C
Dimensions, Weight & Mechanical Design
Weight 2147 kg
Electrical, Power & Drive Specifications
Current consumption 120 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1 1/4 male
Cooling water consumption 5000 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 3 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply tolerance is conditional on frequency tolerance

Voltage may deviate ±5% only if frequency simultaneously deviates ±2%; combinations such as -5% voltage with +2% frequency are not allowed. Verify the site supply quality before connection, because out-of-tolerance combinations can damage the unit or void compliance.

Unistat 950w Technical Data, page 2
Electrical
Power cable and plug are not included above 63 A

Because the unit draws 120 A, it is delivered without cable and without plug. The customer must provide the three-phase power connection, sized for 120 A with 3×125 A fusing, and have the hard-wired connection performed by an electrician.

Unistat 950w Technical Data, page 2; Unistat Operation Manual, page 37
Cooling & Media
Cooling water intake must stay between 5 and 20 °C

The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling power drops as the cooling water temperature rises. If the cooling water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation and corrosion.

Unistat 950w Technical Data, page 2
Cooling & Media
Cooling water circuit materials restrict water quality

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only cooling water compatible with these materials may be used; aggressive or untreated water can corrode the circuit.

Unistat 950w Technical Data, page 2
Mounting
Unit must not be installed inside an ATEX zone

Without an Ex px cabinet the unit is not protected against explosion. Site selection must exclude ATEX zones; operation in an Ex px cabinet requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.

Unistat Operation Manual, Section 1.3.3, page 15-16
Mounting
Transport upright and empty; casters are not for transport

Only transport the unit upright and emptied; horizontal transport damages the compressor and a filled unit risks thermal fluid overflow. Casters, if present, are not suitable for transport and carry only 25% of total mass symmetrically — use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, page 22
Process Limits
Expansion vessel lock state depends on application type

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Verify the lock state before filling and commissioning.

Unistat Operation Manual, page 33
Process Limits
Heat transfer fluid viscosity must not exceed 100 mm²/s

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s. At low temperatures viscosity rises, so fluid selection must ensure this limit is respected across the entire operating range down to -90 °C.

Unistat 950w Technical Data, page 2
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; configure this before relying on Process Safety for exothermic process protection.

Unistat Operation Manual, page 19
Control & I/O
Power-failure restart behavior must be chosen deliberately

After a power outage or when switching on, the response can be set via Pilot ONE: 'Off/Standby' (default, temperature control started only by manual input) or 'ON/Temperature control active' (temperature control always started after turning on, with an INFO shown for a few seconds allowing automatic startup to be disabled). Choose according to whether unattended restart is acceptable for the process.

Unistat Operation Manual, Section 1.8.1, page 21
Maintenance
Shut off cooling water even for overnight downtimes

Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight, to prevent water damage and unnecessary load on the cooling water circuit.

Unistat Operation Manual, page 29
Maintenance
Refrigerant circuit work requires approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. General maintenance personnel cannot perform refrigerant-circuit maintenance. The unit contains R-507 (19.5 kg) and R-23 (7 kg), with CO2 equivalents of 77.7 t and 103.6 t respectively.

Unistat Operation Manual, Section 1.4.2, page 19; Unistat 950w Technical Data, page 1

Selection & Integration Q&A

When should I choose the Unistat 950w Dynamisches instead of the smaller 705w or 815w siblings?

Choose the 950w when the process needs 36 kW heating and up to 240 l/min pump delivery at 4 bar; the 705w and 815w are smaller siblings. The 950w is the largest listed w Dynamisches model, so select it for high heat-load or high-flow external circuits where the smaller units cannot meet the duty point.

What power supply and protection must I provide for this unit, and what is not included?

Provide 400 V 3~ 50 Hz with 120 A max current and 3×125 A fusing. Because the unit draws more than 63 A, it is supplied without cable and without plug; the customer must provide the three-phase connection, performed by an electrician. Voltage tolerance is ±5% only with simultaneous ±2% frequency tolerance.

What are the site and duty-cycle limits for installing the 950w?

Install outside ATEX zones, with ambient temperature 5…40 °C and IP20 protection. Transport upright and empty; casters are not for transport. Allow about 2 hours acclimation when moving between cold and warm environments before power-on. Cooling water intake must be 5…20 °C at 5000 l/h, min 3 bar differential, max 6 bar.

What maintenance items and accessories are not included in base delivery?

The power cable and plug are not included for this >63 A unit. A condensation water/heat transfer fluid drip tray must be installed by the responsible body. Hoses and connections must be periodically checked and replaced as needed. Cooling water must be shut off even overnight. Refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 950w Technical Data
Datasheet ·58,8 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF