Unistat GL P535w Temperature Control System Datasheet

Contents

  1. Overview p. 1
    Introduces the Unistat GL P535w temperature control unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 21 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C2100 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight600 kg
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat GL P535w (Order-No.: 5022.0002.01)

ParameterValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C23 kW
Cooling power with Thermooil at 100°C23 kW
Cooling power with Thermooil at 20°C23 kW
Cooling power with Ethanol at 0°C20 kW
Cooling power with Ethanol at -20°C12,5 kW
Cooling power with Ethanol at -40°C5,5 kW
Cooling power with Ethanol at -50°C2,2 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Refrigerant quantity3,3 kg
CO2 equivalent0,0 t
Circulation pump: MK pump max. delivery201 l/min
Circulation pump: MK pump max. delivery pressure5,3 bar
Circulation pump: MK pump Delivery at 1,0 bar179 l/min
Circulation pump: MK pump Delivery at 2,0 bar154 l/min

Значения приведены дословно; строки охлаждения сведены по параметру и среде (Thermooil/Ethanol) с указанием температуры.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase power connection.

Determines electrical infrastructure and whether a plug or hard wiring is needed.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the installer must provide a suitable connection.

Affects electrical installation planning and procurement of a plug or isolator.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water must be supplied at 5–20 °C inlet temperature with 1–4 bar differential pressure and no more than 6 bar pressure.

Cooling water conditions directly affect cooling performance and safe operation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the dew point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and possible damage or electrical risk.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so actual capacity may differ on site.

Needed to correctly compare quoted cooling capacity with real site conditions.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Performance

Higher cooling water temperature reduces cooling power and may require a higher cooling water flow rate.

Affects sizing of the cooling water supply and expected process performance.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion or damage from unsuitable water chemistry.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions.

Defines acceptable supply quality for reliable operation and warranty.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Compliance

The equipment is EMC class A group 1, which may require special precautions in residential or sensitive environments.

Affects installation location and possible interference with nearby equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.

Determines environmental limits and protection against dust and water.

Source
“Degree of Protection IP20”
Installation

The unit weighs 600 kg and measures 730×804×1738 mm, so floor loading and access routes must be checked.

Needed for transport, rigging and installation space planning.

Source
“Overall dimensions WxDxH ** 730x804x1738 mm”
Operation

The pump connections are M38×1.5 male with bore shape Y (60°) per DIN 3863, requiring matching ball socket and sleeve nut fittings.

Ensures correct hose or pipe fittings are ordered and leak-free connections are made.

Source
“Pump connection M38x1,5 male”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Limits fluid selection and ensures pump performance and protection.

Source
“max. permissible kin. viscosity 100 mm²/s”
Safety

The unit uses R-744 (CO2) refrigerant with GWP 1 and 3.3 kg charge, which has low global warming impact but requires high-pressure awareness.

Relevant for refrigerant handling, safety classification and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280)”