Unistat P545w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat P545w unit
  2. Spec p. 2
    Technical data according to DIN 12876 including performance parameters
  3. p. 2 table, 21 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1630 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity9,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/23

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ХарактеристикаUnistat P545w
Order-No.5012.0002.01
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 250°C35 kW
Cooling power with Thermooil at 200°C35 kW
Cooling power with Thermooil at 100°C35 kW
Cooling power with Thermooil at 20°C35 kW
Cooling power with Ethanol at 0°C32 kW
Cooling power with Ethanol at -20°C16 kW
Cooling power with Ethanol at -40°C4 kW
Cooling power with Ethanol at -50°C1,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
Circulation pump:MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a water-cooled refrigerated heating circulator with 24 kW heating and up to 35 kW cooling, suitable for externally closed and open applications.

Confirms the unit type and that it can serve both closed and open external circuits when selecting for a process.

Source
“Refrigerated Heating Circulator with water-cooled cooling machine and electronical level indicator.”
Selection

The 'P' model pump is designed for high pressure drops, making it suitable for flow-through chemistry or semiconductor applications.

Guides selection toward this model when the external circuit has high pressure drop.

Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation

Cooling water must be supplied between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.

Defines the required cooling water supply conditions and condensation protection for installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit requires a differential pressure between 1 bar and 4 bar, with a maximum cooling water pressure of 6 bar.

Ensures the site cooling water supply meets the minimum and maximum pressure limits for correct operation.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit requires a 400 V three-phase 50 Hz supply and draws high power, so the electrical supply must be sized accordingly.

Critical for electrical planning and ensuring the site provides the correct three-phase supply.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Prevents installation on a supply where the combined voltage/frequency deviation exceeds the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices with current consumption less than 63 A are delivered with cable but without a plug.

The installer must provide and fit a suitable plug or connection for this three-phase unit.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

Cooling power drops as cooling water temperature rises, and cooling water flow rate may need to increase.

Operators must account for reduced cooling capacity if the site cooling water is warmer than the rated 15 °C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation

The pump delivers 201 l/min maximum and 5.3 bar maximum pressure, with delivery falling as pressure rises.

Allows checking whether the pump curve suits the external circuit's flow and pressure drop requirements.

Source
“max. delivery 201 l/min”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Prevents selecting a fluid too viscous for the pump and circuit at operating temperature.

Source
“max. permissible kin. viscosity 100 mm²/s”
Safety

The refrigerant R-452A is classified A1 with GHS hazard H280, and the unit contains 3.6 kg with a CO2 equivalent of 7.7 t.

Needed for refrigerant handling, leak reporting and environmental compliance at the installation site.

Source
“R-452A (A1, H280)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms the safety device required for unattended operation and for protecting the external application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The unit must be operated within an ambient temperature range of 5 °C to 40 °C.

Ensures the installation room stays within the allowed ambient conditions for reliable operation.

Source
“max. ambient temperature 40 °C”
Compliance

The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.

Class A equipment may require EMC precautions in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”