No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P545w Dynamisches

Peter Huber Kältemaschinenbau Unistat P545w Dynamisches — Temperature control units.

The Unistat P545w Dynamisches is a water-cooled refrigerated heating circulator with a magnetically coupled stainless-steel pump built for high pressure drop circuits, covering -55 to 250 °C with 24 kW heating. Its 'P' pump (201 l/min, 5.3 bar max) distinguishes it from standard-pressure siblings for flow-through chemistry and semiconductor applications.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P545w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1630 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and operation

Integration & Operating Notes

Process Limits
Cooling water must stay within 5–20 °C intake

The unit is designed for cooling-water intake between 5 °C and 20 °C; warmer water reduces cooling power and may require higher flow. Insulate cooling-water pipes where water temperature can fall below the condensation point to avoid condensation inside the unit.

Unistat P545w Technical Data, page 2
Electrical
Combined voltage/frequency deviation envelope is restricted

Voltage tolerance is ±5% and frequency ±2%, but the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply against this envelope before connection; the three-phase unit is delivered with cable but without a plug, so the installer must provide the connection.

Unistat P545w Technical Data, page 2
Mounting
Expansion vessel lock state depends on circuit type

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. Confirm the circuit type before commissioning.

Unistat Operation Manual, page 33
Process Limits
Heat transfer fluid viscosity capped at 100 mm²/s

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s; selecting a fluid that exceeds this at operating temperature overloads the pump and circuit. For FL-rated fluids the fire point must be ≥65 °C, and ethanol use permits only supervised operation.

Unistat P545w Technical Data, page 2; Unistat Operation Manual, Section 1.7
Control & I/O
Process safety mode needs compressor set to Permanently ON

In 'Process safety' mode the controller releases maximum cooling capacity at the OT trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON. Configure the restart behavior (Off/Standby vs ON/Temperature control active) according to whether unattended restart is acceptable.

Unistat Operation Manual, page 19–21
Maintenance
Shut off cooling water even for overnight downtime

Close the cooling-water supply during short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Periodically check hoses and connections for leaks and quality and replace as required; work on the refrigerant circuit must be done by approved refrigeration/air-conditioning contractors.

Unistat Operation Manual, page 29; Section 1.4.2
Mounting
ATEX-zone installation is prohibited

The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on the Ex px cabinet rating plate is suitable. Site selection must exclude hazardous areas.

Unistat Operation Manual, Section 1.3.3 / page 15–16

Selection & Integration Q&A

When should I choose the P545w Dynamisches instead of the 510w/515w/520w/525 Dynamisches siblings?

Choose the P545w when the external circuit has high pressure drop (flow-through chemistry, semiconductor) and needs the 'P' high-discharge pump (201 l/min, 5.3 bar) plus water cooling; the 5xx Dynamisches siblings are lower-capacity alternatives for less demanding circuits.

What cooling-water supply and piping must the site provide for this unit?

G3/4 male connection; intake 5–20 °C, differential pressure 1–4 bar, max pressure 6 bar; consumption 1630 l/h at 15 °C, flow 0 °C. Insulate pipes where water can fall below the condensation point and shut off supply during downtimes.

What maintenance and delivery items are not included with the base unit?

Three-phase units below 63 A ship with cable but without a plug; the site must provide the plug/connection and a drip tray. Periodic hose/connection checks and cooling-water shutoff are operator duties; refrigerant work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P545w Technical Data
Datasheet ·86,1 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF