Product introduction and main features of Unistat P825
Specpp. 1-2
Technical data and performance specifications per DIN 12876
Specp. 2
Detailed technical parameters and operating limits
p. 2table, 22 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Circulation pump
MK pump
max. delivery
67 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
61 l/min
Delivery at 1,0 bar
55 l/min
Delivery at 1,5 bar
48 l/min
Delivery at 2,0 bar
39 l/min
Delivery at 2,5 bar
28 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
2,9 l
Filling capacity expansion tank
3,6 l
Overall dimensions WxDxH **
460x604x1465 mm
Net weight
238 kg
Power supply factory configured (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
16 A
Fuse (3 phase)
3x20A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S502294
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P825 Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-85...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power with Thermooil at 250°C
2,3 kW
Cooling power with Thermooil at 200°C
2,3 kW
Cooling power with Thermooil at 100°C
2,3 kW
Cooling power with Ethanol at 0°C
2,2 kW
Cooling power with Ethanol at -20°C
2 kW
Cooling power with Ethanol at -40°C
2 kW
Cooling power with Ethanol at -60°C
1,4 kW
Cooling power with Ethanol at -80°C
0,3 kW
Cooling power with Ethanol at -85°C
0,2 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,8 kg
CO2-Äquvivalent
3,9 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,39 kg
CO2-Äquvivalent 2. Stufe
5,8 t
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator with an air-cooled cooling machine, suitable for both externally closed and externally open applications.
Confirms the product type and that it can serve open external loops, which affects loop design and accessories.
Source
“Refrigerated Heating Circulator with air-cooled cooling machine and optical level indicator.”
Selection
The P models use high-discharge-pressure circulation pumps intended for applications with high pressure drops such as flow-through chemistry or semiconductor industry.
Determines whether this pump variant matches a high-pressure-drop process loop.
Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation
The unit requires a three-phase 400 V 50 Hz supply with 16 A max current and 3x20 A fusing.
Electrical supply and protection must be planned before installation.
Three-phase devices with current consumption less than 63 A are delivered with cable but without a plug.
The installer must provide and fit a suitable plug or hard-wire connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, and hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures correct mating fittings and avoids leaks or mismatched connections.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation
The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient temperature.
Restricts installation location to dry indoor areas within the ambient temperature range.
Source
“Degree of Protection IP20”
Operation
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.
Performance must be derated for warmer rooms; affects process capability.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
Voltage and frequency tolerances are ±5% and ±2% respectively, but combined worst-case opposite tolerances are not allowed.
Defines acceptable mains quality for reliable operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.
Affects electrical safety design and residual current device choice.
Source
“Attention: leakage current > 3,5mA”
Safety
The refrigerant R-452A has GWP 2141 with 1.8 kg charge (3.9 t CO2 equivalent), and the second stage R-23 has GWP 14800 with 0.39 kg (5.8 t CO2 equivalent).
Relevant for refrigerant regulations, leak checks, and environmental compliance.
Source
“Refrigerant quantity 1,8 kg”
Compliance
The equipment is classified Class A, Group 1 for electromagnetic disturbance per EN55011.
Class A equipment may require EMC measures in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Compliance
The equipment complies with US-SNAP and applicable EU laws, with US-SNAP end-use as industrial process refrigeration.
Confirms regulatory acceptance for industrial process refrigeration use.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Maintenance
The refrigeration machine is air-cooled and uses CFC- and HCFC-free refrigerant.
Requires adequate air flow and condenser cleaning; refrigerant type affects service and disposal.
Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Performance
The circulation pump delivers up to 67 l/min and 3 bar, with 61 l/min at 0.5 bar and 28 l/min at 2.5 bar.
Needed to verify the pump can overcome the pressure drop of the external loop.