Unistat P915w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat P915w unit
  2. Spec pp. 2-3
    Technical data according to DIN 12876 including performance parameters
  3. p. 2 table, 28 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
CO2 equivalent 2nd stage18,2 t
Circulation pumpMK pump
max. delivery110 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar89 l/min
Delivery at 1,0 bar75 l/min
Delivery at 1,5 bar60 l/min
Delivery at 2,0 bar43 l/min
Delivery at 2,5 bar24 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C720 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,9 l
Filling capacity expansion tank6,5 l
Overall dimensions WxDxH **630x704x1565 mm
Net weight369 kg
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply requirement400V 3~ 50Hz
max. current27 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S419528

** Overall dimensions WxDxH

Technical data according to DIN 12876

ParameterValue
Operating temperature range-90...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C6,5 kW
Cooling power with Thermooil at 200°C7,5 kW
Cooling power with Thermooil at 150°C9 kW
Cooling power with Thermooil at 100°C11 kW
Cooling power with Ethanol at 0°C11 kW
Cooling power with Ethanol at -20°C11 kW
Cooling power with Ethanol at -40°C8,2 kW
Cooling power with Ethanol at -60°C4,2 kW
Cooling power with Ethanol at -80°C1,3 kW
Cooling power with Ethanol at -90°C0,6 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-507 (A1, H280)
Global Warming Potential (GWP)3985
Refrigerant quantity2,9 kg
CO2 equivalent11,6 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage1,23 kg

Order-No.: 1080.0008.01; Unistat P915w

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a very wide temperature range from -90 to 250 °C, so it can serve both deep-freeze and high-temperature processes with one circulator.

Defines whether the unit can cover the full process window without a second thermostat.

Source
“Operating temperature range -90...250 °C”
Performance

Cooling capacity collapses at low temperatures (0.6 kW at -90 °C versus 11 kW at 0 °C with ethanol), so deep-cold loads must be sized against the low-temperature end.

Prevents oversizing expectations and ensures heat loads are checked at the coldest operating point.

Source
“at -90°C 0,6 kW”
Installation

The refrigeration machine is water-cooled and requires cooling water between 5 and 20 °C, so a suitable chilled-water supply must be available at the installation site.

Site services and cooling-water temperature must be verified before purchase and installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water differential pressure must be at least 1 bar and must not exceed 6 bar, which dictates the supply pump and any pressure-reducing valve.

Ensures adequate flow without overpressuring the unit's cooling circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit draws up to 27 A on a 400 V three-phase supply and needs 3x32 A fusing, so the electrical feed and protection must be sized accordingly.

Determines cable cross-section, breaker rating and whether the existing site supply is sufficient.

Source
“max. current 27 A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which requires proper earthing and may call for a higher-rated residual current device.

Impacts electrical safety design and RCD selection during installation.

Source
“Attention: leakage current > 3,5mA”
Compliance

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.

Prevents operation on unstable or generator-supplied mains that could damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant R-507 has a GWP of 3985 and a CO2 equivalent of 11.6 t, and the second stage R-23 has a GWP of 14800 and 18.2 t CO2e, creating significant F-gas reporting obligations.

Affects environmental compliance, leak-check frequency and total cost of ownership.

Source
“Global Warming Potential (GWP) 3985”
Operation

The pump delivers 110 l/min maximum and 3 bar maximum pressure, with flow falling to 24 l/min at 2.5 bar, so external circuit pressure drop must be matched to the pump curve.

Ensures the external loop and hoses do not exceed the available pump head.

Source
“max. delivery pressure 3 bar”
Installation

The unit requires at least 3.9 l filling capacity and has a 6.5 l expansion tank, so the external volume plus thermal expansion must fit within this budget.

Prevents overflow or insufficient fill when connecting external consumers.

Source
“min. filling capacity 3,9 l”
Installation

The unit weighs 369 kg and measures 630x704x1565 mm, so floor loading, door widths and lifting equipment must be planned.

Determines transport route, floor reinforcement and rigging requirements.

Source
“Net weight 369 kg”
Operation

Ambient temperature must stay between 5 and 40 °C, and the sound pressure level is 64 dB(A), which constrains the room environment and operator exposure.

Ensures the room ventilation and acoustic environment are suitable for continuous operation.

Source
“max. ambient temperature 40 °C”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used to avoid corrosion.

Guides water quality specification and prevents premature fouling or corrosion of the heat exchanger.

Source
“Please use suitable cooling water.”