Peter Huber Kältemaschinenbau Unistat P915w Dynamisches

Peter Huber Kältemaschinenbau Unistat P915w Dynamisches — Temperature control units with Noise level 64 dB.

The Unistat P915w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 250 °C. It is the water-cooled 'w' variant of the 9xx series, distinguished from air-cooled siblings by its cooling-water circuit (G3/4 male, 720 l/h at 15 °C). It serves single-thermostat deep-freeze and high-temperature processes in externally closed or open circuits.

Noise level
64 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P915w Dynamisches Current 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Noise level 64 dB(A)
Control output POKO ONE
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...250 °C
Dimensions, Weight & Mechanical Design
Weight 369 kg
Electrical, Power & Drive Specifications
Current consumption 27 A (max.)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 720 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior across the temperature range

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must stay within 5–20 °C and 1–6 bar differential

The water-cooled refrigeration circuit is designed for intake temperatures between 5 and 20 °C; supply pressure below 1 bar differential starves the condenser, while above 6 bar risks overpressuring the cooling circuit. Verify the site chilled-water supply and fit a pressure-reducing valve if necessary.

Unistat P915w Technical Data, page 2
Electrical
Plan for 27 A draw, 3x32 A fusing and no supplied plug

The unit draws up to 27 A on 400 V three-phase and requires 3x32 A fusing; three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug. Leakage current exceeds 3.5 mA, so earthing and RCD selection must be engineered accordingly.

Unistat P915w Technical Data, page 2
Process Limits
Size deep-cold loads against 0.6 kW at -90 °C, not 11 kW at 0 °C

Cooling capacity collapses at low temperatures (0.6 kW at -90 °C versus 11 kW at 0 °C with ethanol), so heat loads for deep-freeze processes must be checked at the coldest operating point rather than at ambient. Oversizing expectations based on the 0 °C figure will lead to unachievable pull-down times.

Unistat P915w Technical Data, page 1
Control & I/O
Choose power-failure behavior before unattended operation

The Pilot ONE can be set to 'Off/Standby' (default, temperature control started only by manual input) or 'ON/Temperature control active' (automatic restart after power restoration, with a brief INFO allowing cancellation). Select according to whether unattended restart is acceptable for the process; Process Safety mode additionally requires the compressor set to Permanently ON.

Unistat Operation Manual, Section 1.8.1 / page 21

Selection & Integration Q&A

When should I choose the P915w Dynamisches over the air-cooled Unistat 815w Dynamisches or the smaller 705w/1005w/1015w Dynamisches models?

Choose the P915w when a chilled-water supply (5–20 °C, 1–6 bar differential) is available and the 9xx frame's capacity is needed; the 815w is air-cooled at 60 dB(A) and avoids water connections, while the 705w/1005w/1015w are smaller frames for lower loads.

What cooling-water and electrical services must be in place before the P915w can be commissioned?

Cooling water at 5–20 °C with G3/4 male connection, 720 l/h at 15 °C, and 1–6 bar differential; a 400 V 3~ 50 Hz supply fused at 3x32 A with a site-fitted plug (cable supplied without plug), plus earthing for >3.5 mA leakage current.

What maintenance items and accessories are not included in the base delivery and must be planned separately?

The site plug, drip tray, cooling-water hoses and connections, and heat transfer fluid are not included; periodic hose inspection/replacement, cooling-water shutoff for downtimes, and refrigerant-circuit work by approved contractors must be scheduled by the operator.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P915w Technical Data
Datasheet ·80,3 КБ ·3 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF