Introduces the Unistat P920w unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 26 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Circulation pump
MK pump
max. delivery
191 l/min
max. delivery pressure
5,5 bar
Delivery at 1,0 bar
167 l/min
Delivery at 2,0 bar
140 l/min
Delivery at 3,0 bar
107 l/min
Delivery at 4,0 bar
67 l/min
Delivery at 5,0 bar
22 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
730 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
12 l
Filling capacity expansion tank
29 l
Overall dimensions WxDxH **
950x1205x1650 mm
Net weight
916 kg
Power supply requirement
400V 3~ 50Hz
max. current
38 A
Fuse
3x40A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S460015
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Unistat P920w
Order-No.
1061.0011.01
Operating temperature range
-90...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
11 kW
Cooling power with Thermooil at 100°C
11 kW
Cooling power with Thermooil at 20°C
11 kW
Cooling power with Ethanol at 0°C
11 kW
Cooling power with Ethanol at -20°C
11 kW
Cooling power with Ethanol at -40°C
10 kW
Cooling power with Ethanol at -60°C
8 kW
Cooling power with Ethanol at -80°C
2 kW
Cooling power with Ethanol at -90°C
1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
4,55 kg
CO2 equivalent
9,7 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
1,9 kg
CO2 equivalent 2nd stage
28,1 t
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit is a water-cooled refrigerated heating circulator with a magnetic-coupled stainless steel pump, suitable for both externally closed and externally open applications.
Confirms cooling medium and application compatibility before specifying the unit.
Source
“Refrigerated Heating Circulator with water-cooled cooling machine and optical level indicator. Magnetic coupled circulation pump made of stainless steel.”
Selection
The 'P' model pump delivers high discharge pressure for systems with high pressure drops such as flow-through chemistry or semiconductor processes.
Determines whether this pump variant suits a high-pressure-drop loop.
Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation
Cooling water must be supplied between 5°C and 20°C; below the condensation point the cooling water pipes must be insulated.
Sets the facility cooling water requirement and insulation work at installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Rated output data assume 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power.
Prevents overestimating cooling capacity when site conditions differ from the rating basis.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility checks for the cooling water supply.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable supply quality for safe commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
Leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.
Requires appropriate earth connection and residual current device planning.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.
Affects electrical installation scope and ordering of the mains connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Pump connections use bore shape Y (60°) per DIN 3863 with ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures correct mating fittings and hoses are procured for the piping.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Confirms built-in safety function required for unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
The refrigerant is R-452A with GWP 2141 and 4.55 kg charge (9.7 t CO2e), plus R-23 second stage with GWP 14800 and 1.9 kg (28.1 t CO2e).
Relevant for refrigerant handling, leak checks and environmental reporting.