Unistat P930w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat P930w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 25 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
max. delivery191 l/min
max. delivery pressure5,5 bar
Delivery at 1,0 bar167 l/min
Delivery at 2,0 bar140 l/min
Delivery at 3,0 bar107 l/min
Delivery at 4,0 bar67 l/min
Delivery at 5,0 bar22 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1300 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity12 l
Filling capacity expansion tank29 l
Overall dimensions WxDxH **950x1205x1650 mm
Net weight986 kg
Power supply requirement400V 3~ 50Hz
max. current66 A
Fuse3x80A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S449594

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ХарактеристикаUnistat P930w
Order-No.1082.0003.01
Operating temperature range-90...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C19 kW
Cooling power with Thermooil at 100°C19 kW
Cooling power with Thermooil at 20°C19 kW
Cooling power with Ethanol at 0°C20 kW
Cooling power with Ethanol at -20°C20 kW
Cooling power with Ethanol at -40°C20 kW
Cooling power with Ethanol at -60°C15 kW
Cooling power with Ethanol at -80°C5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity5 kg
CO2 equivalent10,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage2,2 kg
CO2 equivalent 2nd stage32,6 t
Circulation pump:MK pump

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a 400 V 3-phase 50 Hz supply with a 3x80 A fuse, so the electrical infrastructure must be sized accordingly.

Determines the site power distribution, cable sizing and protection device rating before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices drawing more than 63 A are supplied without cable and without plug, so the customer must provide the power connection.

Affects procurement and installation planning of the electrical supply.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation

Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the pipes must be insulated if condensation can occur.

Defines the cooling water utility requirements and condensation protection needed at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion and material incompatibility in the cooling water loop.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The pump connection is M38x1.5 male with a Y-shaped bore per DIN 3863, so mating hoses must use ball socket DIN 3863 and sleeve nut DIN 3870.

Ensures correct mechanical mating of process piping and avoids leaks or mismatched fittings.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.

Restricts installation location regarding moisture and ambient temperature.

Source
“Degree of Protection IP20”
Safety

Leakage current exceeds 3.5 mA, so protective earth connection and residual current protection must be designed for high leakage.

Affects earth bonding and RCD selection to avoid nuisance tripping and ensure safety.

Source
“Attention: leakage current > 3,5mA”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination -5% voltage with +2% frequency is not allowed.

Defines the acceptable grid quality for reliable operation and warranty compliance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Selection

Cooling power drops sharply at low temperatures with ethanol, from 20 kW at -40 °C to 5 kW at -80 °C, so low-temperature duty must be checked against the load.

Prevents oversizing expectations and ensures the process heat load is covered at the lowest operating temperature.

Source
“at -80°C 5 kW”
Selection

The refrigerant charge gives a CO2 equivalent of 10.7 t for R-452A and 32.6 t for R-23, which matters for F-gas regulations and leak monitoring.

Determines regulatory obligations for refrigerant containment and periodic leak checks.

Source
“CO2 equivalent 32,6 t”
Performance

Pump delivery falls from 191 l/min at zero head to 22 l/min at 5.0 bar, so the process pressure drop must be matched to the pump curve.

Ensures adequate circulation flow through high-pressure-drop external circuits.

Source
“Delivery at 5,0 bar 22 l/min”
Operation

The unit needs a minimum filling capacity of 12 l and has a 29 l expansion tank, so the external loop volume must be considered when charging heat transfer fluid.

Prevents underfilling and overflow issues during commissioning.

Source
“min. filling capacity 12 l”
Maintenance

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, so fluid selection must respect this limit at the lowest operating temperature.

Avoids pump overload and poor circulation with viscous fluids at low temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”