Peter Huber Kältemaschinenbau Unistat P930w Dynamisches
Peter Huber Kältemaschinenbau Unistat P930w Dynamisches — Temperature control units.
The Unistat P930w Dynamisches is a water-cooled, high-capacity temperature control unit with a 24 kW heater and a -90…200 °C working range, distinguished from smaller Unistat w models by its 3×80 A supply and 986 kg free-standing build. It serves as a central temperature source for large externally closed or open circuits requiring high heating and cooling power.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P930w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 29 l |
| Filling volume | 12 l |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Weight | 986 kg |
| Current consumption | 66 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1300 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing, water-cooled (w) Unistat with magnetically coupled stainless-steel pump and Pilot ONE controller; no plate heat exchanger is listed for this model in the manual, unlike some smaller siblings.
- Within the Unistat w Dynamisches family, the P930w is the largest frame referenced here (986 kg, 3×80 A supply); siblings such as the 705w, 815w, 1005w and 1015w are smaller-capacity alternatives.
- The 'P' designation indicates suitability for applications with high pressure drop, and the pump curve confirms delivery falling from 191 l/min at zero head to 22 l/min at 5.0 bar.
Working range and load behavior
- Temperature range -90…200 °C with 24 kW heating capacity; cooling power is strongly temperature-dependent (e.g. 20 kW at -40 °C down to 5 kW at -80 °C with ethanol), so low-temperature duty must be checked against the actual process load.
- Heat transfer fluid viscosity must not exceed 50 mm²/s at the lowest operating temperature; the 12 l minimum filling volume and 29 l expansion volume define the chargeable circuit size.
- Process Safety mode is available to maintain maximum cooling at the overtemperature trigger instead of tripping all poles, but requires the compressor set to Permanently ON.
Installation, connection and control
- Requires a 400 V 3~ 50 Hz supply with 3×80 A fusing; because current exceeds 63 A, the unit is delivered without cable and plug, so the customer must provide the hard-wired connection by an electrician.
- Cooling water: 5–20 °C intake, 1–6 bar, G3/4 male connection, 1300 l/h consumption at 15 °C; circuit materials (copper, 1.4401, MS, PA, PPE, PTFE, EPDM) restrict water quality, and pipes must be insulated where condensation can occur.
- Pump ports are M38×1.5 male with DIN 3863 Y-bore; mating hoses require DIN 3863 ball socket and DIN 3870 sleeve nut.
- Pilot ONE controller with POKO ONE output, 0.01 K display resolution, Pt100 external sensor input, and Ethernet/USB/RS232 interfaces; IP20 rating and 5–40 °C ambient limit restrict siting.
Integration & Operating Notes
The 3×80 A three-phase draw means the unit arrives without a power cable or plug; the site must provide a hard-wired connection sized for 66 A and protected accordingly, executed by an electrician. Leakage current above 3.5 mA also requires RCD selection and PE bonding designed for high leakage.
The water circuit uses copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Intake temperature outside 5–20 °C, differential pressure below 1 bar or supply above 6 bar will impair cooling or damage components; insulate pipes where condensation can form and shut off water even for overnight downtime.
The magnetically coupled stainless-steel pump cannot handle fluids thicker than 50 mm²/s; select the heat transfer fluid so that this limit is respected at the lowest operating temperature, otherwise circulation stalls and temperature control is lost. Also respect the FL fire-point requirement of ≥65 °C if a flammable fluid is used.
Horizontal transport damages the compressor, and only an emptied unit may be moved. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation-induced electrical faults. Leveling feet must be activated and caster brakes engaged before commissioning.
If Process Safety is used to keep cooling active at the overtemperature trigger, the compressor must be configured to Permanently ON; otherwise the safety function will not deliver maximum cooling. Also decide the power-failure restart behavior (Off/Standby vs ON) according to whether unattended restart is acceptable.
Selection & Integration Q&A
Choose the P930w when the process needs the highest heating capacity (24 kW) and a -90…200 °C range with a large circuit (12 l min fill, 29 l expansion) and high pressure-drop duty. Smaller siblings like the 815w or 1015w suit lower loads and smaller footprints; the P930w also demands a 3×80 A supply, which smaller units may not.
Pump ports are M38×1.5 male with DIN 3863 Y-bore, so use DIN 3863 ball socket and DIN 3870 sleeve nut hoses. Cooling water connects via G3/4 male at 5–20 °C, 1–6 bar, 1300 l/h at 15 °C; ensure the water is compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors, not in an ATEX zone, at 5–40 °C ambient and IP20 conditions; provide a drip tray and level floor. Transport upright and empty, acclimate 2 h before power-on, and respect the ±5% voltage / ±2% frequency tolerance (the -5% V with +2% Hz combination is not allowed).
The unit ships without power cable and plug (3×80 A), so the customer supplies the hard-wired connection. Maintenance includes periodic hose and mechanical-seal inspection, cleaning liquefier fins and the hat-type strainer, and F-gas leak checks; refrigerant work requires certified contractors. A drip tray and compatible heat transfer fluid are also site-provided.