Peter Huber Kältemaschinenbau Unistat P930w Dynamisches

Peter Huber Kältemaschinenbau Unistat P930w Dynamisches — Temperature control units.

The Unistat P930w Dynamisches is a water-cooled, high-capacity temperature control unit with a 24 kW heater and a -90…200 °C working range, distinguished from smaller Unistat w models by its 3×80 A supply and 986 kg free-standing build. It serves as a central temperature source for large externally closed or open circuits requiring high heating and cooling power.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P930w Dynamisches Current —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 29 l
Filling volume 12 l
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...200 °C
Dimensions, Weight & Mechanical Design
Weight 986 kg
Electrical, Power & Drive Specifications
Current consumption 66 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1300 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and load behavior

Installation, connection and control

Integration & Operating Notes

Electrical
Supply exceeds 63 A: no cable or plug supplied

The 3×80 A three-phase draw means the unit arrives without a power cable or plug; the site must provide a hard-wired connection sized for 66 A and protected accordingly, executed by an electrician. Leakage current above 3.5 mA also requires RCD selection and PE bonding designed for high leakage.

Unistat P930w Technical Data, page 2; Operation Manual, Section 1.8 / page 21
Cooling & Media
Cooling water must stay within 5–20 °C and 1–6 bar

The water circuit uses copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Intake temperature outside 5–20 °C, differential pressure below 1 bar or supply above 6 bar will impair cooling or damage components; insulate pipes where condensation can form and shut off water even for overnight downtime.

Unistat P930w Technical Data, page 2; Operation Manual, page 29
Process Limits
Fluid viscosity must not exceed 50 mm²/s at the coldest setpoint

The magnetically coupled stainless-steel pump cannot handle fluids thicker than 50 mm²/s; select the heat transfer fluid so that this limit is respected at the lowest operating temperature, otherwise circulation stalls and temperature control is lost. Also respect the FL fire-point requirement of ≥65 °C if a flammable fluid is used.

Unistat P930w Technical Data, page 2; Operation Manual, Section 1.7 / page 19
Mounting
Transport upright and empty; acclimate 2 h before power-on

Horizontal transport damages the compressor, and only an emptied unit may be moved. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation-induced electrical faults. Leveling feet must be activated and caster brakes engaged before commissioning.

Unistat Operation Manual, page 22 and page 29
Control & I/O
Process Safety mode needs compressor set to Permanently ON

If Process Safety is used to keep cooling active at the overtemperature trigger, the compressor must be configured to Permanently ON; otherwise the safety function will not deliver maximum cooling. Also decide the power-failure restart behavior (Off/Standby vs ON) according to whether unattended restart is acceptable.

Unistat Operation Manual, Section 1.7.2 / page 20 and Section 1.8.1 / page 21

Selection & Integration Q&A

When should I choose the P930w instead of a smaller Unistat w Dynamisches such as the 815w or 1015w?

Choose the P930w when the process needs the highest heating capacity (24 kW) and a -90…200 °C range with a large circuit (12 l min fill, 29 l expansion) and high pressure-drop duty. Smaller siblings like the 815w or 1015w suit lower loads and smaller footprints; the P930w also demands a 3×80 A supply, which smaller units may not.

What must I provide on the process side to connect the pump and cooling water correctly?

Pump ports are M38×1.5 male with DIN 3863 Y-bore, so use DIN 3863 ball socket and DIN 3870 sleeve nut hoses. Cooling water connects via G3/4 male at 5–20 °C, 1–6 bar, 1300 l/h at 15 °C; ensure the water is compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM.

What are the site and duty-cycle limits for installing this unit?

Install indoors, not in an ATEX zone, at 5–40 °C ambient and IP20 conditions; provide a drip tray and level floor. Transport upright and empty, acclimate 2 h before power-on, and respect the ±5% voltage / ±2% frequency tolerance (the -5% V with +2% Hz combination is not allowed).

What maintenance and accessories are required beyond the base delivery?

The unit ships without power cable and plug (3×80 A), so the customer supplies the hard-wired connection. Maintenance includes periodic hose and mechanical-seal inspection, cleaning liquefier fins and the hat-type strainer, and F-gas leak checks; refrigerant work requires certified contractors. A drip tray and compatible heat transfer fluid are also site-provided.