Induztr
Engineering Digest Document Viewer Related Models

Unistat T330w HT Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat T330w HT unit
  2. Spec pp. 1-2
    Technical performance data, dimensions, and operating parameters
  3. Quality p. 2
    Technical data according to DIN 12876 classification and safety standards
  4. p. 2 table, 7 rows
  5. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Filling capacity expansion tank24 l
Overall dimensions WxDxH **540x749x1184 mm
Power supply (3 Phase)400V 3~ 50Hz
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:1.0/25

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat T330w HT — Technical data according to DIN 12876

CharacteristicValue
Operating temperature range65...300 °C
Minimum temperature with water cooling15 °C
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power at 300°C18 kW
Cooling power at 200°C18 kW
Cooling power at 100°C10 kW
Circulation pump: MK pump
max. delivery96 l/min
max. delivery pressure3,5 bar
Delivery at 0,5 bar82 l/min
Delivery at 1,0 bar72 l/min
Delivery at 1,5 bar60 l/min
Delivery at 2,0 bar45 l/min
Delivery at 2,5 bar33 l/min
Delivery at 3,0 bar20 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water consumption at water temp. 15°C860 l/h
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity7 l

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit has no active refrigeration, so achieving temperatures down to 65°C requires either sufficient heat loss from the application or suitable water cooling.

Critical for selecting the unit: it cannot actively cool below 65°C and depends on process heat loss or water cooling.

Source
“These devices do not have an active refrigeration system.”
Installation

The cooling water supply must be between 5°C and 20°C; if it can drop below the dew point, the cooling water pipes inside the unit must be insulated.

Defines the required cooling water infrastructure and condensation protection during installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water connection is G3/4 male and requires a minimum differential pressure of 1 bar and maximum pressure of 6 bar.

Needed to specify matching fittings and ensure the water supply meets pressure requirements.

Source
“Cooling water connection G3/4 male”
Installation

The pump connection is M30x1.5 male and requires pipework or hoses with ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures correct mating of process hoses and avoids leaks or mechanical mismatch.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without a plug, so the customer must provide a suitable plug and connection.

Affects electrical installation planning and the need to procure a compatible plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has a leakage current greater than 3.5 mA, which must be considered for protective earth and RCD selection.

Important for electrical safety design and compliance with local installation codes.

Source
“Attention: leakage current > 3,5mA”
Operation

The cooling water exit temperature is limited to 55°C and no steam exits when cooling with circulation temperatures over 100°C.

Relevant for safe operation and for planning the cooling water return line.

Source
“Water exit temperature limited to 55°C. No steam exits on cooling with circulation temperatures over 100°C.”
Operation

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.

Limits the choice of heat transfer fluid and ensures pump performance and protection.

Source
“max. permissible kin. viscosity 50 mm²/s”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects sizing of the cooling water supply and expected performance at elevated water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Compliance

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but the combination of -5% voltage and +2% frequency is not allowed.

Must be checked against the site power quality to avoid invalid supply conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit has protection degree IP20 and must be installed in an environment with ambient temperature between 5°C and 40°C.

Defines environmental limits and protection against ingress for installation location.

Source
“Degree of Protection IP20”
Selection

The minimum filling capacity is 7 l and the expansion tank capacity is 24 l, which must be considered for system volume and expansion.

Needed to ensure the external system volume and expansion requirements are compatible.

Source
“min. filling capacity 7 l”