Peter Huber Kältemaschinenbau Unistat T330w HT Dynamisches
Peter Huber Kältemaschinenbau Unistat T330w HT Dynamisches — Temperature control units.
The Unistat T330w HT is a water-cooled, dynamically temperature-controlled unit with a 24 kW heater and a 65–300 °C range (15–300 °C with water cooling), using a magnetically coupled stainless-steel pump. It has no active refrigeration, so sub-65 °C operation depends on process heat loss or cooling water. It serves high-temperature process loops requiring precise, fast temperature control.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T330w HT Dynamisches Current | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 24 l |
| Filling volume | 7 l (min) |
| Heating capacity | 24 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C with water cooling: 15...300 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unistat T330w HT is a water-cooled high-temperature variant in the Unistat T3xx series. Unlike the T305/T305 HT siblings, it provides 24 kW heating capacity and a 65–300 °C range (15–300 °C with water cooling). The 'w' denotes water cooling; the 'HT' denotes high-temperature capability. The unit uses a magnetically coupled stainless-steel pump with adjustable speed, a Pilot ONE controller with POKO ONE control output, and a Pt100 external sensor input. It has no active refrigeration system, so it cannot actively cool below 65 °C without water cooling or sufficient process heat loss.
Behavior under load and media
Heating capacity is 24 kW. The temperature range is 65–300 °C without water cooling and 15–300 °C with water cooling. Cooling power decreases as cooling water temperature rises, and the required cooling water flow rate may increase. The cooling water exit temperature is limited to 55 °C, and no steam exits when cooling with circulation temperatures over 100 °C. The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s. The expansion tank capacity is 24 l, and the minimum filling capacity is 7 l.
Installation, connection and control
Cooling water connection is G3/4 male, requiring a minimum differential pressure of 1 bar and maximum pressure of 6 bar; supply temperature must be 5–20 °C, and pipes must be insulated if condensation is possible. Pump connection is M30x1.5 male, requiring ball socket per DIN 3863 and sleeve nut per DIN 3870. Power supply is 400 V 3~ 50 Hz; three-phase devices under 63 A are supplied with cable but without a plug. Leakage current exceeds 3.5 mA. Protection rating is IP20; ambient temperature must be 5–40 °C. The unit must not be installed in an ATEX zone. Transport and store upright only; move only when emptied. Allow about 2 hours acclimation after temperature changes before power-on. For externally closed applications, the expansion vessel must remain unlocked; for externally open applications, it must be locked. Hard-wired power connection must be done by an electrician. Interfaces include Ethernet, USB (Host and Device), and RS232. The controller supports Process Safety mode (requires compressor set to Permanently ON) and Power Failure Auto (resumes temperature control after power outage if active). Shut off cooling water supply even for short downtimes.
Integration & Operating Notes
The unit cannot actively cool below 65 °C. To reach temperatures down to 15 °C, water cooling must be connected and operating, or the application must provide sufficient heat loss. Exceeding this limit without cooling water risks invalid results or unit damage.
Cooling water supply temperature must be between 5 °C and 20 °C. If the water temperature can drop below the ambient dew point, internal cooling water pipes must be insulated to prevent condensation. Cooling water pressure must be 1–6 bar differential.
The unit's leakage current exceeds 3.5 mA. Protective earth and residual current device (RCD) selection must account for this to ensure electrical safety and compliance with local installation codes.
Devices with current consumption less than 63 A are delivered with a cable but no plug. The customer must provide a suitable plug and have the connection made by an electrician. Supply voltage and frequency must match the rating plate.
The unit must not be installed or started inside an ATEX zone. Explosion protection requires an Ex px cabinet. Ignoring this creates an immediate ignition hazard.
Transport and store the unit upright only; horizontal transport damages the compressor. Move only an emptied unit to avoid thermal fluid overflow. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck.
For externally closed applications, the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications, the expansion vessel must be locked to prevent fluid loss.
When using Process Safety mode, the compressor must be set to Permanently ON to keep cooling active at the over-temperature trigger. This ensures cooling remains available for exothermic process protection.
If temperature control is active during a power outage, the process will automatically continue after the power outage when Power Failure Auto is enabled. Assess unattended operation risks accordingly.
Shut off the cooling water supply to the unit even for shorter downtimes such as overnight. This prevents water damage and unnecessary load on the cooling water circuit.
Hoses and hose connections must be checked periodically for leaks and quality, and replaced as needed. This preventive maintenance avoids fluid loss and ensures safe operation.
Selection & Integration Q&A
Choose the T330w HT when you need 24 kW heating and a 65–300 °C range (15–300 °C with water cooling) for high-temperature processes. The T305 variants have lower heating capacity; select them for smaller loads or lower temperature requirements.
Cooling water: G3/4 male, 5–20 °C, 1–6 bar differential pressure. Pump: M30x1.5 male, requiring ball socket per DIN 3863 and sleeve nut per DIN 3870. Insulate cooling water pipes if below dew point.
Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water even overnight. A power plug is not included for three-phase units under 63 A; customer must provide. No other consumables specified.