Contents

  1. Overview p. 1
    Product name and general introduction to the Unistat T350 HT unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 25 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range65... 300 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power96 kW
Cooling power at 300°C30 kW
Circulation pump:MK pump
max. delivery200 l/min
max. delivery pressure5,5 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Filling capacity expansion tank56 l
Overall dimensions WxDxH **800x1007x2200 mm
Power supply (3 Phase)400V 3~ 50Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/25

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a very wide operating temperature range from 65 to 300 °C, so it suits high-temperature processes rather than sub-ambient work.

Defines the process window the circulator can serve and rules it out for cooling below 65 °C.

Source
“Operating temperature range 65... 300 °C”
Performance

Heating power is 96 kW and cooling power at 300 °C is 30 kW, so the heat exchanger and facility power must be sized for these loads.

Determines electrical supply sizing and the achievable heat-up/cool-down rates for the process.

Source
“Heating power 96 kW”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase feed.

Dictates the electrical infrastructure and rules out single-phase installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.

Affects site preparation and the electrical bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and opposite-sign deviations are not allowed.

Prevents selecting a supply point whose combined voltage/frequency deviation would violate the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for high leakage.

Impacts earth connection design and choice of residual current protection.

Source
“Attention: leakage current > 3,5mA”
Compliance

The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.

Constrains where the unit may be installed and what EMC measures are needed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The enclosure is IP20, so it must be installed in a dry, protected location away from water and dust ingress.

Limits the permissible installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 and 40 °C, so the room climate must be controlled for reliable operation.

Defines the site climate envelope and cooling/ventilation needs of the installation room.

Source
“min. ambient temperature 5 °C”
Installation

The unit is 800 x 1007 x 2200 mm, so space and access around it must be planned accordingly.

Needed for layout, transport route and maintenance access planning.

Source
“Overall dimensions WxDxH ** 800x1007x2200 mm”
Selection

The pump delivers up to 200 l/min at 5.5 bar, so external circuit pressure drop must be checked against this curve.

Determines whether the pump can serve the external loop and what pressure losses are acceptable.

Source
“max. delivery 200 l/min”
Selection

Maximum permissible kinematic viscosity is 50 mm²/s, so the thermofluid must be chosen within this limit at operating temperature.

Constrains thermofluid selection and pump loading.

Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance

The expansion tank holds 56 l and is not thermoregulated, so fluid expansion volume and venting must be planned.

Affects fluid fill volume, expansion allowance and system venting design.

Source
“Filling capacity expansion tank 56 l”
Safety

The unit has adjustable overtemperature protection per DIN 12876, so the safety limit must be set for the application.

Requires commissioning step to set the overtemperature cut-out for safe operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”