Peter Huber Kältemaschinenbau Unistat T350 HT Dynamisches
Peter Huber Kältemaschinenbau Unistat T350 HT Dynamisches — Temperature control units.
The Unistat T350 HT Dynamisches is a high-temperature dynamic temperature control unit with a 65–300 °C working range, 96 kW heating and 300 kW cooling at 30 °C. It is the larger sibling of the T305/T305 HT models, distinguished by its 300 kW cooling capacity and 56 l expansion volume. It serves high-temperature process loops requiring rapid, precise temperature control via the Pilot ONE controller.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T350 HT Dynamisches Current | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 56 l |
| Heating capacity | 96 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 300 kW (at 30 °C) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The T350 HT Dynamisches is a floor-standing unit (800 × 1007 × 2200 mm) with a magnetically coupled stainless steel pump and M38×1.5 male pump connections.
- Within the Unistat series, it sits above the T305 and T305 HT variants; the T305w HT is a different configuration. The T350 HT is the high-capacity member with 300 kW cooling at 30 °C and 96 kW heating.
- The unit uses the Pilot ONE controller with POKO ONE control output, 0.01 K display resolution, and Ethernet, USB (host and device) and RS232 interfaces.
Behavior and process window
- Operating range is 65 to 300 °C, so it is not suitable for sub-ambient or low-temperature work; it is intended for high-temperature processes.
- The pump delivers up to 200 l/min at 5.5 bar; external circuit pressure drop must be checked against this curve. Pump speed is adjustable.
- Maximum permissible kinematic viscosity is 50 mm²/s, constraining thermofluid selection at operating temperature.
- The expansion tank holds 56 l and is not thermoregulated; fluid expansion and venting must be planned.
Installation, control and operation
- Requires a 400 V three-phase 50 Hz supply; three-phase devices under 63 A are delivered with cable but without a plug. Hard-wired connection must be by an electrician.
- Leakage current exceeds 3.5 mA, so earthing and RCD selection must account for high leakage. EMC class A group 1 for industrial environments.
- IP20 enclosure requires a dry, protected location; ambient temperature must stay 5–40 °C.
- The unit has adjustable overtemperature protection per DIN 12876; the safety limit must be set at commissioning.
- For externally closed applications the expansion vessel must remain unlocked; for externally open applications it must be locked.
Integration & Operating Notes
The unit requires 400 V 3~ 50 Hz. Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and opposite-sign deviations (e.g. -5 % voltage with +2 % frequency) are not allowed. Verify the site supply against this envelope before connection.
Three-phase devices drawing less than 63 A are delivered with cable but without a plug. The customer must provide and fit a suitable plug, and the hard-wired connection must be carried out by an electrician.
Leakage current exceeds 3.5 mA. Protective earthing must be designed accordingly, and residual current device selection must account for high leakage to avoid nuisance tripping or inadequate protection.
Only transport and store the unit upright and emptied; horizontal transport damages the compressor and filled transport risks thermal fluid overflow. Casters are not suitable for transport and carry only 25 % of total mass symmetrically; use an industrial truck. Activate parking brakes and leveling feet before commissioning.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss. Set this correctly before operation.
Maximum permissible kinematic viscosity is 50 mm²/s. Select a thermofluid whose viscosity at the lowest operating temperature remains within this limit to avoid pump overload and poor heat transfer.
The pump delivers up to 200 l/min at 5.5 bar. Check the pressure drop of the external circuit (hoses, heat exchanger, application) against the pump curve to ensure the required flow is achieved; pump speed is adjustable.
When using Process Safety mode, the compressor must be set to Permanently ON so that cooling remains active at the overtemperature trigger. Configure this in the controller before relying on the safety function.
If temperature control is active during a power outage, the process will automatically continue after power is restored when Power Failure Auto is enabled. Assess whether unattended restart is acceptable for the process.
Shut off the cooling water supply to the unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit.
Check periodically for leaks and the quality of the hose and hose connections and replace as required. This is part of preventive maintenance for the fluid circuit.
Do not install or start up the temperature control unit within an ATEX zone. Explosion protection requires an Ex px cabinet. Verify the area classification before installation.
When moving the unit between cold and warm environments, allow about 2 hours acclimation before power-on to prevent condensation-induced electrical faults.
Selection & Integration Q&A
Choose the T350 HT when your process needs the higher cooling capacity (300 kW at 30 °C) and 96 kW heating in the 65–300 °C range. The T305/T305 HT are smaller siblings; the T305w HT is a different configuration. Select based on required heat removal and flow (up to 200 l/min).
Provide a 400 V 3~ 50 Hz supply with voltage/frequency tolerances coupled (±5 % V only with ±2 % F, no opposite-sign deviations). Leakage current >3.5 mA requires earth and RCD review. Devices under 63 A come with cable but no plug; hard-wiring by an electrician.
Install upright and level on a dry, protected location (IP20), ambient 5–40 °C, not in an ATEX zone. Transport only upright and emptied; casters not for transport. Allow 2 h acclimation after temperature changes. Activate parking brakes and leveling feet before commissioning.
Periodically check hoses and connections for leaks and quality, replace as needed; shut off cooling water even overnight. The base delivery includes cable but no plug for three-phase devices under 63 A. Thermofluid and external circuit components are not included and must be selected within 50 mm²/s viscosity.