Check ventilation to pump. Check ambient temperature.
Debris in pump
Check intake screen. Flush pump. Disassemble pump and inspect.
Pump runs intermittently
Motor thermal protector is cycling open and closed.*
Allow the motor to cool. Identify the cause of the overload.
* The IDP-15 is equipped with an auto-reset thermal motor protector. This protector automatically shuts down the pump when it detects an overload condition, and automatically restarts the pump when the motor has cooled to within an acceptable temperature range.
Table 9 IDP-15 Service Kits and Advance Exchange Pumps
Part Number
Description
Contents
X3815-67000
Replacement Tip Seal Set
Replacement Tip Seal, O-ring, and grease
X3815-69000
Exchange IDP-15 Standard
IDP-15 Exchange Pumps
X3815-69010
Exchange IDP-15 Pumps with VPI Valve
IDP-15 VPI Valve Exchange Pumps
Request for Return Form (Health and Safety Certification)
Field
Value
1) CUSTOMER INFORMATION — Company Name
Contact Name
Tel
Email
Fax
Customer Ship To
Customer Bill To
Europe only: VAT reg. Number
USA/Canada only: Taxable
USA/Canada only: Non-taxable
2) PRODUCT IDENTIFICATION — Product Description
Agilent P/N
Agilent S/N
Original Purchasing Reference
3) TYPE OF RETURN — 3A. Non-Billable
3A. Billable
3A. New PO # (hard copy must be submitted with this form)
3B. Exchange
3B. Repair
3B. Upgrade
3B. Consignment/Demo
3B. Calibration
3B. Evaluation
3B. Return for Credit
4) HEALTH and SAFETY CERTIFICATION — HAS NOT pumped or been exposed to any toxic or hazardous materials
HAS pumped or been exposed to the following toxic or hazardous materials
Toxic
Corrosive
Reactive
Flammable
Explosive
Biological
Radioactive
List all toxic/hazardous materials. Include product name, chemical name, and chemical symbol or formula
Print Name
Authorized Signature
Date
5) FAILURE INFORMATION — Failure Mode (REQUIRED FIELD. See next page for suggestions of failure terms)
Detailed Description of Malfunction: (Please provide the error message)
Application (system and model)
Форма содержит чекбоксы и поля для заполнения; пустые ячейки оставлены пустыми строками.
Engineering Facts
14 facts · every claim linked to a page
Selection
The pump is only suitable for air or inert gases and must not be used on corrosive, explosive, or particulate-forming gases.
Determines whether the pump can be applied to a given process gas at all.
Source
“This pump is suitable for pumping air or inert gases. The pump is not intended to pump corrosive, explosive, or particulate-forming gases.”
Installation
Shipping locks must be disengaged before operation or the pump will not float on its isolation mounts.
Prevents damage and ensures vibration isolation is active at startup.
Source
“Before operating the IDP-15 disengage the shipping locks.”
Installation
Shipping locks must be re-engaged for transport, long moves, or long-term storage.
Protects the scroll set and isolation mounts during handling and storage.
Source
“Engage the shipping locks (Figure 3) when shipping, moving a long distance or during long-term storage.”
Installation
The voltage selector must be set to match the actual mains supply before powering up.
Wrong setting risks motor damage and invalidates the electrical setup.
Source
“Slide the voltage selector switch to 115 for 100-120 V mains, or to 230 for 200-240 V mains.”
Installation
The pump must be fed from a 15 A circuit at 100/120 V to avoid nuisance breaker trips on start.
Sizing the branch circuit correctly avoids spurious trips during motor start.
Source
“For 100/120 V service, connect the IDP-15 to a 15 A circuit.”
Installation
A minimum 2-inch clearance at front and rear is required to keep fan ducts unobstructed.
Prevents overheating and shortened pump life in enclosures or racks.
Source
“Leave a minimum gap of 2" between front and back of the pump.”
Installation
Exhaust must not be restricted by long small-diameter tubing to avoid overheating.
Guides exhaust plumbing diameter and length during system integration.
Source
“To avoid overheating the pump, do not restrict the exhaust flow with long lengths of small diameter tubing.”
Installation
A bellows between pump and chamber provides both vibration isolation and strain relief.
Protects the chamber and piping from pump vibration and mechanical stress.
Source
“Insert a bellows between the pump and vacuum chamber to provide both vibration isolation and strain relief.”
Operation
The pump needs about an hour of running with the inlet sealed to reach ultimate vacuum.
Sets realistic expectations for commissioning and acceptance testing.
Source
“In order to reach maximum vacuum, the pump must be left running for about an hour with the inlet sealed.”
Operation
Gas ballast port 2 handles high water vapor loads while port 1 suits moderate loads.
Selecting the correct ballast port prevents excessive tip seal wear.
Source
“Gas ballast port 2 is used for high water vapor loads”
Operation
For air-sensitive applications dry nitrogen at about 5 l/min can be fed to the gas ballast inlet.
Allows inert ballast purge to avoid air ingress in sensitive processes.
Source
“dry nitrogen at a flow rate of approximately 5 l/min can be provided to the gas ballast inlet”
Maintenance
The pump has no oil for cleaning, so periodic short runs with the inlet open flush out debris.
Defines a simple preventive maintenance routine for contaminated processes.
Source
“Run the pump periodically with the inlet open to atmosphere for a minute or two to flush it out.”
Maintenance
Newly installed tip seals may need several hours of run time to seat and meet specifications.
Prevents premature rejection of a rebuilt pump during performance checks.
Source
“Newly installed tip seals may require several hours of run time to seat properly”
Safety
The gas ballast must be sealed whenever pumping gas not intended to be vented to atmosphere.
Avoids uncontrolled release of hazardous process gas to the room.
Source
“The gas ballast must be sealed whenever pumping any gas not intended to be vented to the atmosphere”