Labconco Protector® Laboratory Fume Hoods catalog covering Protector XStream® (4', 5', 6', 8' widths), Protector Premier® with remote blower (4', 5', 6', 8') and with built-in blower (4', 5', 6'), Protector XL™ (3', 4', 5', 6', 7', 8', 10', 12', 16' widths; exterior depths 31.7", 37.7", 43.7"), plus SpillStopper™ work surfaces and cupsinks.

Technical Data

Parameter Value Condition
Hood width 4' ft p.10 ↗ XStream, 4' Hood
Hood width 5' ft p.10 ↗ XStream, 5' Hood
Hood width 6' ft p.10 ↗ XStream, 6' Hood
Hood width 8' ft p.10 ↗ XStream, 8' Hood
Airflow at 100 fpm sash open 705 CFM p.10 ↗ XStream, 4' Hood
Airflow at 100 fpm sash open 930 CFM p.10 ↗ XStream, 5' Hood
Airflow at 100 fpm sash open 1150 CFM p.10 ↗ XStream, 6' Hood
Airflow at 100 fpm sash open 1600 CFM p.10 ↗ XStream, 8' Hood
Static pressure at 100 fpm sash open 0.26 in. s.p. p.10 ↗ XStream, 4' Hood
Static pressure at 100 fpm sash open 0.32 in. s.p. p.10 ↗ XStream, 5' Hood
Static pressure at 100 fpm sash open 0.41 in. s.p. p.10 ↗ XStream, 6' Hood
Static pressure at 100 fpm sash open 0.29 in. s.p. p.10 ↗ XStream, 8' Hood
Airflow at 100 fpm sash open 725 CFM p.12 ↗ Premier, 4' Hood
Airflow at 100 fpm sash open 955 CFM p.12 ↗ Premier, 5' Hood
Airflow at 100 fpm sash open 1180 CFM p.12 ↗ Premier, 6' Hood
Airflow at 100 fpm sash open 1640 CFM p.12 ↗ Premier, 8' Hood
Static pressure at 100 fpm sash open 0.22 in. s.p. p.12 ↗ Premier, 4' Hood
Static pressure at 100 fpm sash open 0.31 in. s.p. p.12 ↗ Premier, 5' Hood
Static pressure at 100 fpm sash open 0.41 in. s.p. p.12 ↗ Premier, 6' Hood
Static pressure at 100 fpm sash open 0.28 in. s.p. p.12 ↗ Premier, 8' Hood
Built-in blower airflow 725 CFM p.14 ↗ Premier built-in blower, 4 Feet
Built-in blower airflow 955 CFM p.14 ↗ Premier built-in blower, 5 Feet
Built-in blower airflow 1180 CFM p.14 ↗ Premier built-in blower, 6 Feet
Built-in blower maximum external static pressure 0.17 in. S.P. p.14 ↗ Premier built-in blower, 4 Feet
Built-in blower motor HP 1/3 HP p.14 ↗ Premier built-in blower, 4 Feet

Operating Limits

  • Electrical requirement 100-115V: 50/60 Hz, 10A p.10 ↗
  • Electrical requirement 208-230V: 50/60 Hz, 5A p.10 ↗
  • Built-in blower electrical requirement 100-115V: 60 Hz, 10A p.15 ↗
  • Built-in blower electrical requirement 208-230V: 50 Hz, 5A p.15 ↗
  • Built-in blower electrical requirement 208-230V: 60 Hz, 5A p.15 ↗
  • LED light rated life: 50,000 hours p.5 ↗

Type Code Decoder

Position Token Meaning
5-6 p.10 ↗ 00, 20 Electrical requirements: 00 = 100-115V, 50/60 Hz, 10A; 20 = 208-230V, 50/60 Hz, 5A
7-8 p.10 ↗ 00, 02, 20, 21 Service fixtures/receptacle: 00 = no fixtures; 02 = two fixtures GFCI duplex; 20 = two fixtures; 21 = two fixtures GFCI duplex
5-6 p.13 ↗ 00, 20 Electrical requirements: 00 = 100-115V, 50/60 Hz, 10A; 20 = 208-230V, 50/60 Hz, 5A
7-8 p.13 ↗ 00, 02, 20, 21 Service fixtures/receptacle: 00 = no fixtures; 02 = two fixtures GFCI duplex; 20 = two fixtures; 21 = two fixtures GFCI duplex