HiPace 1200 with TC 1200 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiPace 1200 and HiPace 1500 turbomolecular vacuum pumps (3-page datasheet/brochure, dated 03.07.2025). Covers both models' performance data, flange configurations (DN 200 ISO-K/ISO-F/CF-F for HiPace 1200; DN 250 ISO-K/ISO-F/CF-F for HiPace 1500), and accessory/options portfolio (electrical adapters, mains/connecting cables, display unit, safety/mounting kits, venting and sealing gas accessories).
Section-by-Section Technical Breakdown
Positioning of the HiPace 1200/1500 series as high-performance turbomolecular pumps. Emphasizes high pumping speed, excellent compression for light gases, hybrid bearing technology, and nickel-coated surfaces enabling use in corrosive applications. Lists accessory and option categories.
- High pumping speed with excellent compression for light gases
- Best-in-class hybrid bearing technology for maximum uptime
- Nickel-coated surfaces for corrosive applications
- Accessories/spare parts/options: electrical adapter, mains cable, connecting cable, display unit, safety and mounting kits, venting accessories, sealing gas accessories
Tabulated performance data for N₂, H₂, He and Ar for both models. Pumping speeds range 1100–1400 l/s depending on gas and model; compression ratios span 6·10³ (H₂) to 1·10⁸ (N₂, Ar). Gas throughput values are given at final rotational speed and at 0.1 hPa HV pressure, plus maximum forevacuum pressure.
- Pumping speed N₂: 1250 l/s (HiPace 1200) / 1400 l/s (HiPace 1500)
- Pumping speed H₂: 1100 / 1150 l/s; He: 1300 / 1350 l/s; Ar: 1200 / 1350 l/s
- Compression ratio: N₂ 1·10⁸, H₂ 6·10³, He 6·10⁴, Ar 1·10⁸ (identical for both models)
- Gas throughput at final rotational speed: N₂ 20 hPa·l/s, H₂ 30 hPa·l/s, He 30 hPa·l/s, Ar 11 hPa·l/s
- Gas throughput at 0.1 hPa HV pressure: He 15 hPa·l/s, H₂ 8 hPa·l/s, N₂ 26 hPa·l/s
- Maximum forevacuum pressure for N₂: 2 hPa (both models)
General specifications: run-up time, noise level, weight, communication interfaces, allowed mounting orientations, and flange sizes. Both models are very similar; HiPace 1500 uses a larger DN 250 inlet flange and weighs slightly more.
- Run-up time: 2.5 min for both models
- Sound pressure level (EN ISO 2151): 50 dBA
- Weight: approx. 30.5 kg (HiPace 1200) / 31.2 kg (HiPace 1500)
- Standard communication interfaces: RS-485, remote
- Mounting orientation: 0° (HV flange top) – 90° / 90° – 180° (HV flange bottom)
- Gas inlet/outlet: HiPace 1200 DN 200 ISO-K/ISO-F/CF-F / DN 40 ISO-KF; HiPace 1500 DN 250 ISO-K/ISO-F/CF-F / DN 40 ISO-KF
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed for N₂ is 1250 l/s (HiPace 1200) and 1400 l/s (HiPace 1500). p.2 ↗ | Size the backing pump and process chamber conductance around these values; the 1500 offers ~12% more N₂ speed for larger chambers. | HiPace 1200 / HiPace 1500 |
| Compression ratio for H₂ is only 6·10³ while N₂ and Ar reach 1·10⁸. p.2 ↗ | For hydrogen-containing processes, ultimate pressure is fore-vacuum limited; a low base pressure backing pump is critical. | Both models |
| Maximum permissible forevacuum pressure for N₂ is 2 hPa. p.2 ↗ | The backing pump must maintain ≤2 hPa at the foreline under full gas throughput or the pump will not sustain operation. | Both models |
| Gas throughput at final rotational speed for Ar is 11 hPa·l/s versus 30 hPa·l/s for H₂/He. p.2 ↗ | Heavy-gas (Ar, e.g. sputtering) processes are throughput-limited; check process gas load against the Ar limit, not the light-gas figures. | Both models |
| Run-up time is 2.5 min for both models. p.3 ↗ | Venting/valve sequencing and process start interlocks must allow ~2.5 min from start to full speed. | Both models |
| Sound pressure level is 50 dBA per EN ISO 2151. p.3 ↗ | No special acoustic enclosure needed for typical lab noise limits; relevant for workplace noise assessments. | Both models |
| Weights are approx. 30.5 kg (HiPace 1200) and 31.2 kg (HiPace 1500). p.3 ↗ | Mounting frames/rigging must support ~31 kg per pump; single-person handling with lifting aids. | HiPace 1200 / HiPace 1500 |
| Standard communication interfaces are RS-485 and remote. p.3 ↗ | Integration into PLC/monitoring systems via RS-485; the optional display unit accessory adds local HMI. | Both models |
| Permitted mounting orientation spans 0° (HV flange top) to 90° and 90° to 180° (HV flange bottom). p.3 ↗ | The pump can be mounted with the high-vacuum flange pointing up, sideways, or down, giving flexibility in chamber integration. | Both models |
| Gas inlet flanges: DN 200 ISO-K/ISO-F/CF-F (HiPace 1200) and DN 250 ISO-K/ISO-F/CF-F (HiPace 1500); fore-vacuum outlet DN 40 ISO-KF on both. p.3 ↗ | Select mating chamber flange hardware accordingly; the larger DN 250 inlet of the 1500 requires different clamping/ sealing components. | HiPace 1200 (DN 200) / HiPace 1500 (DN 250) |
| Surfaces are nickel-coated for corrosive applications. p.1 ↗ | Suitable for processes with corrosive gases without custom coating; still verify compatibility with specific media and use purges/venting accessories as needed. | Both models |
| Hybrid bearing technology is cited as best-in-class for maximum uptime. p.1 ↗ | Hybrid (magnetic/mechanical) bearings reduce maintenance intervals; factor into service planning versus fully oil-lubricated alternatives. | Both models |
| Venting accessories and sealing gas accessories are available as options. p.1 ↗ | Plan venting (for run-down protection) and inert sealing gas purge (for corrosive/particle processes) in the installation design. | Both models |