PVG-500 and PVG-502 Vacuum Gauge User Manual

Contents

  1. p. 2
    Legal and copyright notices for the document
  2. Overview p. 3
    Table of contents listing manual sections
  3. Procedure p. 4
    Instructions for connecting power to the gauge
  4. Procedure p. 5
    General operating instructions for the gauge
  5. Overview p. 6
    Purpose and structure of the manual
  6. Overview p. 7
    Key terms and definitions used in the manual
  7. Warning p. 8
    Explanation of safety warning symbols
  8. Warning p. 9
    General safety information and precautions
  9. Warning p. 10
    Warnings against incorrect use of the gauge
  10. Warning p. 11
    Specific safety guidelines for PVG-500 and PVG-502
  11. p. 12
    Instructions for proper disposal of the product
  12. Services p. 13
    Service and support contact information
  13. Spec p. 14
    Technical data and operating parameters
  14. Overview p. 15
    Description of the gauge and its components
  15. Spec pp. 16-17
    Detailed technical specifications and limits
  16. Spec p. 18
    Dimensional drawings for PVG-500 and PVG-502
  17. Spec pp. 19-20
    Measurement behavior for different gases
  18. Procedure pp. 21-22
    Installation instructions and notes
  19. Procedure p. 23
    Electrical power connection procedure
  20. Procedure p. 24
    Operating instructions for the gauge
  21. Procedure p. 25
    Calibration and adjustment procedures
  22. Procedure p. 26
    Configuration of switching functions
  23. Procedure p. 27
    Procedure for setting switching setpoints
  24. Procedure p. 28
    Instructions for removing the gauge
  25. Procedure p. 29
    Maintenance and repair guidance
  26. Ordering p. 30
    List of accessories and spare parts
  27. pp. 31-34
    Vacuum Products Division contact details
  28. Overview p. 35
    Summary of the manual contents
  29. p. 16 table, 28 rows
  30. p. 30 table, 12 rows
  31. p. 17 table, 30 rows
  32. p. 14 table, 16 rows
  33. p. 5 table, 11 rows
  34. p. 19 table, 1 rows
  35. p. 20 table, 5 rows
  36. p. 33 table, 11 rows

Table 1 Technical Specifications

PVG-500 and PVG-502 SpecificationsValue
Measurement principlethermal conductance according to Pirani
Measurement range (air, O2, CO, N2)5×10-4 … 1000 mbar
Accuracy (N2)
1×10-3 … 100 mbar±15% of reading
5×10-4 … 1×10-3 mbar±50% of reading
100 … 1000 mbar±50% of reading
Resolution1% of reading
Repeatability
1×10-3 … 100 mbar2% of reading
Output signal (measurement signal)
Voltage rangeVDC 0 … +10.3
Measurement rangeVDC +1.9 … +10.0
Voltage vs. pressureLogarithmic 1.286 V/decade
Error signal
Filament ruptureV 0 … +0.5
V +0.1
Output impedanceΩ 2×4.7
Minimum loaded impedancekΩ 10, short-circuit proof
Response timems 80
Gauge identification27.0 kΩ, referenced to supply common (voltage at pin 4 ≤5 V)
Adjustmentone tactile switch for ATM and HV adjustment
Switching functions
Threshold value indication and setting
Setting rangeSP1, SP2 one tactile switch at measurement value output. Press briefly for threshold indication. Keep pressing or press repeatedly for threshold setting. 2×10-3 … 500 mbar
Hysteresis10% above lower threshold
Relay contact
closed30 V, 0.5 ADC, floating at low pressure (LED is lit)
openat high pressure, error, missing supply

Table 3 Accessories and Spare Parts for PVG-500 and PVG-502

Sensor for gaugeOrdering number
PVG500KF16
PVG500KF16S
PVG500CF16
PVG500CF16S
PVG500KF16RS
PVG500CF16RS
PVG502KF16S
PVG502CF16S
PVG502KF16RS
VG502CF16RS

Table 2 Technical Specifications (continued)

PVG-500 and PVG-502 SpecificationsValue
Supply voltage
At gaugeVDC +14 … +30
RippleVpp ≤1
Current consumptionmA <500 (max. starting current)
Power consumptionW ≤1
Fuse required 1)AT (slow) 1
Electrical connection
Sensor cableFCC 68 / RJ45 appliance connector, 8-pin, male
Cable length8-pin plus shielding ≤100 m (8×0.14 mm2)
Grounding concept
Vacuum connection to signal commonconnected via 1 MΩ (voltage difference <15 V)
Supply common to signal commonconducted separately, for differential measurement → "Power Connection"
Materials exposed to vacuum
FilamentPVG-500 W
PVG-502 Ni
DIN 1.4301, DIN 1.4305, DIN 1.4435, glass, Ni, NiFe
Internal volumecm3 ≈1.5
Admissible pressurebar (abs.) 10, limited to inert gases
Admissible temperatures
Operation°C +5 … +60
Vacuum connection°C 80
Filament°C 110
Storage°C –20 … +65
Relative humidity% ≤80 at temperatures up to ≤+31 °C, decreasing to 50 at +40 °C
Useindoors only, altitude up to 2000 m NN
Mounting orientationany
Degree of protectionIP40
Weight
DN 16 ISO-KFg 80
DN 16 CF-Rg 100

1) Agilent controllers fulfill these requirements.

Technical Information

РазделСтраница
Product Description15
Technical Specifications16
PVG-500 and PVG-502 Outline18
Measurement Signal vs. Pressure19
Gas Type Dependence20
Installation21
Vacuum Connection21
Power Connection23
Operation24
Gas Type Dependence24
Adjusting the Gauge25
Switching functions26
Adjusting the Setpoints27
Deinstallation28
Maintenance, Repair29
Accessories and Spare Parts30

Instructions for Use

РазделСтраница
About this manual6
Validity6
Definitions and terminology7
Definition of Caution, Warning and Note7
Warning Symbols8
Safety9
Proper use9
Improper use10
Safety guideline for PVG-500 and PVG-50211
Disposal12
Service13

Measurement Signal vs. Pressure

U [V]p [mbar]cU [V]p [micron]cU [V]p [µbar]cU [V]p [Pa]c
6.1432.4482.2873.5726.3047.4292.448

valid in the range 5×10-4 mbar < p < 1000 mbar; 3.75×10-4 Torr < p < 750 Torr; 5×10-2 Pa < p < 1×105 Pa. where p pressure, U measurement signal, c constant (depending on pressure unit)

Calibration factors for the pressure range below 1 mbar

Gas typeCalibration factor CGas typeCalibration factor C
He0.8H20.5
Ne1.4air, O2, CO, N21.0
Ar1.7CO20.9
Kr2.4water vapor0.5
Xe3.0freon 120.7

peff = C × pressure reading

Health and safety — The product has been exposed to the following substances: (by selecting ‘YES’ you MUST complete the table to the right)

Substances (please refer to MSDS forms)Trade nameChemical nameChemical SymbolCAS Number
Toxic
Harmful
Corrosive
Reactive
Flammable
Explosive (*)
Radioactive (*)
Biological (*)
Oxidizing
Sensitizer
Other dangerous substances

* Agilent will not accept delivery of any product that is exposed to radioactive, biological, explosive substances or dioxins, PCB’s without written evidence of decontamination.

Engineering Facts

14 facts · every claim linked to a page
Selection

The gauge covers a vacuum measurement range of 5×10-4 to 1000 mbar, so it suits rough-to-medium vacuum applications only.

Defines the usable pressure envelope when selecting the gauge for a chamber.

Source
“The Pirani Standard Gauges PVG-500 and PVG-502 have been designed for vacuum measurement of gases in the pressure range of 5×10-4 … 1000 mbar.”
Safety

The gauge must not be used to measure flammable or combustible gas mixtures with oxidants inside the explosion range.

Rules out the gauge for hazardous gas service during selection.

Source
“They must not be used for measuring flammable or combustible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within the explosion range.”
Selection

The gauge is available with tungsten or nickel filament and with or without switching functions, so part number must be chosen accordingly.

Part-number variants determine filament material and relay/setpoint capability.

Source
“W Filament: PVG500KF16 (DN 16 ISO-KF, w/o switching functions)”
Compliance

The gauge may only be powered from a grounded SELV supply and the connection must be fused.

Electrical safety and fuse sizing must be verified before wiring the gauge.

Source
“The gauge may only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extralow voltage (SELV). The connection to the gauge has to be fused 1).”
Installation

The gauge can be mounted in any orientation, but horizontal-to-upright is preferred to keep condensates and particles out.

Mounting orientation affects contamination and long-term reading stability.

Source
“The gauge may be mounted in any orientation. To keep condensates and particles from getting into the measuring chamber preferably choose a horizontal to upright position”
Installation

For KF connections a conductive metallic clamping ring is required to achieve protective grounding.

Without it the protective ground requirement of EN 61010 is not met.

Source
“For gauges with KF connections, use a conductive metallic clamping ring.”
Installation

KF flange connections with elastomer seals cannot withstand overpressure above 2.5 bar, so O-rings with outer centering ring must be used.

Determines seal/clamp selection for systems that may see overpressure.

Source
“KF flange connections with elastomer seals (e.g. O-rings) cannot withstand such pressures. Process media can thus leak and possibly damage your health.”
Operation

The measurement signal is available between pins 3 and 5 and a stabilization period of at least 10 minutes is required.

Wiring and warm-up time must be planned for reliable readings.

Source
“Allow a stabilization period of at least 10 minutes. It is advisable to operate the gauge continuously, irrespective of the pressure.”
Performance

Readings are gas-dependent and must be corrected with calibration factors for gases other than air, O2, CO and N2.

Process gas composition dictates whether a correction factor must be applied.

Source
“The measurement value is gas dependent. The pressure reading applies to dry air, O2, CO and N2.”
Maintenance

The gauge is factory calibrated but zero drift can occur, so periodic zero (ATM/HV) adjustment is required.

Plans periodic recalibration to maintain accuracy over service life.

Source
“The gauge is factory calibrated. Due to long time operation or contamination, a zero drift could occur. Periodically check the zero and adjust it if necessary.”
Operation

Pressing an SP button suppresses the measurement signal and outputs the threshold value instead, which can disturb process control.

Operators must avoid pressing SP buttons while the signal drives a process.

Source
“If processes are controlled via the signal output, keep in mind that by pressing a button <SP> the measururement signal is suppressed and that the corresponding threshold value is output instead.”
Maintenance

Gauge failures from contamination or wear and expendable parts such as the filament are not covered by warranty.

Affects lifecycle cost and spare-part planning at selection.

Source
“Gauge failures due to contamination or wear and tear as well as expendable parts (e.g. filament) are not covered by the warranty.”
Safety

The Pirani filament reaches 120 °C, so possible reactions of process media with that heat must be considered.

Hot-filament temperature limits use with thermally sensitive or reactive media.

Source
“Consider possible reactions (e.g. explosion) of the process media due to the heat generated by the product (Pirani filament 120° C).”
Installation

The gauge must be electrically connected to the grounded vacuum chamber to satisfy protective connection requirements.

Grounding is mandatory for fault safety and must be designed into the installation.

Source
“The gauge must be electrically connected to the grounded vacuum chamber.”