Specific safety guidelines for PVG-500 and PVG-502
p. 12
Instructions for proper disposal of the product
Servicesp. 13
Service and support contact information
Specp. 14
Technical data and operating parameters
Overviewp. 15
Description of the gauge and its components
Specpp. 16-17
Detailed technical specifications and limits
Specp. 18
Dimensional drawings for PVG-500 and PVG-502
Specpp. 19-20
Measurement behavior for different gases
Procedurepp. 21-22
Installation instructions and notes
Procedurep. 23
Electrical power connection procedure
Procedurep. 24
Operating instructions for the gauge
Procedurep. 25
Calibration and adjustment procedures
Procedurep. 26
Configuration of switching functions
Procedurep. 27
Procedure for setting switching setpoints
Procedurep. 28
Instructions for removing the gauge
Procedurep. 29
Maintenance and repair guidance
Orderingp. 30
List of accessories and spare parts
pp. 31-34
Vacuum Products Division contact details
Overviewp. 35
Summary of the manual contents
p. 16table, 28 rows
p. 30table, 12 rows
p. 17table, 30 rows
p. 14table, 16 rows
p. 5table, 11 rows
p. 19table, 1 rows
p. 20table, 5 rows
p. 33table, 11 rows
Table 1 Technical Specifications
PVG-500 and PVG-502 Specifications
Value
Measurement principle
thermal conductance according to Pirani
Measurement range (air, O2, CO, N2)
5×10-4 … 1000 mbar
Accuracy (N2)
1×10-3 … 100 mbar
±15% of reading
5×10-4 … 1×10-3 mbar
±50% of reading
100 … 1000 mbar
±50% of reading
Resolution
1% of reading
Repeatability
1×10-3 … 100 mbar
2% of reading
Output signal (measurement signal)
Voltage range
VDC 0 … +10.3
Measurement range
VDC +1.9 … +10.0
Voltage vs. pressure
Logarithmic 1.286 V/decade
Error signal
Filament rupture
V 0 … +0.5
V +0.1
Output impedance
Ω 2×4.7
Minimum loaded impedance
kΩ 10, short-circuit proof
Response time
ms 80
Gauge identification
27.0 kΩ, referenced to supply common (voltage at pin 4 ≤5 V)
Adjustment
one tactile switch for ATM and HV adjustment
Switching functions
Threshold value indication and setting
Setting range
SP1, SP2 one tactile switch at measurement value output. Press briefly for threshold indication. Keep pressing or press repeatedly for threshold setting. 2×10-3 … 500 mbar
Hysteresis
10% above lower threshold
Relay contact
closed
30 V, 0.5 ADC, floating at low pressure (LED is lit)
open
at high pressure, error, missing supply
Table 3 Accessories and Spare Parts for PVG-500 and PVG-502
Sensor for gauge
Ordering number
PVG500KF16
PVG500KF16S
PVG500CF16
PVG500CF16S
PVG500KF16RS
PVG500CF16RS
PVG502KF16S
PVG502CF16S
PVG502KF16RS
VG502CF16RS
Table 2 Technical Specifications (continued)
PVG-500 and PVG-502 Specifications
Value
Supply voltage
At gauge
VDC +14 … +30
Ripple
Vpp ≤1
Current consumption
mA <500 (max. starting current)
Power consumption
W ≤1
Fuse required 1)
AT (slow) 1
Electrical connection
Sensor cable
FCC 68 / RJ45 appliance connector, 8-pin, male
Cable length
8-pin plus shielding ≤100 m (8×0.14 mm2)
Grounding concept
Vacuum connection to signal common
connected via 1 MΩ (voltage difference <15 V)
Supply common to signal common
conducted separately, for differential measurement → "Power Connection"
Materials exposed to vacuum
Filament
PVG-500 W
PVG-502 Ni
DIN 1.4301, DIN 1.4305, DIN 1.4435, glass, Ni, NiFe
Internal volume
cm3 ≈1.5
Admissible pressure
bar (abs.) 10, limited to inert gases
Admissible temperatures
Operation
°C +5 … +60
Vacuum connection
°C 80
Filament
°C 110
Storage
°C –20 … +65
Relative humidity
% ≤80 at temperatures up to ≤+31 °C, decreasing to 50 at +40 °C
Use
indoors only, altitude up to 2000 m NN
Mounting orientation
any
Degree of protection
IP40
Weight
DN 16 ISO-KF
g 80
DN 16 CF-R
g 100
1) Agilent controllers fulfill these requirements.
Technical Information
Раздел
Страница
Product Description
15
Technical Specifications
16
PVG-500 and PVG-502 Outline
18
Measurement Signal vs. Pressure
19
Gas Type Dependence
20
Installation
21
Vacuum Connection
21
Power Connection
23
Operation
24
Gas Type Dependence
24
Adjusting the Gauge
25
Switching functions
26
Adjusting the Setpoints
27
Deinstallation
28
Maintenance, Repair
29
Accessories and Spare Parts
30
Instructions for Use
Раздел
Страница
About this manual
6
Validity
6
Definitions and terminology
7
Definition of Caution, Warning and Note
7
Warning Symbols
8
Safety
9
Proper use
9
Improper use
10
Safety guideline for PVG-500 and PVG-502
11
Disposal
12
Service
13
Measurement Signal vs. Pressure
U [V]
p [mbar]
c
U [V]
p [micron]
c
U [V]
p [µbar]
c
U [V]
p [Pa]
c
6.143
2.448
2.287
3.572
6.304
7.429
2.448
valid in the range 5×10-4 mbar < p < 1000 mbar; 3.75×10-4 Torr < p < 750 Torr; 5×10-2 Pa < p < 1×105 Pa. where p pressure, U measurement signal, c constant (depending on pressure unit)
Calibration factors for the pressure range below 1 mbar
Gas type
Calibration factor C
Gas type
Calibration factor C
He
0.8
H2
0.5
Ne
1.4
air, O2, CO, N2
1.0
Ar
1.7
CO2
0.9
Kr
2.4
water vapor
0.5
Xe
3.0
freon 12
0.7
peff = C × pressure reading
Health and safety — The product has been exposed to the following substances: (by selecting ‘YES’ you MUST complete the table to the right)
Substances (please refer to MSDS forms)
Trade name
Chemical name
Chemical Symbol
CAS Number
Toxic
Harmful
Corrosive
Reactive
Flammable
Explosive (*)
Radioactive (*)
Biological (*)
Oxidizing
Sensitizer
Other dangerous substances
* Agilent will not accept delivery of any product that is exposed to radioactive, biological, explosive substances or dioxins, PCB’s without written evidence of decontamination.
Engineering Facts
14 facts · every claim linked to a page
Selection
The gauge covers a vacuum measurement range of 5×10-4 to 1000 mbar, so it suits rough-to-medium vacuum applications only.
Defines the usable pressure envelope when selecting the gauge for a chamber.
Source
“The Pirani Standard Gauges PVG-500 and PVG-502 have been designed for vacuum measurement of gases in the pressure range of 5×10-4 … 1000 mbar.”
Safety
The gauge must not be used to measure flammable or combustible gas mixtures with oxidants inside the explosion range.
Rules out the gauge for hazardous gas service during selection.
Source
“They must not be used for measuring flammable or combustible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within the explosion range.”
Selection
The gauge is available with tungsten or nickel filament and with or without switching functions, so part number must be chosen accordingly.
Part-number variants determine filament material and relay/setpoint capability.
The gauge may only be powered from a grounded SELV supply and the connection must be fused.
Electrical safety and fuse sizing must be verified before wiring the gauge.
Source
“The gauge may only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extralow voltage (SELV). The connection to the gauge has to be fused 1).”
Installation
The gauge can be mounted in any orientation, but horizontal-to-upright is preferred to keep condensates and particles out.
Mounting orientation affects contamination and long-term reading stability.
Source
“The gauge may be mounted in any orientation. To keep condensates and particles from getting into the measuring chamber preferably choose a horizontal to upright position”
Installation
For KF connections a conductive metallic clamping ring is required to achieve protective grounding.
Without it the protective ground requirement of EN 61010 is not met.
Source
“For gauges with KF connections, use a conductive metallic clamping ring.”
Installation
KF flange connections with elastomer seals cannot withstand overpressure above 2.5 bar, so O-rings with outer centering ring must be used.
Determines seal/clamp selection for systems that may see overpressure.
Source
“KF flange connections with elastomer seals (e.g. O-rings) cannot withstand such pressures. Process media can thus leak and possibly damage your health.”
Operation
The measurement signal is available between pins 3 and 5 and a stabilization period of at least 10 minutes is required.
Wiring and warm-up time must be planned for reliable readings.
Source
“Allow a stabilization period of at least 10 minutes. It is advisable to operate the gauge continuously, irrespective of the pressure.”
Performance
Readings are gas-dependent and must be corrected with calibration factors for gases other than air, O2, CO and N2.
Process gas composition dictates whether a correction factor must be applied.
Source
“The measurement value is gas dependent. The pressure reading applies to dry air, O2, CO and N2.”
Maintenance
The gauge is factory calibrated but zero drift can occur, so periodic zero (ATM/HV) adjustment is required.
Plans periodic recalibration to maintain accuracy over service life.
Source
“The gauge is factory calibrated. Due to long time operation or contamination, a zero drift could occur. Periodically check the zero and adjust it if necessary.”
Operation
Pressing an SP button suppresses the measurement signal and outputs the threshold value instead, which can disturb process control.
Operators must avoid pressing SP buttons while the signal drives a process.
Source
“If processes are controlled via the signal output, keep in mind that by pressing a button <SP> the measururement signal is suppressed and that the corresponding threshold value is output instead.”
Maintenance
Gauge failures from contamination or wear and expendable parts such as the filament are not covered by warranty.
Affects lifecycle cost and spare-part planning at selection.
Source
“Gauge failures due to contamination or wear and tear as well as expendable parts (e.g. filament) are not covered by the warranty.”
Safety
The Pirani filament reaches 120 °C, so possible reactions of process media with that heat must be considered.
Hot-filament temperature limits use with thermally sensitive or reactive media.
Source
“Consider possible reactions (e.g. explosion) of the process media due to the heat generated by the product (Pirani filament 120° C).”
Installation
The gauge must be electrically connected to the grounded vacuum chamber to satisfy protective connection requirements.
Grounding is mandatory for fault safety and must be designed into the installation.
Source
“The gauge must be electrically connected to the grounded vacuum chamber.”