Instruction Manual - CFT-EX Series

Technical Reference Summary

Covered Models & Document Scope

This manual covers the Shimadzu FLOWTESTER CFT-EX series constant-pressure extrusion capillary rheometer, specifically models CFT-500EX (part no. 344-04260-58) and CFT-100EX (part no. 344-04261-58). Both models share the same appearance and operating software; the EMC section explicitly applies only to these two model numbers.

Section-by-Section Technical Breakdown

pp. 30
Overview and Test Modes

The CFT-EX is a constant-pressure extrusion capillary rheometer that measures flow rate of fluid materials under controlled temperature and pressure and calculates viscosity. It supports a constant-temperature mode and a constant-heating-rate mode, the latter allowing determination of softening temperature and flow-beginning temperature.

  • Measures flow rate of melted materials passing through a die orifice and calculates viscosity.
  • Two test modes: constant temperature mode and constant heating rate mode.
  • Constant heating rate mode measures softening temperature and flow beginning temperature.
  • Applicable to synthetic resin, ceramics and food; approximates actual molding pressure conditions.
pp. 40
Operation Unit Controls and Indicators

The operation unit contains four buttons (START, STOP, AIR, HEAT) and two LED indicators (ALARM and POWER). The POWER light blinks during start-up and lights green when running; the ALARM light is orange and indicates error status.

  • [START] starts a test; [STOP] stops pre-heating or test.
  • [AIR] bleeds air; [HEAT] turns heater ON/OFF and should only be set ON when software displays current temperature.
  • POWER light: blinking = system in start-up; lighting = system running; color green.
  • ALARM light color is orange; refer to Chapter 10.1 for details.
  • If ALARM1, ALARM2 and POWER lights blink simultaneously, the CPU board has a problem; restart once, and if blinking continues contact Shimadzu.
pp. 43-44
Installation Site and Power Requirements

The installation site must meet environmental and electrical conditions including temperature, humidity, vibration, grounding and supply voltage tolerance. The main unit weighs about 160 kg and requires a solid floor with a horizontal tolerance of 1/500.

  • Operating temperature: +10 °C to +35 °C; humidity 60 % or less, no condensation.
  • Vibration acceleration must be 0.005 G or less.
  • Supply voltage fluctuation must be within ±10 %.
  • D class grounding with ground resistance of 100 Ω max.
  • Installation surface must be within 1/500 horizontal degree; main unit weight about 160 kg.
  • Necessary power: single-phase AC 100 V, 700 VA, 50/60 Hz.
pp. 49
Supplying Air Pressure and Turning Power ON

Before operation, open the stop valve and verify the manometer reads 0.4 ± 0.02 MPa; adjust if necessary per section 10.3. Turn the power switch on the right side of the testing machine; the POWER light flashes then lights, and the ALARM light should be checked.

  • Open stop valve counterclockwise; manometer must read 0.4 ± 0.02 MPa.
  • Before turning OFF power, close the stop valve clockwise.
  • POWER light begins to flash and lights in a few seconds after power ON.
  • Check ALARM light after power ON.
pp. 50-51
System Start-up, Connection and Shutdown

Start the system via the [On-Line] button in the program manager. The system loads initial test conditions and checks communication with the tester; if the tester is OFF or communication fails, an error message is displayed and the on-line main window is not opened. Shutdown is via [File]-[Exit] or the window close button, with prompts to save unsaved test data.

  • Click [On-Line] in the CFT-500EX/100DEX Program Manager window.
  • If tester power is OFF or communication is disabled, check power and USB connection and retry.
  • Shutdown: [File]-[Exit] or window [CLOSE] button.
  • If quitting during a test, a confirmation message appears; unsaved results prompt a save dialog.
pp. 193
Troubleshooting

Table 10.2 lists problems, causes and solutions for heating failures, stroke display failures, and non-smooth flow curves. Many causes require contacting a Shimadzu representative; others can be resolved by adjusting air pressure, removing the die stopper, cleaning the die orifice, or resetting test conditions.

  • Heating unit not heated (temperature near room temperature): thermal fuse melted, band heater failed, A/D board chip failed, or control unit overheating prevention circuit activated.
  • Heating unit not heated (temperature not near room temperature): resistance bulb failed, connector at rear of control unit loosened, PID constant improper, heater improperly installed, or resistance bulb incompletely inserted.
  • Stroke display value does not change: potentiometer failed/caught, A/D board part failed, cylinder failed, air pressure zero, die stopper attached, die orifice clogged, or test force/temperature too low.
  • Flow curve not smooth: scratched piston/cylinder, potentiometer not smooth, cylinder speed not adjusted, material adhered, or improper air pressure.
pp. 27, 193
Adjusting Air Pressure

Section 10.3 provides the procedure for adjusting air pressure when the manometer does not read 0.4 ± 0.02 MPa. This is referenced from troubleshooting for zero or improper air pressure.

  • Referenced from troubleshooting for air pressure zero and improper air pressure.
  • Target manometer reading is 0.4 ± 0.02 MPa as stated in section 5.1.1.
pp. 22, 27
Consumables and Maintenance Parts

Chapter 11 lists consumables and maintenance parts. Replacement cycles described for periodic replacement parts are rough estimates and may require earlier replacement depending on usage environment and frequency.

  • Chapter 11.1 covers consumables; Chapter 11.2 covers maintenance parts.
  • Replacement cycles are rough estimates; actual replacement may be earlier depending on environment and usage frequency.
  • Periodic inspection/calibration should be requested to a Shimadzu representative.
pp. 27
Technical Data and Specifications

Chapter 12 covers principle of operation, test modes, viscosity correction, shear sensitivity N, data file extension codes, and specifications. These sections provide the theoretical basis and data formats for measurement and reanalysis.

  • 12.1 Principle; 12.2 Test modes; 12.3 Viscosity correction; 12.4 Shear sensitivity N.
  • 12.5 Data file extension code list; 12.6 Specifications.
pp. 27
Off-line Software (Data Reanalysis)

Chapter 7 describes off-line software for reanalyzing saved test results, including recalculation in constant temperature and constant heating rate modes, viscosity/shear stress vs shear rate curves, overlays, zoom, viscosity correction, and ASCII export.

  • Constant temperature mode recalculation: stroke-time.
  • Constant heating-rate mode recalculation: stroke-temperature/time.
  • Viscosity/shear stress - shear rate curve, viscosity/shear rate - temperature curve, viscosity/shear rate - time curve.
  • Overlay and zoom functions; viscosity correction; ASCII save of analytical display result.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Operating temperature range is +10 °C to +35 °C with minimal daily variation. p.43 ↗ Install in a temperature-controlled room; excursions outside this range may affect measurement stability. CFT-500EX and CFT-100EX
Humidity must be 60 % or less with no condensation. p.43 ↗ Use dehumidification or air conditioning if ambient humidity exceeds 60 %; condensation can damage electronics. CFT-500EX and CFT-100EX
Vibration acceleration at the installation site must be 0.005 G or less. p.43 ↗ Avoid locations near heavy machinery or traffic; vibration can disturb the stroke measurement. CFT-500EX and CFT-100EX
Supply voltage fluctuation must be within ±10 %. p.43 ↗ Verify power quality; unstable supply may prevent rated performance or cause faults. CFT-500EX and CFT-100EX
Required power is single-phase AC 100 V, 700 VA, 50/60 Hz. p.44 ↗ Provide a dedicated circuit with sufficient capacity; use the power cord specified or provided by Shimadzu. CFT-500EX and CFT-100EX
Grounding must be D class with ground resistance of 100 Ω max. p.43 ↗ Proper grounding is required to prevent electric shock and ensure stable operation; ground the PC and optional units with equal potential grounds. CFT-500EX and CFT-100EX
Main unit weight is about 160 kg; installation surface must be within 1/500 horizontal degree. p.43 ↗ Mount on a solid floor; special foundation work is not required but the floor must support the weight and be level. CFT-500EX and CFT-100EX
Air pressure must be set to 0.4 ± 0.02 MPa as read on the manometer. p.49 ↗ Check and adjust air pressure before each test; incorrect pressure affects cylinder speed and flow curve smoothness. CFT-500EX and CFT-100EX
Before turning OFF the power switch, close the stop valve clockwise. p.49 ↗ Prevents air pressure loss and protects the pneumatic system during shutdown. CFT-500EX and CFT-100EX
The [HEAT] button should only be set to ON when the software is displaying the current temperature. p.40 ↗ Prevents heater activation without temperature monitoring, reducing risk of overheating. CFT-500EX and CFT-100EX
If ALARM1, ALARM2 and POWER lights blink simultaneously, the CPU board has a problem. p.40 ↗ Restart the system once; if blinking continues, contact Shimadzu. CFT-500EX and CFT-100EX
EMC compliance: EN61326:1997 + A1:1998 + A2:2001 + A3:2003, class A emissions, minimum immunity requirements. p.25 ↗ Class A product not designed for residential environments; maintain distance from sensitive devices and use a separate power source from strong EMI noise sources. CFT-500EX (344-04260-58) and CFT-100EX (344-04261-58)
Immunity test levels: ESD air/contact 4 kV; radiated RF 3 V/m; EFT 1 kV to AC power line and ground; surge 0.5 kV line-to-line, 0.5/1 kV line-to-ground; conducted RF 3 V; voltage dip 100 % for 1 cycle. p.25 ↗ Interference stronger than these levels may cause malfunction; locate away from strong EMI sources. CFT-500EX and CFT-100EX
Replacement parts are available for seven years after product discontinuation. p.21 ↗ Plan long-term maintenance and spare parts procurement within this window. CFT-500EX and CFT-100EX
Do not store weights inside the circle shown in Figure 4.3 when storing them in the main body. p.46 ↗ Improper weight storage may affect test data if the weight sling drops during a test. CFT-500EX and CFT-100EX
The press joint can bend at 90° and the knurlizer adjusts the load head up/down to set piston contact level depending on sample volume. p.38 ↗ Adjust piston contact level before testing to accommodate different sample volumes. CFT-500EX and CFT-100EX

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
The heating unit is not heated up. (Temperature display remains near room temperature.) p.193 ↗ Thermal fuse is melted. Please contact Shimadzu representative. Table 10.2
The heating unit is not heated up. (Temperature display remains near room temperature.) p.193 ↗ Band heater is failed. Please contact Shimadzu representative. Table 10.2
The heating unit is not heated up. (Temperature display remains near room temperature.) p.193 ↗ A/D board chip is failed. Please contact Shimadzu representative. Table 10.2
The heating unit is not heated up. (Temperature display remains near room temperature.) p.193 ↗ Control unit overheating prevention circuit is activated. Turn power OFF and ON after 1 hour. Table 10.2
The heating unit is not heated up. (Temperature display is not near room temperature.) p.193 ↗ Resistance bulb is failed. Please contact Shimadzu representative. Table 10.2
The heating unit is not heated up. (Temperature display is not near room temperature.) p.193 ↗ Connector at rear of control unit is loosened. Please contact Shimadzu representative. Table 10.2
Temperature control is failed. p.193 ↗ PID constant is not proper. Please contact Shimadzu representative. Table 10.2
Temperature control is failed. p.193 ↗ Heater is improperly installed. Please contact Shimadzu representative. Table 10.2
Temperature control is failed. p.193 ↗ Resistance bulb is incompletely inserted. Please contact Shimadzu representative. Table 10.2
The stroke display value does not change. p.193 ↗ Main unit potentiometer for stroke detection is failed or caught. Please contact Shimadzu representative. Table 10.2
The stroke display value does not change. p.193 ↗ A/D board part is failed. Please contact Shimadzu representative. Table 10.2
The stroke display value does not change. p.193 ↗ Cylinder is failed. Please contact Shimadzu representative. Table 10.2
The stroke display value does not change. p.193 ↗ Air pressure is zero. Refer to 10.3 and adjust the air pressure. Table 10.2
The stroke display value does not change. p.193 ↗ Die stopper is still attached. Remove the die stopper. Table 10.2
The stroke display value does not change. p.193 ↗ Die orifice is clogged. Refer to 5.8.12 and clean the die orifice. Table 10.2
The stroke display value does not change. p.193 ↗ Test force is too small or temperature is too low. Please reset the test conditions. Table 10.2
The flow curve (stroke-time curve) is not smooth. p.193 ↗ Piston or cylinder parts are scratched. Please contact Shimadzu representative. Table 10.2
The flow curve (stroke-time curve) is not smooth. p.193 ↗ Potentiometer for stroke detection does not operate smoothly. Please contact Shimadzu representative. Table 10.2
The flow curve (stroke-time curve) is not smooth. p.193 ↗ Cylinder speed is not adjusted. Please contact Shimadzu representative. Table 10.2
The flow curve (stroke-time curve) is not smooth. p.193 ↗ Material is adhered to piston or cylinder. Refer to 5.8.12 and clean the piston or cylinder. Table 10.2
The flow curve (stroke-time curve) is not smooth. p.193 ↗ Air pressure is improper. Refer to 10.3 and adjust the air pressure. Table 10.2
ALARM1, ALARM2, and POWER lights are blinking at the same time p.40 ↗ The CPU board has a problem. Please re-start the system once. If the blinking continues, please contact to Shimadzu agency. Section 3.4