Induztr
Engineering Digest Document Viewer Related Models

Instruction Manual - Installation

Technical Reference Summary

Covered Models & Document Scope

This document (347-05511C, May 2021) covers the installation of CFT-EX PC software and USB driver for the SHIMADZU FLOWTESTER CFT-500EX/100EX. It applies to Windows 10 (64-bit) and Windows 7 (32-bit and 64-bit) host PCs.

Section-by-Section Technical Breakdown

pp. 5-7
Installation of CFT-EX Software

Describes installing the CFT-EX PC software from the Setup CD-ROM. The PC must be started with administrator authority. The procedure covers launching setup.exe, selecting language, choosing an installation folder, and completing the Setup Wizard.

  • Start the PC with administrator authority before installation.
  • Insert the CFT-EX Setup CD-ROM and run [\DISK1\setup.exe] (may display as 'setup').
  • Select the setup language in the Select Setup Language dialog and click [OK].
  • The default installation destination is C:\CFT; sample test data are installed in C:\CFT\Sample.
  • To uninstall, use Control Panel > Uninstall a program and select 'CFT-EX'.
pp. 9
Installation of USB Driver

Explains how to install the USB driver required for PC-to-instrument communication. The correct driver executable depends on the Windows version and system type (32-bit or 64-bit).

  • For Windows 10, run \Driver\Windows10\DPInst64.exe.
  • For Windows 7 32-bit, run \Driver\Windows7\32bit\DPInst.exe.
  • For Windows 7 64-bit, run \Driver\Windows7\64bit\DPInst64.exe.
  • Confirm the system type via Control Panel > System and Security > System.
  • Follow the installation wizard instructions to complete driver installation.
pp. 11
Connection of the USB Cable

The instrument has a USB connector on the rear panel. Connect the PC and the unit using the USB cable.

  • USB connector is located on the back of the unit.
  • Connect the PC and the unit with the USB cable.
pp. 13-16
Setting for Using Touch Keyboard

Describes enabling and using the Windows touch keyboard and On-Screen keyboard for touch panel operation without a mouse or physical keyboard. Notes feature differences and the need to install the touch panel display driver first.

  • Touch keyboard: large keys for finger touch; numeric keypad and system keys cannot be used.
  • On-Screen keyboard: supports numeric keypad and system keys like a physical keyboard.
  • Enable touch keyboard via right-click taskbar > 'Show touch keyboard button'.
  • Enable On-Screen keyboard via Settings > Ease of Access > Keyboard > 'Use the On-Screen Keyboard'.
  • The touch panel display driver must be installed before touch panel functions can be enabled.
pp. 7
Power Options for Stable Communication

Warns that PC sleep and hard disk turn-off settings can cause loss of communication between the PC and testing machine. Sets these power options to 'Never'.

  • Set [Sleep] and [Turn off hard disks] to [Never] via Control Panel > System and Maintenance > Power Options.
  • Failure to do so may cause connection loss during communication.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
The PC must start by the administrator authority for CFT-EX software installation and uninstallation. p.5 ↗ Ensure administrator rights are available before starting installation or removal. CFT-500EX/100EX with CFT-EX software
Setup executable is located at [\DISK1\setup.exe] on the CFT-EX Setup CD-ROM. p.5 ↗ Launch this file to begin installation; it may be displayed as 'setup'. CFT-EX software
Default installation destination is C:\CFT; sample test data are installed in C:\CFT\Sample. p.7 ↗ Sample data are for explanation only; plan disk space and folder permissions accordingly. CFT-EX software
For Windows 10, the USB driver executable is \Driver\Windows10\DPInst64.exe. p.9 ↗ Use this driver on Windows 10 systems. CFT-500EX/100EX USB driver
For Windows 7 32-bit, the USB driver executable is \Driver\Windows7\32bit\DPInst.exe. p.9 ↗ Confirm 32-bit system type before running this driver. CFT-500EX/100EX USB driver
For Windows 7 64-bit, the USB driver executable is \Driver\Windows7\64bit\DPInst64.exe. p.9 ↗ Confirm 64-bit system type before running this driver. CFT-500EX/100EX USB driver
System type (32-bit or 64-bit) is checked via Control Panel > System and Security > System. p.9 ↗ Verify OS architecture before selecting the USB driver folder. CFT-500EX/100EX USB driver
USB connector is located on the back of the unit. p.11 ↗ Route the USB cable to the rear panel for PC connection. CFT-500EX/100EX
Set [Sleep] and [Turn off hard disks] to [Never] in Power Options. p.7 ↗ Prevents loss of communication between PC and testing machine during idle periods. CFT-500EX/100EX with PC
Touch keyboard cannot use the numeric keypad and system keys (non-character keys). p.13 ↗ Use On-Screen keyboard when numeric keypad or system keys are required. Touch panel display operation
On-Screen keyboard supports keys with the same function as a physical keyboard, including numeric keypad and system keys. p.13 ↗ Select On-Screen keyboard for full key functionality on touch panel displays. Touch panel display operation
The touch panel display driver must be installed before touch panel functions can be enabled. p.16 ↗ Install the display driver per the display's manual prior to configuring touch keyboard. Touch panel display operation
Once On-Screen Keyboard is closed, it must be re-enabled from the Settings window. p.16 ↗ Minimize rather than close the On-Screen keyboard to avoid repeated setup. Touch panel display operation

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
connection may be lost during communication between the PC and testing machine p.7 ↗ Depending on the [Sleep] and [Turn off hard disks] settings on the PC Select [Start] - [Control Panel] - [System and Maintenance] - [Power Options]. Set the [Sleep] and [Turn off hard disks] settings to [Never]. CAUTION on p.7

Related models