Technical Reference Summary

Covered Models & Document Scope

This manual covers the Shimadzu SERVOPULSER MMT-V Series micro-load testing system, including models MMT-11NV-2 (part no. 348-20800-01), MMT-101NV-10 (348-20801-00), MMT-250NV-10 (348-20802-00), and MMT-500NV-10 (348-20803-00). It describes the loading frame, electromagnetic actuator, power amplifier, load cells, controller 4830, installation, operation, maintenance, and residual risks.

Section-by-Section Technical Breakdown

pp. 3-4
System Performance and Specifications

This section provides the main performance specifications for the four MMT-V models, including maximum force, stroke, control waveforms, repetition frequency, allowable side force, sensor amplifier channels, indication accuracy, weight, and power supply requirements. It also includes notes on force indication accuracy for the MMT-11NV-2 and power cable limitations.

  • Max. force: ±10 N (MMT-11NV-2), ±100 N (MMT-101NV-10), ±250 N (MMT-250NV-10), ±500 N (MMT-500NV-10).
  • Max. stroke: ±2 mm for MMT-11NV-2; ±10 mm for the other models.
  • Control waveforms: sine, triangle, rectangle, haver-sine, ramp, and external wave.
  • Repetition frequency: max. 60 Hz for MMT-11NV-2; max. 100 Hz for MMT-101NV-10, MMT-250NV-10, and MMT-500NV-10.
  • Allowable side force: 0.5 N, 5 N, 12 N, and 25 N for the respective models.
  • Force indication accuracy: within ±0.5 % of indicated value or 0.02 % of max. force (MMT-11NV-2: within ±1.0 % of indicated value or 0.02 % of max. force).
  • Stroke indication accuracy: within ±1.0 % of indicated value or 0.1 % of max. stroke.
  • Power supply: single phase 100/230 V AC, 50/60 Hz; 500 VA for MMT-11NV-2, 1000 VA for other models.
pp. 5-6
Individual Unit Specifications

This section details the specifications of the loading frame, actuator, power amplifier, and load cell. It includes part numbers, dimensions, weights, structural details, and actuator performance data.

  • Loading frame: twin-pillar arch type, hand wheel crosshead UP/DOWN, friction clamp by bolt tightening, pillar-to-pillar distance 360 mm.
  • Flange spacing: 60–460 mm for 11N/101N/250N models; 0–350 mm for 500N model.
  • Frame dimensions (W:D:H): approx. 510 mm x 415 mm x 1150 mm; weight approx. 48 kg.
  • Actuator part numbers: 339-86055-01 (MMT-11NV-2), 339-86056-01 (MMT-101NV-10), 339-85922-02 (MMT-250NV-10), 339-86415-01 (MMT-500NV-10).
  • Actuator maximum stroke: ±2 mm for MMT-11NV-2; ±10 mm for others.
  • Power amplifier part no. 346-98456-42: maximum output ±10 A, external control start/stop/alarm reset/emergency stop, alarm detection over-current/over-heating/control errors.
  • Power amplifier dimensions (W:D:H): approx. 350 mm x 465 mm x 135 mm; weight approx. 15 kg.
pp. 11
Installation Requirements

This section specifies the environmental and power supply conditions required for installation, including temperature, humidity, vibration, level, and grounding. It also notes that installation is performed by Shimadzu and that stand/fall prevention fixtures are available as an option.

  • Indoor use only; temperature 10 to 35 °C with variation during test within ±2 °C.
  • Humidity 10 to 75 % (no condensation).
  • Power supply voltage fluctuation ±10 % or less; no high-capacity noise-producing equipment on the same line.
  • Vibration amplitude less than 5 μm; level 3 mm/1000 mm or less.
  • Grounding: Class D grounding recommended (100 Ω or less).
  • Use a regulated power supply if power fluctuation cannot be avoided.
pp. 12
Loading Jigs

This section describes the mounting of test jigs and load cells, including thread sizes, tightening torques, and precautions to avoid damaging the load cell or actuator. It specifies allowable tightening torques for load cell mounting.

  • Standard attaching part is internal thread; standard adapters convert thread size to M6 or M5.
  • Allowable tightening torque for load cell: 130 N·cm for MMT-11N-2, MMT-101NV-10, MMT-250NV-10; 300 N·cm for MMT-500NV-10.
  • Use a spanner on the piston to receive reaction force when attaching jigs to the actuator; do not rely on the piston rotation stopper.
  • Be careful not to apply overload to the load cell when attaching jigs.
pp. 14
Wiring up the System

This section covers the wiring connections between the controller, power amplifier, load cell, stroke sensor, and actuator. It emphasizes checking cable connections before power-on and never plugging or unplugging cables with power on.

  • Check cable connections before turning ON the power.
  • Never plug or unplug cables with the power turned on.
  • Improper sensor cable connections can cause abrupt piston movement.
  • Be careful not to let sensor and control cables receive tension due to piston movement.
pp. 15-25
Operating the MMT-V Series

This section describes the operating procedure from confirmation before use through starting the control unit, loading the specimen, starting and ending the test, unloading, and shutting down the power amplifier and system. It includes safety checks and precautions for each step.

  • Before starting the test, check that the specimen is correctly mounted, no one is near the test system, hands are away, and no mechanical stopper obstructs piston stroke.
  • Set alarm (limit conditions) for test force (TD1), stroke (TD2), etc.
  • Adjust test force to zero before installing the test piece; do not adjust zero after mounting the specimen.
  • When using Controller 4830, set control mode to piston displacement mode to enable jog dial for mounting specimen.
  • Before switching off power, ensure power amplifier is set to STOP.
  • If power amplifier shuts down (STOP), actuator will fall under its own weight.
pp. 26-29
Maintenance and Inspection

This section covers daily maintenance, force calibration, and maintenance parts for the 4830 Controller and Power Amplifier. It emphasizes periodic inspection and calibration by Shimadzu representatives.

  • Daily maintenance and inspection are necessary to maintain performance.
  • Periodic inspection/calibration should be requested from Shimadzu representative.
  • Replacement parts are available for seven years after product discontinuation.
  • Use specified fuses and parts; never remove the main cover.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Allowable tightening torque for load cell mounting is 130 N·cm for MMT-11N-2, MMT-101NV-10, and MMT-250NV-10, and 300 N·cm for MMT-500NV-10. p.8 ↗ Use a torque wrench to tighten load cell bolts to these values to avoid damage or insufficient clamping. MMT-11NV-2, MMT-101NV-10, MMT-250NV-10, MMT-500NV-10
Installation environment requires temperature 10 to 35 °C with variation during test within ±2 °C, humidity 10 to 75 % non-condensing, vibration amplitude less than 5 μm, and level 3 mm/1000 mm or less. p.11 ↗ Site the instrument in a stable, temperature-controlled room away from vibration and direct sunlight to maintain micro-load accuracy. whole series
Power supply voltage fluctuation must be ±10 % or less; grounding Class D (100 Ω or less) is recommended. p.11 ↗ Ensure a stable power line and proper grounding to prevent malfunction and ensure safety. whole series
Maximum repetition frequency is 60 Hz for MMT-11NV-2 and 100 Hz for MMT-101NV-10, MMT-250NV-10, and MMT-500NV-10. p.3 ↗ Select the model based on required test frequency; do not exceed the maximum frequency for the model. model-specific
Allowable side force is 0.5 N (MMT-11NV-2), 5 N (MMT-101NV-10), 12 N (MMT-250NV-10), and 25 N (MMT-500NV-10) at neutral piston position. p.3 ↗ Avoid lateral loads exceeding these values; consider external bearings if eccentric loads cannot be avoided. model-specific
Force indication accuracy is within ±0.5 % of indicated value or 0.02 % of max. force (MMT-11NV-2: within ±1.0 % of indicated value or 0.02 % of max. force). p.3 ↗ Account for this accuracy when interpreting test results, especially for small loads. model-specific
Stroke indication accuracy is within ±1.0 % of indicated value or 0.1 % of max. stroke. p.3 ↗ Consider this tolerance for displacement measurements. whole series
Power amplifier maximum output is ±10 A; alarm detection includes over-current, over-heating, and control errors. p.6 ↗ Ensure power amplifier is adequately ventilated and not overloaded to prevent alarms. whole series
Do not use AGC (function of controller 4830) at frequencies lower than 2 Hz. p.1 ↗ Disable AGC for low-frequency tests to avoid unstable control. whole series with 4830 controller
Amplitude performance is specified at the neutral position of piston stroke; amplitude may be reduced if operated away from neutral position. p.1 ↗ Center the piston stroke for tests requiring full amplitude. whole series
Set control gain to the value obtained in a preliminary test; too high gain may cause actuator hunting and damage load cell or specimen. p.12 ↗ Tune control gain carefully and verify stability before running tests. whole series
Before switching off power, ensure power amplifier is set to STOP; switching off in RUN may cause actuator to go out of control. p.12 ↗ Follow proper shutdown sequence to prevent damage or injury. whole series
If power amplifier shuts down (STOP), actuator will fall under its own weight. p.12 ↗ Take precautions to prevent load cell damage when power amplifier stops unexpectedly. whole series
Use a hook wrench to secure the piston when mounting/removing jigs; the piston restraint is not designed for large torques. p.13 ↗ Avoid applying torque to the piston restraint to prevent damage. whole series
Load cell warranty is void if customer drops the load cell, tightens it too much, or damages it by overload. p.8 ↗ Handle load cells with care and follow torque specifications. whole series
Replacement parts are available for seven years after product discontinuation. Plan for long-term maintenance and spare parts procurement. whole series