SFC-40P Instruction Manual - Back Pressure Regulator
Technical Reference Summary
Covered Models & Document Scope
This manual (228-92879) covers the SFC-40P Back Pressure Regulator (BPR) for Shimadzu Nexera UC Prep Series preparative supercritical fluid chromatography. It is intended for use with the LC-40P SF CO2 Solvent Delivery Module or similar. The document describes basic operation, accessories, and options; system-level details are in the Nexera UC Prep Series System Guide (228-92882).
Section-by-Section Technical Breakdown
Lists the parts supplied with the SFC-40P, including the instrument, Safety Guideline, DVD-ROM, drain tubing, fittings, wrenches, joints, drain adapters, and stop plug. Used for unpacking verification.
- SFC-40P main unit (no part number listed)
- Safety Guideline booklet, part 228-92880
- Instruction Manual / System Guide DVD-ROM, part 228-97175-41
- DRAIN TUBE SI (L1000), part 228-25162-03
- FITTING, L-TYPE 2005-10, part 035-61561-12
- WRENCH, 13 x 17, part 086-03047-10 (qty 2)
- STRAIGHT JOINT, part 228-28163
- DRAIN ADAPTER, part 228-42204
- DRAIN OUT, part 228-42205
- DRAIN CTO, part 228-42206
- STOP PLUG, PEEK, part 228-53071
Identifies front-panel features: operation area, front cover, and power button. The operation area is touch-sensitive and displays operation keys when its central part is touched.
- Operation area: touch central part to display operation keys
- Front cover: covers the flow line
- Power button: turns power on/off
Describes internal components behind the front cover, including the BPR pressure sensor, BPR cover, BPR block, thermo-sensor, BPR 2nd heater inlet, makeup inlet, leak sensor, and BPR 1st heater.
- BPR pressure sensor monitors pressure
- BPR cover covers the BPR block and first heater
- BPR block is the pressure control block
- Thermo-sensor measures BPR 1st heater temperature
- BPR 2nd heater inlet
- Makeup inlet connects to makeup pump
- Leak sensor detects liquid leaks
- BPR 1st heater warms tubing from BPR block
Identifies rear-panel connectors and switches: REMOTE, external I/O terminals, COOLER, HEX, PC, power cord connector, and main power switch.
- [REMOTE] connector for system controller
- External input/output terminals for external equipment
- [COOLER] connector for CO2 cooler
- [HEX] connector for heat exchanger
- [PC] connector for service personnel (normally unused)
- Power cord connector
- Main power switch: normally keep on; turn off if not used for a long time
Describes the operation panel display area, status LED, link LED, power button, direct keys, and keypad functions. The panel has arrow-key and numeric-key input modes switched by [Enter].
- Status LED colors: Green = analysis ready; Red = error; Blue = analysis in progress; Yellow = preparation for analysis in progress; Orange = sleep
- Link LED illuminates when controlled by system controller
- Power on: press and hold power button at least 3 seconds
- Power off: hold power button 3 seconds or more, then hold again 1 second on confirmation screen
- Direct keys: [PRESS], [HEAT], [COOL], [HEX]
- [CE] clears input, initializes screen, clears error messages and cancels alarms
- [del] deletes input values
- [Enter] confirms input values and switches input modes
- Arrow keys move cursor or switch screens
- Numeric keys [0]-[9] input values
Provides procedures to set BPR pressure and BPR temperature via the FUNCTION > PARAMETER menu. BPR pressure range is 1.00 to 40.00 MPa (initial 10.00 MPa); BPR temperature range is 0 to 70 °C (initial 50 °C).
- BPR pressure [BPR PRS]: initial 10.00 MPa, settable 1.00 to 40.00 MPa
- BPR temperature [BPR TEMP]: initial 50 °C, settable 0 to 70 °C
- Access: press [→] twice on initial screen, select PARAMETER, press [Enter]
- Use numeric keypad and [Enter] to confirm; [CE] clears entered value
Describes starting operation: verify temperature and pressure increase after starting heaters and CO2 pump, and stopping pressure control with [PRESS].
- Check that displayed temperature starts increasing after starting temperature control
- Check that displayed pressure starts increasing after CO2 pump starts pumping
- Press [PRESS] to stop pressure control
Describes the initial screen, monitor screen, auxiliary functions screen (FUNCTION), and VP functions screen. Monitoring screens display pump head temperatures, CO2 temperature, BPR pressure, and heat exchanger temperature.
- Initial screen: press [↓] for monitoring screens; press [→] to select FUNCTION or VP
- PUMP HEAD: displays right and left pump head temperatures
- CO2: displays CO2 temperature drawn into CO2 pump and cooling block temperature
- BPR PRS: displays BPR pressure; [PZ] and [MT] normally unused
- HEX: displays heat exchanger temperature
- When turning pump head cooling off and on again, wait until cooling block temperature reaches 5 °C or higher to avoid freezing in piping
Lists and describes the FUNCTION setting groups: PARAMETER (analysis parameters), CONTROL (instrument control), SYSTEM (system configuration), and UTILITY (display and buzzer). Includes setting ranges and initial values.
- PARAMETER: BPR PRS (1.00-40.00 MPa, init 10.00), BPR TEMP (0.0-70.0 °C, init 50.0), HEX TEMP (0.0-70.0 °C, init 40.0)
- CONTROL: FORCE PRESS OFF, ZERO ADJUST, BPR HOME, BPR BACK
- SYSTEM: LOCAL MODE (0=Remote, 1=Local, init 0), LINK ADDRESS (1-12, init 4), USE COOL UNIT (0=Use, 1=Not Use, init 0), USE HEX (0=Use, 1=Not Use, init 1)
- SYSTEM: BPR RDY CHECK (0=Off, 1=On, init 1), BPR RDY RNG (0.5-1.0 MPa, init 0.5), BPR WAIT TIME (0-10 min, init 1)
- SYSTEM: COLUMN PROTECT (0=Off, 1=On, init 1), DEPRESS TIME (0.1-60.0 min, init 5.0)
- SYSTEM: COOL RDY CHECK (0=Off, 1=On, init 1), COOL RDY RNG (1-10 °C, init 5), COOL W.TIME (0-10 min, init 5)
- SYSTEM: PRESSURE UNIT (0=MPa, 1=kgf/cm2, 2=bar, 3=psi, init 0)
- UTILITY: KEY CLOSE, BRIGHTNESS (1-4, init 4), BEEP MODE (0=On, 1=Alarm, 2=Off, init 0), VOLUME (1-3, init 2), DISP OFF TIME (0-10 min, init 1), DIRECT KEY MODE (0=Confirm, 1=Direct, init 0)
Describes VP functions for product information, maintenance logs and usage timers, validation tests, and calibration settings. Includes password-protected calibration group.
- PRODUCT INFO: serial numbers and software versions for instrument, CO2 cooler, CO2 pump, heat exchanger
- MAINTENANCE: TOTAL OP TIME, BPR BLOCK USED, BPR DIAPHRAGM USED, COOL PUMP USED, PART REPLACEMENT, MAINTENANCE LOG (20 records), OPERATION LOG (10 records), ERROR LOG (10 records), COOL/HEX logs
- VALIDATION: DATE, TIME, MEMORY CHECK, LEAK SENSOR CHECK, PULSE CHECK (acceptance criterion 0.05-1.00 MPa), BPR HEATER1 TEST, BPR HEATER2 TEST, PUMP COOLER TEST, HEX HEATER TEST
- CALIBRATION: password (default 00000), BPR BLOCK replacement time (0-9999 h, init 2000), BPR DIAPHRAGM replacement time (0-9999 h, init 2000), COOLANT PUMP replacement time (0-9999 h, init 2000), COOL TEMP (0-10 °C), CANCEL COVER SW (0=detect, 1=do not detect), DATE FORMAT, INITIALIZE PARAM, PARAMETER LOCK, PASSWORD SETTING, CBM PARAMETER, LEAK THOLD, PRESS COMP FACTOR
Describes setting the instrument for connection to a system controller or workstation, including local/remote mode and link address. Also covers external input/output terminals and remote cable connection.
- LOCAL MODE: 0=Remote (via system controller), 1=Local (independent)
- LINK ADDRESS: 1 to 12
- External I/O terminals on rear panel
- Remote cable connection described in section 3.8.2
Covers periodic inspection and maintenance, removing the right panel cover and filter holder, replacing the BPR diaphragm (and BPR block), resetting usage time, replacing the air filter, cleaning the leak tray, and cleaning the exterior.
- Periodic inspection and maintenance list
- Removing right panel cover and filter holder
- Replacing diaphragm (replacing BPR block)
- Resetting usage time of diaphragm/BPR block
- Replacing air filter
- Cleaning leak tray
- Cleaning exterior
Provides instrument specifications and lists consumable and replacement parts.
- Specifications section
- Consumable parts list
- Replacement parts list
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| BPR pressure setting range is 1.00 to 40.00 MPa; initial value is 10.00 MPa. p.16 ↗ | Select operating pressure within this range; default is 10.00 MPa. | SFC-40P |
| BPR temperature setting range is 0 °C to 70 °C; initial value is 50 °C. p.17 ↗ | Set BPR temperature within 0-70 °C to prevent freezing and maintain control. | SFC-40P |
| Heat exchanger temperature setting range is 0 °C to 70 °C in increments of 1 °C; initial value is 40.0 °C. p.24 ↗ | HEX temperature can be set in 1 °C steps within 0-70 °C. | SFC-40P with heat exchanger |
| When turning pump head cooling off and then on again, wait until the cooling block temperature reaches 5 °C or higher to avoid freezing in the piping. p.20 ↗ | Prevents ice formation and flow blockage in the CO2 pump head cooling circuit. | SFC-40P with CO2 pump |
| BPR ready check range (BPR RDY RNG) is 0.5 to 1.0 MPa; initial value 0.5 MPa. A pressure within set pressure ± range is considered stable. p.27 ↗ | Adjusts the tolerance for declaring BPR pressure ready. | SFC-40P |
| BPR WAIT TIME is 0 to 10 min; initial value 1 min. If set to 0, READY is displayed unconditionally. p.27 ↗ | Controls delay before READY after pressure reaches setpoint. | SFC-40P |
| Column protection depressurizing time (DEPRESS TIME) is 0.1 to 60.0 min; initial value 5.0 min. p.28 ↗ | Gradual depressurization time to protect columns after pressure control is turned off. | SFC-40P |
| COOL RDY RNG is 1 to 10 °C; initial value 5 °C. COOL W.TIME is 0 to 10 min; initial value 5 min. p.28 ↗ | Sets pump head temperature stability criteria for READY status. | SFC-40P with CO2 pump |
| Pressure unit conversion: 1 kgf/cm2 = 0.098 MPa = 0.98 bar = 14.2 psi. p.29 ↗ | Use for converting displayed pressure units. | SFC-40P |
| PULSE CHECK acceptance criterion range is 0.05 to 1.00 MPa; measurement duration is 1 minute. p.41 ↗ | Used to verify pressure control accuracy. | SFC-40P |
| Estimated replacement time for BPR block, BPR diaphragm, and refrigerant pump is 2000 hours (initial value); settable 0 to 9999 hours. p.43 ↗ | Maintenance scheduling based on usage time; reset after replacement. | SFC-40P |
| COOL TEMP setting range for pump head temperature is 0 °C to 10 °C. p.43 ↗ | Sets the CO2 pump head cooling temperature. | SFC-40P with CO2 pump |
| Default calibration password is 00000 (five digits). p.42 ↗ | Required to access CALIBRATION group settings. | SFC-40P |
| CANCEL COVER SW setting: 0 = automatically detect BPR cover (initial), 1 = do not detect. Setting 1 keeps temperature control ON without cover, causing dangerous situation and possible burns. p.44 ↗ | Do not change without special reason; safety hazard. | SFC-40P |
| LINK ADDRESS range is 1 to 12; initial value 4. p.26 ↗ | Sets channel number for system controller connection. | SFC-40P |
| USE COOL UNIT: 0 = Use (initial), 1 = Not Use. USE HEX: 0 = Use, 1 = Not Use (initial). p.26 ↗ | Must be set according to whether CO2 cooler and heat exchanger are connected. | SFC-40P |
| Maintenance log stores 20 records; operation log and error log store 10 records each. p.36 ↗ | Log capacity for tracking maintenance and errors. | SFC-40P |
| Power on: press and hold power button for at least 3 seconds. Power off: hold 3 seconds or more, then hold again 1 second on confirmation screen. p.13 ↗ | Correct power sequencing to avoid abrupt shutdown. | SFC-40P |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| SENSOR NO GOOD p.41 ↗ | Leak sensor detection level not adjusted properly. | Adjust the detection level with [LEAK THOLD] function in the Calibration Support Group. | 3.6.4 / p.41 |
| PASSWORD WRONG p.45 ↗ | Incorrect password input when changing password. | Password has not been changed; re-enter correct password. | 3.6.5 / p.45 |