Induztr
Engineering Digest Document Viewer Related Models

c190-e291

Technical Reference Summary

Covered Models & Document Scope

Nexera UC and Nexera UC Prep analytical/preparative supercritical fluid chromatograph and supercritical fluid extraction systems. Document covers system controller CBM-40, solvent delivery pumps LC-40D, LC-40D XR, LC-40D XS, LC-40D X3, LC-20AR, LC-20AP, degassing unit DGU-403, autosampler SIL-40C XR, and detectors SPD-40, SPD-40V, SPD-M40.

Technical Data

Parameter Value Condition
Allowable maximum pressure 44 MPa p.2 ↗ LC-40D
Allowable maximum pressure 70 MPa p.2 ↗ LC-40D XR
Allowable maximum pressure 105 MPa p.2 ↗ LC-40D XS
Allowable maximum pressure 130 MPa p.2 ↗ LC-40D X3
Flow rate settings range 0.0001 to 5.0000 mL/min p.2 ↗ LC-40D, 1.0 to 44 MPa
Flow rate settings range 5.0001 to 10.0000 mL/min p.2 ↗ LC-40D, 1.0 to 22 MPa
Flow rate accuracy ≤ ± 1% or ± 2 μL/min, whichever greater p.2 ↗ under specified conditions
Flow rate precision ≤ 0.06% RSD or 0.02 minSD, whichever greater p.2 ↗
Gradient range of set concentrations 0 to 100 % p.2 ↗ 0.1% step
Gradient concentration accuracy ± 0.5 % p.2 ↗ under specified conditions
Available pH range 1 to 14 p.2 ↗
Weight 10 kg p.2 ↗ LC-40D
Weight 12 kg p.2 ↗ LC-40D XR
Dimensions W 260 × D 500 × H 140 mm p.2 ↗ LC-40D
Operating temperature range 4 to 35 °C p.2 ↗ LC-40D series
Power supply 100 to 240 V AC, 150 VA, 50/60 Hz p.2 ↗ LC-40D series
Plunger capacity 47 μL p.2 ↗ LC-20AR
Flow rate setting range 0.001 to 5 mL/min p.2 ↗ LC-20AR, 1.0 to 50 MPa
Flow rate setting range 5.001 to 20 mL/min p.2 ↗ LC-20AR, 1.0 to 25 MPa
Flow rate accuracy No more than ±1% or ±10 μL/min, whichever is greater p.2 ↗ LC-20AR, 0.1 to 5.0 mL/min
Flow rate precision No more than 0.08% RSD or 0.02 min SD, whichever is greater p.2 ↗ LC-20AR
Number of degassed solvents 3 p.3 ↗ DGU-403
Degassed flow line capacity 400 µL/1 line p.3 ↗ DGU-403
Injection volume accuracy ≤ ± 1 % p.3 ↗ SIL-40C XR, 5 µL injection, n = 20
Wavelength range 190 to 700 nm p.4 ↗ SPD-40

Operating Limits

  • Operating temperature range: 4 to 35°C p.2 ↗
  • Allowable maximum pressure: 130 MPa p.2 ↗
  • Available pH range: 1 to 14 p.2 ↗