Instruction Manual - Nexis GC-2030
Technical Reference Summary
Covered Models & Document Scope
This document is the Instruction Manual for the Shimadzu Nexis™ GC-2030 gas chromatograph. It covers troubleshooting and maintenance procedures for the GC-2030 system, including injection ports (SPL, PTV, OCI), detectors (FID, TCD, PTCD, FTD, FPD), flow controllers (AFC, APC), and related components. No specific model variants or ordering codes are listed in the provided pages.
Section-by-Section Technical Breakdown
This section addresses problems where pressure or flow rate cannot be set or does not reach the set value, and where gas leaks cannot be stopped. It provides causes and solutions, including supply pressure ranges, leak check procedures for capillary and packed columns, and corrective actions for leaking parts.
- Supply pressure ranges: carrier gas (other than hydrogen) 300–980 kPa; carrier gas (hydrogen) 150–500 kPa; detector gas 300–980 kPa; hydrogen 150–300 kPa; air 300–500 kPa.
- Carrier gas leak check for capillary columns: set inlet pressure to 150 kPa, total flow 100 mL/min, wait 10 minutes, and compare pressure drop against leak judgment table.
- Leak judgment: pressure drop >2.5 kPa (below 150 kPa), >5.0 kPa (150–450 kPa), >7.5 kPa (450–600 kPa), >10.0 kPa (600–750 kPa), >12.5 kPa (above 750 kPa).
- Leak check points: septum, around injection port, tubing connections, split vent, purge vent, column connection.
- Corrective actions: replace septum, replace O-ring of glass insert, replace seal of blank nut, replace blank nut, replace ferrule/column nut/ClickTek connector.
- Do NOT use leak detecting fluid or soapy water on connections above the carrier and detector gas controllers (AFC/APC).
- The instrument allows 0.2 mL/min leakage from split/purge vent.
This section covers the problem of temperature not increasing or not reaching the set value. It lists causes such as GC not started, long start time, heat loss, low maximum temperature setting, detector temperature lower than column temperature, and mismatched INJ/oven programs, along with solutions.
- Press [Start GC] on the [GC Start Sequence] screen to start the GC.
- Set start time at a lower value if heating has not started.
- Close the oven door and close the hole of the insulation to prevent heat loss.
- Increase the maximum temperature setting for the column and other units; set [Max Usable Temp] on the [Column Information] screen correctly.
- Set the detector temperature larger than the column temperature.
- Set the total time of the INJ program to match the total time of the oven program.
This section provides troubleshooting for common detector issues: no peaks or extremely small peaks, high baseline noise, baseline drift or long-term fluctuation, and zero adjustment error. Causes and solutions are listed for all detectors.
- Check signal cable connections of detector and data processing unit.
- Turn [On] the detector control.
- Check for gas leaks at detector-column connection and tighten.
- Install column correctly with correct insertion length on detector side.
- Set detector attenuation, range, and current at appropriate values.
- Perform Auto Zero if zero level is far below '0'.
- Increase injection volume (normally less than 2 µL for liquid samples and less than 1 mL for gas samples) or increase sample concentration.
- Reduce split ratio for split analysis.
- Check syringe for clogging or leaks; replace if necessary.
- Replace glass insert with deactivated type or acid/alkaline treated one depending on component.
- Set detector gas flow and filter time constant at proper (recommended) values.
- Change analytical conditions such as temperature or column flow if incorrect.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Carrier gas supply pressure (other than hydrogen): 300 to 980 kPa. p.55 ↗ | Ensure supply pressure is within this range to allow proper pressure/flow control. | GC-2030 series |
| Carrier gas (hydrogen) supply pressure: 150 to 500 kPa. p.55 ↗ | Use this range when hydrogen is used as carrier gas. | GC-2030 series |
| Detector gas supply pressure: 300 to 980 kPa. p.55 ↗ | Ensure detector gas supply is within this range. | GC-2030 series |
| Hydrogen supply pressure for detector: 150 to 300 kPa. p.55 ↗ | Use this range for hydrogen detector gas. | GC-2030 series |
| Air supply pressure: 300 to 500 kPa. p.55 ↗ | Ensure air supply is within this range. | GC-2030 series |
| Carrier gas leak check for capillary columns: set inlet pressure to 150 kPa, total flow to 100 mL/min, wait 10 minutes, and compare pressure drop to leak judgment table. p.57 ↗ | Use this procedure to verify system leak-tightness. | GC-2030 with capillary column |
| Leak judgment: if initial inlet pressure is below 150 kPa, a decrease of more than 2.5 kPa indicates a leak. p.58 ↗ | Use this threshold to determine if a leak exists. | GC-2030 series |
| Leak judgment: if initial inlet pressure is 150 to 450 kPa, a decrease of more than 5.0 kPa indicates a leak. p.58 ↗ | Use this threshold to determine if a leak exists. | GC-2030 series |
| Leak judgment: if initial inlet pressure is 450 to 600 kPa, a decrease of more than 7.5 kPa indicates a leak. p.58 ↗ | Use this threshold to determine if a leak exists. | GC-2030 series |
| Leak judgment: if initial inlet pressure is 600 to 750 kPa, a decrease of more than 10.0 kPa indicates a leak. p.58 ↗ | Use this threshold to determine if a leak exists. | GC-2030 series |
| Leak judgment: if initial inlet pressure is above 750 kPa, a decrease of more than 12.5 kPa indicates a leak. p.58 ↗ | Use this threshold to determine if a leak exists. | GC-2030 series |
| The instrument allows 0.2 mL/min leakage from split/purge vent. p.60 ↗ | This is an acceptable leak rate; higher leakage may require maintenance. | GC-2030 series |
| Do NOT use leak detecting fluid or soapy water on connections above the carrier and detector gas controllers (AFC/APC). p.60 ↗ | Drips may damage the controller; use a leak detector instead. | GC-2030 series |
| Normally, injection volume should be less than 2 µL for liquid samples and less than 1 mL for gas samples. p.62 ↗ | Exceeding these volumes may contaminate the injection port. | GC-2030 series |
| For leak check, use blank nut (G-type P/N: 221-35566-92) and column gasket (P/N: 201-35184) on purge vent; for SPL or PTV, also install blank nut on split vent. p.57 ↗ | These parts are required to seal vents during leak testing. | GC-2030 with SPL/PTV |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| No gas is supplied. p.85 ↗ | No gas is supplied. | Open the gas cylinder main valve to supply gas. Turn [On] the detector/injection port control. Turn [On] the detector gas control. | See also 7.5 Injection Port (P.196), 7.4 Detector (P.139), 7.4.9 Settings for Detector Gas Program (P.175) |
| Supply pressure is low. p.85 ↗ | Supply pressure is low. | Change the supply pressure of the gas. Carrier gas (other than hydrogen): 300 to 980 kPa; Carrier gas (Hydrogen): 150 to 500 kPa; Detector gas: 300 to 980 kPa; Hydrogen: 150 to 300 kPa; Air: 300 to 500 kPa. | See also 8.3.2 Gas Supply Pressure (P.348) |
| Gas leaks. p.85 ↗ | Gas leaks. | Check for gas leaks, and tighten leaking connections. Replace the gasket or septum. Additionally tighten the graphite ferrule. If there is leakage even after additional tightening, replace it. | See also Carrier gas leak check (Capillary) (P.57) |
| Set value is incorrect. p.86 ↗ | Set value is incorrect. | Set pressure at a proper value. | See also 8.6 AFC, APC (P.372) |
| Set value is outside controllable range. p.86 ↗ | Carrier gas type or column information is incorrect. | Set the carrier gas type or column information correctly. | See also 7.5 Injection Port (P.196), 7.6.2 [Column Information] Screen (P.241) |
| Total flow is too low, and pressure cannot increase. p.86 ↗ | Total flow is too low, and pressure cannot increase. | Increase the total flow. | See also 8.6 AFC, APC (P.372) |
| Split ratio is set at a low value and total flow is low. p.86 ↗ | Split ratio is set at a low value and total flow is low. | Increase split ratio to increase the total flow. | |
| Split line filter is clogged. p.86 ↗ | Split line filter is clogged. | Replace the filter of the split line. | See Maintenance Help |
| Even if total flow is increased, the actual flow of the split vent is low and the pressure is high. p.86 ↗ | Split line filter is clogged. | Replace the filter of the split line. | See Maintenance Help |
| Part is incorrectly installed. p.86 ↗ | Part is incorrectly installed. | Install parts correctly. | |
| Graphite ferrule or other seal such as O-ring is damaged. p.86 ↗ | Graphite ferrule or other seal such as O-ring is damaged. | Replace O-ring. Additionally tighten the graphite ferrule. If there is leakage even after additional tightening, replace it. | See Maintenance Help |
| Tubing including pipes and joints is damaged. p.86 ↗ | Tubing including pipes and joints is damaged. | Replacement of the tubing part is required. Turn off the main power of the instrument, shut off gas supplies, and contact your Shimadzu sales/service representative. | |
| GC has not started yet. p.91 ↗ | GC has not started yet. | Press [Start GC] on the [GC Start Sequence] screen to start the GC. | See also 7.1.1 [GC Start Sequence] Screen (P.98) |
| Because start time is too long, heating has not started yet. p.91 ↗ | Because start time is too long, heating has not started yet. | Set the start time at a lower value. Immediately after this setup value is changed, the new value is in effect. | See also 7.1.1 [GC Start Sequence] Screen (P.98) |
| Heat loss is large. p.91 ↗ | Heat loss is large. | Close the oven door. Close the hole of the insulation. | |
| Because maximum temperature is set at a very low value, an overheat error occurs. p.91 ↗ | Because maximum temperature is set at a very low value, an overheat error occurs. | Increase the maximum temperature setting for the column and other units. Set [Max Usable Temp] on the [Column Information] screen correctly. | See also [Max Temperature] Screen (P.284), 7.6.2 [Column Information] Screen (P.241) |
| Detector temperature is lower than the set value of column temperature. p.91 ↗ | Detector temperature is lower than the set value of column temperature. | Set the detector temperature larger than the column temperature. | See also 7.4 Detector (P.139) |
| The INJ program does not match the oven program. p.91 ↗ | The INJ program does not match the oven program. | Set the total time of the INJ program to match the total time of the oven program. | See also 7.5.4 [INJ/FLOW] Screen (OCI) (P.208), 7.5.5 [INJ/FLOW] Screen (PTV) (P.212) |
| Signal cable of the detector is connected incorrectly. p.91 ↗ | Signal cable of the detector is connected incorrectly. | Connect the signal cable correctly. | See Maintenance Help |
| Detector control is turned [Off]. p.91 ↗ | Detector control is turned [Off]. | Turn [On] the detector control. | See also 7.4 Detector (P.139) |
| Signal cable of the data processing unit is connected incorrectly. p.91 ↗ | Signal cable of the data processing unit is connected incorrectly. | Connect the signal cable correctly. | See also 8.9 Connecting to External Devices (P.404) |
| There is a leak at the connection of the detector with the column. p.91 ↗ | There is a leak at the connection of the detector with the column. | Check for gas leaks, and tighten leaking connections. | See also Carrier gas leak check (Capillary) (P.57) |
| The length of the inserted column on the detector side is incorrect. p.91 ↗ | The length of the inserted column on the detector side is incorrect. | Install the column correctly. | See Maintenance Help |
| Detector attenuation, range, or current is not appropriate. p.92 ↗ | Detector attenuation, range, or current is not appropriate. | Set the detector attenuation, range, and current at appropriate values. | See also 7.4 Detector (P.139) |
| Zero level is far below "0". p.92 ↗ | Zero level is far below "0". | Check the signal level on the Monitor screen and perform Auto Zero. | See also 7.2 Monitor (P.111) |
| Injection volume is too low. Or sample concentration is low. p.92 ↗ | Injection volume is too low. Or sample concentration is low. | Increase the injection volume. (Normally, less than 2 μL for liquid samples and less than 1 mL for gas samples.) Large injection volume may contaminate the injection port. Increase the sample concentration. | |
| Split ratio is too high. (For split analysis) p.92 ↗ | Split ratio is too high. (For split analysis) | Reduce the split ratio. | See also 7.5.1 [INJ/FLOW] Screen (SPL) (P.196), 7.5.5 [INJ/FLOW] Screen (PTV) (P.212) |
| Sample is injected incorrectly. p.92 ↗ | Sample is injected incorrectly. | Check the clogging or leaks of the syringe and repair it. Replace the syringe. | See Maintenance Help |
| Component is adsorbed in the glass insert. p.92 ↗ | Component is adsorbed in the glass insert. | Replace the glass insert with deactivated type. Replace the glass insert with acid or alkaline treated one depending on the component. | See Maintenance Help |
| Detector gas flow is incorrect. p.92 ↗ | Detector gas flow is incorrect. | Set it at a proper (recommended) value. | See Maintenance Help |
| Detector filter time constant is incorrect. p.92 ↗ | Detector filter time constant is incorrect. | Set it at a proper (recommended) value. | See also 7.4.10 [Signal Settings] Screen (P.177) |
| Analytical conditions, such as temperature or column flow are incorrect. p.92 ↗ | Analytical conditions, such as temperature or column flow are incorrect. | Change the analytical conditions. |