| 00 | | R/W | Clear Variable Command Word. | |
| 01 | RESERVED | | | |
| 02 | CPLOCKOUT | R/W | The control panel lockout feature, ON or OFF. | Word 5: 0=OFF, 1= ON |
| 03 | CURRDATE | R/W | The system date as kept by the I/O PCB and transferred to the CPU on a Write. | Word 5: mm, Word 6: dd, Word 7: yyyy |
| 04 | DISPLAY | R/W | The display bar graph mode, log or lin. | Word 5: 1 = LOG, 0 = LIN |
| 05 | EXPONENT | R/W | The manual-mode exponent. | Word 5: xx i.e. 03 or 10 |
| 06 | EXTLEAK | R/W | The external calibrated leak.value. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent) Word 7: 0 = neg, 1 = pos exponent |
| 07 | GAIN | R/W | The system gain value. | Word 5: xx (gain X 10 i.e. 22 = 2.2) |
| 08 | RESERVED | | | |
| 09 | TIME | R/W | The system time as kept by the I/O PCB and transferred to the CPU on a Write. | Word 5: hhmm,Word 6: ssap (ap = 1 for pm) |
| 10 | RESERVED | | | |
| 11 | LKEXPIRE | R/W | The internal calibrated leak expiration date mm/dd/yyyy. | Word 5: mm, Word 6: dd, Word 7: yyyy |
| 12 | LEAKTEMP | R/W | The internal calibrated leak temperature that the factory calibration was performed at (degrees C). | Word 5: tt (temp X 10) |
| 13 | TPPRESSURE | R | The current Test Port pressure in mTorr. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent) Word 7: 0 = neg exponent |
| 14 | SYPRESSURE | R | The current System Pressure in mTorr. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent) Word 7: 0 = neg exponent |
| 15 | RESERVED | | | |
| 16 | RESERVED | | | |
| 17 | RANGE | R/W | The number indicating the exponent of the least sensitive range of detectable leak. The reported leak rates are in this range and three lower (more negative exponent) ranges. | Word 5: yy (exponent), Word 6: 0 = neg, 1 = pos exponent |
| 18 | RANGESTOP | R/W | The Rangestop variable. | Word 5: yy (exponent), Word 6: 0 = neg, 1 = pos exponent |
| 19 | 1REJECT | R/W | The reject 1 leak rate setpoint. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent |
| 20 | 2REJECT | R/W | The reject 2 leak rate setpoint. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent |
| 21 | 3REJECT | R/W | The reject 3 leak rate setpoint. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent |
| 22 | 4REJECT | R/W | The reject 4 leak rate setpoint. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent |
| 23 | RESERVED | | Reports the currently stored Test Sequencer rough time used as part of setting up the Auto Sequencer. | |
| 24 | STDLEAK | R/W | The calibration standard leak rate. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent |
| 25 | STDLEAKt | R | Reports the temperature corrected value of the internal calibrated leak. If the value is out of range or the temp cannot be read, the STDLEAK value is returned. | Word 5: xxx (mantissa X 1000), Word 6: yy (exponent), Word 7: 0 = neg, 1 = pos exponent |
| 26 | RESERVED | | | |
| 27 | SYSPWD | R | The value of the system password. | |
| 28 | RESERVED | | | |
| 29 | TEMPFACTOR | R/W | Coefficient of temperature change of the cal leak (Determined at the factory when the cal leak was calibrated). | Word 5: tt (temp X 10), Word 6: 1 = neg |
| 30 | RESERVED | | | |
| 31 | VER | R | The CPU software version in the form LC02.05. | Word 5: xxxx (Revision X 100), Word 6: yy (Chksum MSByte), Word 7: Chksum LSByte ex |
| 32 | CHKSUM | R | The checksum of the software in the CPU PCB. | Word 5: Hi Byte, Word 6: Lo Byte |
| 33 | WHYNOCAL | R | The calibration failure diagnostics. | Word 5: 0 to 5 0 = Cal Ok, 1 = No Ion Peak, 2 = No E Peak, 3 = GainTooHigh, 4 = GainTooLow, 5 = LostPower |