Breeze QS HPLC System Getting Started Guide

Technical Reference Summary

Covered Models & Document Scope

This Getting Started Guide (715006394 Rev. A, July 29, 2019) covers the Breeze QS HPLC System software workflow in Empower QS, including single injections, sample sets, processing methods, reporting, and power-down. Pump priming references the Waters 1500-series HPLC pumps: 1515 isocratic, 1525 binary, 1525µ, and 1525EF (extended-flow) binary pumps.

Section-by-Section Technical Breakdown

pp. 6-7
Powering-on and logging in

Describes the power-on sequence for chromatographic instruments and peripheral devices, then starting Empower QS and logging in. After login, the Select Project and System dialog box appears and a project is selected to open the QuickStart interface.

  • Power on chromatographic instruments and wait until all internal calibration and diagnostic routines are complete before powering on the next instrument.
  • Power on peripheral devices connected to the Empower QS system, such as a printer.
  • Start Empower QS by double-clicking the Empower QS icon on the desktop.
  • Enter user name and password in the Login dialog box; the Select Project and System dialog box appears.
  • Select the project of interest and click OK; the QuickStart interface appears and should be expanded for optimum viewing.
pp. 9-10
Equilibrating the system

Explains how to equilibrate the HPLC system and monitor the baseline before an injection. The instrument method is selected in the Setup Equilibrate/System Monitor dialog box, and monitoring is stopped once the baseline stabilizes.

  • Ensure View Acquisition is selected on the navigation bar.
  • Click Equilibrate System/Monitor Baseline in the Acquisition bar.
  • Select the appropriate instrument method from the Instrument Method list and click Equilibrate/Monitor.
  • A baseline appears in the real-time plot; baseline appearance differs for each system.
  • After the baseline stabilizes (becomes relatively flat), click Abort to stop monitoring.
  • Because the software auto-scales the Real-Time plot, inspect the y-axis scale to adjust minimum and maximum values.
pp. 11
Making a single injection

Covers placing a sample vial in the autosampler, specifying single-injection parameters, and starting acquisition. The status bar indicates injection progress and the chromatographic data is displayed in the real-time plot.

  • Place the sample vial in the autosampler in the first position.
  • Click the Inject tool.
  • In the Specify Single Inject Parameters dialog box, specify sample name, method set, vial, injection volume, and run time.
  • Click Inject to begin data acquisition.
  • The status bar indicates when the injection is made and running; data is acquired and displayed in the real-time plot.
  • After the single injection is complete, the system is ready for the next injection.
pp. 11-13, 17-19
Viewing data with the processing method

Describes opening a processing method and reviewing acquired data. Peaks are integrated and named using the Integrate and Calibrate/Quantitate tools, and changes to integration and peak identification parameters are saved to the processing method.

  • On the left-hand navigation pane, click Browse Project.
  • Click the Injections tab and click Update on the toolbar.
  • Highlight the new sample set and click View Data.
  • From the main menu, click File > Open > Processing Method, select the appropriate method, and click Open.
  • Use the Integrate and Calibrate/Quantitate tools to ensure peaks are integrated and named according to the SOP.
  • Make necessary changes to integration and peak identification parameters and save them to the processing method.
pp. 14-16
Loading and running a sample

Explains loading vials into the autosampler and running a sample set. The sample set method is opened, the run is started from the Inject menu, and the status bar tracks progress to Sample Set Complete.

  • Ensure vial locations, sample types, and amounts are correct, then load the vials into the autosampler.
  • Click Open Sample Set Method on the toolbar and select the appropriate sample set method.
  • From the main menu, click Inject > Run Sample Set.
  • In the Run Sample Set dialog box, type the sample set name and click Run.
  • The status bar indicates Sample Set Running and then Sample Set Complete once all injections are made.
pp. 20-22
Processing the sample set

Describes verifying sample information and peak amounts, then processing the sample set. The processing method or method set is selected with How set to Calibrate and Quantitate and Clear Calibration selected.

  • Right-click the new sample set and select Alter Sample.
  • Verify sample-related information such as sample weights and dilutions and make necessary changes.
  • Click Amount on the toolbar, verify peak names and values, and click OK in the Component Editor window.
  • Save all changes and exit the Alter Sample tool.
  • Right-click the sample set and select Process.
  • Per SOP, select a method set or processing method, set How to Calibrate and Quantitate, select Clear Calibration, and click OK.
pp. 23-28
Reporting the results

Covers updating result sets, viewing data with a report method, navigating multi-page and multi-sample reports, and printing individual or batch reports.

  • From the Result Sets menu, click Update.
  • Highlight the new result set and click View Data; the Open Report Method dialog box opens.
  • Select Use the Following Report Method, choose the appropriate report method, and click OK.
  • Use Next Page to navigate a report longer than one page and Next Report to navigate reports for all samples.
  • Click Print for a hard copy; for batch printing, right-click the result set in Browse Project, select Print, choose a printer, select a method set or report method, and click OK.
pp. 29
Powering-down the system

Describes flushing the system with the appropriate mobile phase, slowing or stopping flow, and closing Empower QS. A confirmation prompt appears when exiting QuickStart and logging out.

  • Per SOP, flush the system with the appropriate mobile phase.
  • Either slow the flow rate or stop the flow of mobile phase.
  • Close the Empower QS window.
  • Confirm the message Are you sure you want to exit QuickStart and log out of Empower? by clicking Yes.
pp. 30
Priming the pump

Directs the user to the appropriate Waters 1500-series pump documentation for priming instructions, depending on the pump model installed.

  • For the 1515 isocratic pump or 1525 binary pump, see the 1500-Series Pump Installation and Maintenance Guide.
  • For the 1525µ or 1525EF (extended-flow) binary pump, see the 1525µ and EF HPLC Pump Installation and Maintenance Guide.
pp. 31-46
Creating a processing method

Explains that a processing method defines integration, calibration of standards, and quantitation of unknowns, and is included in a method set. The Processing Method Wizard is used to define the integration region, peak width and threshold, peak rejection, calibration names and retention times, and default amounts.

  • Data must be acquired before defining processing and reporting methods.
  • A processing method defines how Empower QS integrates data, calibrates standards, and quantitates unknowns.
  • The integration region is defined by zooming in on the area of interest; right-click the plot for Full View/Unzoom.
  • Empower QS determines optimum peak width and threshold values for the selected region.
  • Peak rejection criteria can be set by minimum area or height; Empower QS calculates 95% of a selected peak's height for Minimum Height.
  • The Processing Method Wizard allows only a single-point calibration curve; multi-point curves are described later.
  • External Standard Calibration is the default selection in the Calibration - Internal Standards dialog box.
  • After creation, the processing method is added to the method set via View Method > Method Set, File > Open, selecting Training Method Set, and setting Training Processing Method as the Default Processing Method.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Wait until all internal calibration and diagnostic routines are complete before powering on the next chromatographic instrument. p.6 ↗ Prevents initialization conflicts and ensures instruments are ready before software login. Whole series
Power on peripheral devices connected to the Empower QS system, such as a printer, after the chromatographic instruments. p.6 ↗ Ensures peripherals are available when Empower QS starts. Whole series
The software auto-scales the Real-Time plot, so the baseline can appear noisy; inspect the y-axis scale to adjust minimum and maximum values. p.10 ↗ Operators must verify axis scaling before judging baseline stability. Whole series
In the Specify Single Inject Parameters dialog box, specify sample name, method set, vial, injection volume, and run time before clicking Inject. p.11 ↗ All acquisition parameters for a single injection are defined in one dialog. Whole series
The status bar changes to Sample Set Running and then to Sample Set Complete once all injections are made. p.15 ↗ Operators can monitor sample set progress from the status bar. Whole series
When processing a sample set, set How to Calibrate and Quantitate and select Clear Calibration per SOP. p.22 ↗ Ensures calibration is recalculated for the processed sample set. Whole series
For batch printing, right-click the result set in Browse Project, select Print, choose a printer, select a method set or report method, and click OK. p.26 ↗ Batch reports can be generated without opening each result set individually. Whole series
Before power-down, flush the system with the appropriate mobile phase and either slow the flow rate or stop the flow of mobile phase. p.29 ↗ Protects the column and flow path from residual mobile phase. Whole series
For the 1515 isocratic pump or 1525 binary pump, priming instructions are in the 1500-Series Pump Installation and Maintenance Guide. p.30 ↗ Pump priming procedure depends on the installed pump model. 1515, 1525
For the 1525µ or 1525EF (extended-flow) binary pump, priming instructions are in the 1525µ and EF HPLC Pump Installation and Maintenance Guide. p.30 ↗ Pump priming procedure depends on the installed pump model. 1525µ, 1525EF
Data must be acquired before defining the processing and reporting methods. p.31 ↗ A test injection or existing data is required before processing method development. Whole series
The Processing Method Wizard allows only a single-point calibration curve; multi-point calibration curves are described in a later section. p.41 ↗ Multi-point calibration requires interactive method development beyond the wizard. Whole series
External Standard Calibration is the default selection in the Calibration - Internal Standards dialog box. p.41 ↗ Default calibration type is external standard unless changed by the user. Whole series
Empower QS calculates 95% of the selected peak's height for the Minimum Height field and reintegrates the chromatogram. p.37 ↗ Peak rejection threshold is automatically derived from the shortest peak selected. Whole series
If values are manually specified in the peak rejection fields, click Test to view how the values affect integration; if software specifies values, the Test button remains disabled. p.37 ↗ Manual peak rejection settings should be validated with the Test function. Whole series
For PDA detector data, a 2D chromatogram must be extracted from the 3D data to complete the processing method section. p.32 ↗ PDA users require an additional extraction step before processing method development. Whole series with PDA detector
After creating the processing method, add it to the method set by selecting Training Processing Method from the Default Processing Method list and clicking Save. p.45 ↗ The processing method must be linked to the method set for use in sample processing. Whole series