Agilent Technologies 1260 Infinity III Bio-Inert LC System
Agilent Technologies 1260 Infinity III Bio-Inert LC System — Liquid chromatography with Flow rate 0.01 liter/minute, Max pressure 60 megapascal, Pumping speed 6×10⁻⁴ m³/h.
The 1260 Infinity III Bio-inert LC System is a metal-free, titanium-based HPLC system rated to 600 bar with a pH range of 1–13 and 2 M salt tolerance. It combines a Bio-inert pump (10 mL/min at 200 bar), multisampler (carryover <9 ppm), and thermostatted column compartment (10 °C below ambient to 85 °C). It is intended for SEC, ion-exchange, and peptide mapping of biotherapeutics where sample contact with metal must be avoided.
Agilent Technologies Bio LC Solutions Series — Model Comparison & Selection
| Model | Flow rate, liter/minute | Mass range | Max pressure, megapascal | Pumping speed, m³/h | Wavelength range | Sample volume, milliliter |
|---|---|---|---|---|---|---|
| 1260 Infinity III Bio-Inert LC System Current | 0.01 | — | 60 | 6×10⁻⁴ | 190...950 | — |
| 1260 Infinity Amino Acid Analysis System | — | — | — | — | — | — |
| 1290 Infinity III Bio LC | Biocompatible UHPLC system | — | 300...3,200 | — | — | — | 0.005 |
Full specifications
| Flow rate | 10 mL/min |
| Max pressure | 600 bar |
| Pumping speed | 10 mL/min |
| Wavelength range | 190...950 nm |
| Carryover | 9 ppm |
| Data Rate | 74 Hz |
| Internal Volume | 1.0 µL |
| Maximum Data Acquisition Rate | 120 Hz |
| Maximum Individual Bottle Volume | 2 L |
| Maximum System Pressure | 600 bar |
| Salt Tolerance | 2 M |
| Sample Capacity | 6144 samples |
| Solvent Cabinet Maximum Volume | 8 L |
| Maximum system pressure (metal-free sample flow path) mfr | 600 bar |
| Optional injection volume range mfr | 5 mL |
| Double mfr | 7116 A/B |
| Single mfr | 7114 A/B |
| pH short term mfr | 14 |
| Maximum torque for UHP-FF fitting mfr | 0.8 Nm |
| Capillary jut-out before fitting installation mfr | 5 mm |
| Capillary protrusion when sliding fitting on capillary mfr | 5 mm |
| Pressure stability of stainless-steel cladded PEEK capillaries mfr | 600 bar |
| Column Compartment Temperature | 10...85 °C |
| Standard injection volume mfr | 20 μL |
Description
Construction and series positioning
- The system is built around a titanium-based Bio-inert Pump (G5654A) and a 100% metal-free sample flow path using ceramic needle, PEEK needle seat, and stainless-steel-clad PEEK capillaries.
- It sits between the standard 1260 Infinity III Prime Bio LC and the higher-pressure 1290 Infinity III Bio LC; the 1260 Bio-inert is distinguished by its 600 bar pressure limit and full bio-inert flow path, whereas the 1290 Bio LC extends to 1300 bar for UHPLC separations.
- All modules are stackable; configurations include Single Stack (max 4 modules), Two Stacks, InfinityLab Benchtop, and Flex Bench.
Behavior under load and media
- The pump delivers up to 10 mL/min at 200 bar and maintains 600 bar maximum system pressure; flow rate and pressure are interdependent, so high flow rates reduce achievable pressure.
- The flow path tolerates pH 1–13 (short-term 14) and salt concentrations up to 2 M, enabling harsh mobile phases used in bio-purification.
- The Multicolumn Thermostat (G7116A) provides column thermostatting from 10 °C below ambient to 85 °C, supporting sub-ambient SEC and elevated-temperature methods.
- The Bio-inert Multisampler (G5668A) reduces carryover to <9 ppm with multiwash and can hold up to 6144 samples in vials or microtiter plates.
Detection and control
- Detector options include the Diode Array Detector WR (G7115A) with 120 Hz data acquisition and 190–950 nm range, and the Fluorescence Detector (G7121A) with up to 74 Hz data rate.
- LAN configuration must be performed from the module with the highest data rate, typically the detector; the configuration switch at the rear sets the IP method (default 192.168.254.11, DHCP, or stored address).
- The leak sensor stops the running system when the leak detection level is reached; drain connectors must be installed on all modules except the lowest, which connects to waste.
Integration & Operating Notes
DMF corrodes the PVDF leak sensor, leading to false leak detection or sensor failure. If DMF-containing methods are required, consult Agilent for alternative sensor material or additional protective measures; regular inspection of the leak sensor for corrosion is mandatory.
Selection & Integration Q&A
Choose the 1260 Bio-inert for methods requiring up to 600 bar and full metal-free flow path with pH 1–13 and 2 M salt tolerance. The 1290 Bio LC offers higher pressure (up to 1300 bar) for UHPLC; the 1260 Prime Bio LC is a lower-cost biocompatible option but may not have the same inertness. Select based on pressure needs and sample sensitivity to metal.
Use only bio-inert capillaries and fittings; do not mix with bio/biocompatible parts. UHP-FF fittings require 5 mm capillary protrusion and max 0.8 Nm torque. The Bio-inert ZDV union (5067-4741) must be installed in the prescribed sequence. Ensure the flow path is metal-free throughout, using stainless-steel-clad PEEK capillaries rated to 600 bar.
The 1260 Infinity II Max Uptime Kit Bio-inert (G5654-68707) includes capillaries, ZDV unions, fittings, solvent inlet filters, and tools. Valve kits (e.g., G5631A, G5632A, G5639A, G4235A) are separate. The HPLC System Tool Kit is included. Not included: columns, mobile phases, sample vials, and the InfinityLab Level Sensing (G7175A) unless ordered. Regular maintenance includes leak sensor inspection and capillary/fitting replacement.