Agilent Technologies 1260 Infinity II Multidetector GPC/SEC System
Agilent Technologies 1260 Infinity II Multidetector GPC/SEC System — Liquid chromatography.
The 1260 Infinity II Multidetector GPC/SEC System is a high-temperature gel-permeation chromatography platform built around a 220 °C column oven, a 40-vial autosampler with two heated zones, and a differential refractive index detector. It is the high-temperature modification of the 1260 Infinity II GPC/SEC series, distinguished from the ambient Bio-SEC siblings by its 230 VAC/50 Hz supply, 136 kg frame, and solvent-compatible oven. It is used to separate and characterize synthetic polymers by molecular weight in THF or TCB at temperatures up to 220 °C.
Agilent Technologies GPC SEC Solutions Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Sample volume, milliliter | Column particle size, µm | Column temperature, °C | Column inner diameter, mm | Column length, m |
|---|---|---|---|---|---|---|
| 1260 Infinity II Multidetector GPC/SEC System Current | — | 5...500 | — | 220 | — | — |
| 1260 Infinity II Bio-SEC System | 0.024 | — | 1.8 | — | 2.1 | 0.05 |
| 1260 Infinity II Bio-SEC system | Biological polymer analysis | 4.8×10⁻⁵ | 0.01 | 2.7 | 30 | — | — |
| 1290 Infinity II GPC/SEC system | 3.6×10⁻⁶ | 0.02 | 3 | 35 | 4.6 | 0.25 |
Full specifications
| Wavelength range | 890 nm |
| Column Pore Size | 300 Å |
| Column temperature | 220 °C |
| Precision | 1 millimeter |
| Sample volume | 5...500 µL |
| Column dimensions mfr | 7.8 mm |
| Monomer content before sonication mfr | 76.12 |
| Dimer content before sonication mfr | 16.45 |
| Aggregates content before sonication mfr | 7.44 |
| Monomer content after sonication mfr | 83.80 |
| Dimer content after sonication mfr | 12.25 |
| Aggregates content after sonication mfr | 3.95 |
| Safety standards mfr | IEC, CSA, UL UL |
| Flow Rate Reproducibility mfr | 0.07 SD |
| Column Capacity mfr | 6 columns |
| Cell Volume mfr | 8 μL |
| Carousel Capacity mfr | 40 vials |
| Deflection mfr | Differential Refractometer Refractometer |
| Instrument Control mfr | Microprocessor Microprocessor |
| Flow Rate Reproducibility1 mfr | 0.07 temperature |
| Detector Differential Refractometer Deflection Cell Volume mfr | 8 μL |
| Noise2 mfr | 100 μV |
| Drift2 mfr | 10 mV/h |
| Autosampler Carousel Capacity mfr | 40 Vials |
| Precision1 mfr | 1 |
| DMSO mfr | 215 |
| Dibutoxymethane mfr | 220 |
| Tetrahydrofuran mfr | 321 |
| Dimethylformamide mfr | 445 |
| Toluene mfr | 480 |
| Acetone mfr | 465 |
| m-Cresol mfr | 558 |
| Chloronaphthalene mfr | 558 |
| DRI cell volume mfr | 8 μL |
| Carousel capacity mfr | 40 Vials |
| Syringe sample volume mfr | 800 μL |
| DRI reference purge time mfr | 20 min |
| Solvent filter pore size mfr | 0.45 μm |
| Weight | 136 kg |
| Model protein monomer mass mfr | 150 kDa |
| Model protein dimer mass mfr | 300 kDa |
| Size (width × depth × height) mfr | 900 mm |
| Packaged Size (width × depth × height) mfr | 930 mm |
| Bench width mfr | 90 cm |
| Bench depth mfr | 53 cm |
| Bench height mfr | 62 cm |
| Oven open height mfr | 97 cm |
| Line voltage mfr | 230 VAC |
| Line frequency mfr | 50 Hz |
| Power consumption mfr | 3300 BTU |
| Instrument Interface mfr | 7 bit |
| Noise | 100 µV |
| Operating Altitude | 2000 m |
| Non-operating altitude mfr | 4600 m |
| Humidity mfr | 25...30 °C |
| Safety standards: IEC, CSA, UL mfr | 2 |
Description
Construction and positioning
- The system is a floor-standing or bench-mounted assembly (900 mm width, 136 kg) comprising a column oven with eight active heaters (three base, four lid, one fan-mounted), a G7110B isocratic pump, a differential refractive index detector, and a 40-position autosampler.
- Within the 1260 Infinity II GPC/SEC series, this Multidetector model is the high-temperature variant: the Bio-SEC siblings operate at ambient or low temperature with biological mobile phases, while this unit is specified for THF at 40 °C and TCB at 160 °C, with an oven range to 220 °C.
- The oven accepts up to six 30 cm columns; the autosampler carousel holds 40 × 2 mL vials, with position 0 reserved as an index/waste vial.
Behavior under load and in different modes
- The pump provides three independent flow rates (Idle, Low, Run) over 0.0–5.0 mL/min; the Low flow is used in RUN mode while the oven is unstable, and the Run flow once the oven is stable.
- Oven heat rate is programmable from 0.4 to 2 °C/min; below 100 °C the rate is limited to 1 °C/min to prevent overshoot. Temperature stability is <0.05 °C/h.
- The RI detector has an 8 µL cell, noise ≤100 µV, drift ≤10 mV/h at 890 nm; automatic RI conditioning purges and autozeros once the oven is within ±1 °C of setpoint for 10 minutes, repeating until baseline drift falls within the 0.1–1.0 V RI limits.
- Injection volume is 5–500 µL with ≤1% precision; the autosampler hot zone (30–220 °C at vial position 1) and warm zone (25–150 °C at vial position 19) allow a temperature gradient across the carousel, but heaters are enabled only in RUN mode.
Installation, connection and control
- The system requires a dedicated 230 VAC ±10% single-phase supply with earth and RCCB; power consumption is 3300 VA (2310 W/7900 BTU). No voltage selector is present.
- Fan-forced exhausts (4-inch diameter) must be routed to fume extraction with airflow <1.0 SLM.
- Installation must be performed by an Agilent service representative; the pump must be turned on before the HT GPC system to avoid communication errors.
- Control is via a graphical user interface with STANDBY, IDLE, and RUN modes; configuration editors set pump flows and pressure limits, oven IDLE/RUN temperatures and heat rate, vapour sensor threshold, RI autozero/purge cycles, and autosampler zone temperatures and vial positions.
Integration & Operating Notes
The system draws 3300 VA (2310 W) at 230 VAC ±10%, 50 or 60 Hz ±5%, and has no voltage selector. A shared or undersized circuit will trip or cause unstable detector baselines; provide a dedicated single-phase supply with earth leakage protection.
At 136 kg (300 lbs.), the instrument requires at least four persons or lifting equipment to place it. Use a sturdy bench of at least 90 cm width, preferably with wheels, and plan the final location before unpacking to avoid moving the unit after installation.
The four-inch fan-forced exhausts must be connected to fume extraction or a hood with airflow <1.0 SLM. Higher extraction rates can cool the oven unevenly and disturb the RI baseline; verify the extraction balance before running temperature-sensitive methods.
Turning on the G7110B pump first ensures it is initialized and ready before the HT GPC system starts communication. Reversing the order can cause communication errors and require a full restart of the system and PC.
Programmable heat rate is 0.4–2 °C/min, but setpoints below 100 °C must use ≤1 °C/min to prevent overshoot and prolonged stabilization. Above 100 °C, 1 °C/min is recommended; faster rates risk tripping the over-temperature cut-out set about 5 °C above operating temperature.
After a temperature change or solvent switch, the system typically needs 8 hours to reach a stable RI baseline. Plan equilibration overnight before analysis; automatic RI conditioning can then purge and autozero until drift falls within the configured RI limits.
The autosampler zero position is an index/waste vial and must remain empty with a clean, shiny cap. A blocked or contaminated zero vial can cause polymer precipitation and injection errors; inspect and replace the cap as part of routine maintenance.
Selection & Integration Q&A
Choose this model for high-temperature GPC/SEC in THF or TCB up to 220 °C with up to six 30 cm columns. The Bio-SEC siblings are for ambient biological mobile phases; the standard 1260 Infinity II GPC/SEC lacks the 220 °C oven and heated autosampler zones.
Provide a dedicated 230 VAC ±10% single-phase circuit with earth and RCCB, 50 or 60 Hz ±5%, rated for 3300 VA. Route the four-inch fan-forced exhausts to fume extraction with airflow <1.0 SLM. No voltage selector is present, so the supply must match the nameplate.
Operate at 15–30 °C ambient and 25–30% humidity, at altitudes up to 2000 m. Non-operating range is -40 to 70 °C and 4600 m. If shipped cold, let the box warm to room temperature and wait 1 h after unpacking before applying power to avoid condensation damage.
Plan for 8 h RI baseline equilibration after temperature changes, periodic autozero/purge cycles (0–38 cycles, purge 0–3600 s), and replacement of the zero-position vial cap. The base delivery excludes the G7821B viscometer and G7822B dual-angle light scattering detectors, which are optional add-ons.