Agilent Technologies 1290 Infinity II GPC/SEC system
Agilent Technologies 1290 Infinity II GPC/SEC system — Liquid chromatography with Flow rate 3×10⁻⁴ liter/minute, Pumping speed 3.6×10⁻⁶ m³/h.
The 1290 Infinity II GPC/SEC System is a high-performance gel permeation chromatography system configured for isocratic THF operation with a 1290 Infinity II pump and RID detection. It supports flow rates from 0.06 mL/min (micro-GPC) to 1.0 mL/min and column temperatures up to 35 °C, distinguishing it from lower-pressure 1260 Infinity II siblings. It is used for molecular-weight distribution analysis of polymers and resins.
Agilent Technologies GPC SEC Solutions Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Sample volume, milliliter | Column particle size, µm | Column temperature, °C | Column inner diameter, mm | Column length, m |
|---|---|---|---|---|---|---|
| 1290 Infinity II GPC/SEC system Current | 3.6×10⁻⁶ | 0.02 | 3 | 35 | 4.6 | 0.25 |
| 1260 Infinity II Bio-SEC System | 0.024 | — | 1.8 | — | 2.1 | 0.05 |
| 1260 Infinity II Bio-SEC system | Biological polymer analysis | 4.8×10⁻⁵ | 0.01 | 2.7 | 30 | — | — |
| 1260 Infinity II Multidetector GPC/SEC System | — | 5...500 | — | 220 | — | — |
Full specifications
| Flow rate | 0.3 mL/min |
| Pumping speed | 0.06 mL/min |
| Column Particle Size | 3 µm |
| Column temperature | 35 °C |
| Data Rate | 9.25 Hz |
| Filter Membrane Diameter | 47 mm |
| Filter Membrane Pore Size | 0.2 μm |
| Ptfe Frits Per Pm Kit | 5 pk |
| Retention Time | 13.805 min |
| Sample volume | 20 µL |
| Solvent Bottle Volume | 1 L |
| Solvent Filtration Assembly Flask Volume | 1 L |
| Solvent Filtration Assembly Funnel Volume | 250 mL |
| Solvent Inlet Filter Pore Size | 10 μm |
| Well-plate Closing Mat Quantity | 50 pk |
| Well-plate Volume | 0.5 mL |
| Signal polarity mfr | Positive Positive |
| Quick Connect heat exchanger flow rate mfr | 2.5 mL/min |
| Capillary dimensions mfr | 0.12 mm |
| Preventive maintenance cost reduction mfr | 30 |
| Pump seals per PM kit mfr | 4 |
| Seal cap assembly sets per PM kit mfr | 2 |
| Gold seals for outlet valve per PM kit mfr | 2 |
| Seal washes per PM kit mfr | 4 |
| Solvent bottles per convenience kit mfr | 3 amber |
| Mp mfr | 1314 g/mol |
| Mn mfr | 1048 g/mol |
| Mw mfr | 1265 g/mol |
| Solvent consumption per sample mfr | 30 |
| Flow Rates mfr | 0.06 mL/min |
| Stop Time mfr | 22 minutes |
| RID mfr | 35 positive |
| Connection capillary id mfr | 75 μm |
| Micro GPC flow rate mfr | 60 μL/min |
| Column Inner Diameter | 4.6 mm |
| Column Length | 250 mm |
| Connection capillary length mfr | 105 mm |
| Quick Turn Fitting Pressure Limit | 800 bar |
| Solvent bottle thread mfr | 45 |
| Calibration standard Mp mfr | 47,190 g/mol |
| Standard RID cell volume mfr | 8 μL |
Description
Construction and positioning
- The system is built around a 1290 Infinity II high-pressure pump, an autosampler (20 µL injection), a thermostatted column compartment (35 °C), and a refractive index detector (RID) with optical unit at 35 °C.
- It is the high-performance GPC/SEC variant in the Infinity II line; sibling models such as the 1260 Infinity II LC System and 1220 Infinity II LC System are lower-pressure, analytical-scale platforms, while the 1260 Analytical-Scale LC Purification System is for purification, not GPC/SEC.
- The 1290 Infinity II GPC/SEC System is specifically configured for isocratic THF (Channel B) and is compatible with both the 1290 Infinity II Micro-RID (G7162A) and the 1260 Infinity II RID (G7162B) per the application note.
Behavior under load and in different modes
- Flow rate range spans 0.06 mL/min (micro-GPC with 2.1 mm columns) to 1.0 mL/min (conventional 4.6 mm columns), allowing the same system to run both micro and analytical GPC/SEC methods.
- The Quick Connect heat exchanger is rated for 2.5 mL/min, providing thermal equilibration of the mobile phase before the column.
- Quick-turn fittings are rated to 800 bar, supporting the high-pressure operation of the 1290 Infinity II pump.
- Sample volume of 20 µL and well-plate volume of 0.5 mL define the typical injection and sample storage format.
Installation, connection, and operation
- Solvent bottles are 1 L with a 45 mm thread; the solvent inlet filter has 10 µm pores to protect the pump.
- Solvent filtration assembly uses a 1 L flask and 250 mL funnel with 47 mm, 0.2 µm membranes.
- Capillary dimensions are 0.17 mm ID, and the system uses PTFE frits (5 per PM kit) and well-plate closing mats (50 per pack).
- Preventive maintenance kits include 2 pump seals, 2 seal cap assembly sets, 2 gold seals for the outlet valve, and 4 seal washes; the PM program is stated to reduce maintenance cost by 30%.
Integration & Operating Notes
Quick-turn fittings are rated to 800 bar; operating above this pressure risks fitting failure and leakage. The 1290 Infinity II pump can generate high pressures, so method development must respect this limit, especially with narrow-bore columns.
The Quick Connect heat exchanger is specified for 2.5 mL/min. Higher flow rates may reduce thermal equilibration efficiency and cause retention-time drift, so methods above this flow should be validated for column temperature stability.
The preventive maintenance kit contains 2 pump seals, 2 seal cap assembly sets, 2 gold seals for the outlet valve, and 4 seal washes. These consumables should be replaced at the manufacturer-recommended interval to avoid leaks and pressure fluctuations; the PM program is stated to reduce maintenance cost by 30%.
Solvent inlet filters have 10 µm pores, and solvent filtration assemblies use 47 mm, 0.2 µm membranes. Particulates in the mobile phase can damage the pump and column; filtering THF and using the inlet filter are essential for reliable GPC/SEC results.
Selection & Integration Q&A
Choose the 1290 Infinity II GPC/SEC System for high-performance GPC/SEC with flow rates down to 0.06 mL/min and pressures up to 800 bar, especially for micro-GPC with 2.1 mm columns. The 1260 Infinity II LC System is a lower-pressure analytical HPLC platform not optimized for GPC/SEC.
The application note describes use with either the 1290 Infinity II Micro-RID (G7162A) or the 1260 Infinity II RID (G7162B). Columns tested include a single MesoPore 2.1 × 250 mm (PL1913-5325) at 0.06 mL/min and two PLgel 3 µm MiniMixE 4.6 × 250 mm (PL1510-5300) at 1.0 mL/min.