Agilent Technologies 240FS AA
Agilent Technologies 240FS AA — Atomic spectroscopy.
The Agilent 240FS AA is a Fast Sequential flame atomic absorption spectrometer with transverse Zeeman background correction and an 8-position lamp turret, measuring up to 10 elements per sample in about 2 minutes. It sits below the 280FS AA in the flame series and is distinct from the 240Z/280Z graphite-furnace Zeeman instruments. It is used for sequential multi-element trace metal determination in liquid samples by air/acetylene or nitrous-oxide/acetylene flame.
Agilent Technologies Flame Atomic Absorption Instruments Series — Model Comparison & Selection
Full specifications
| Acetylene Purity | 99.5 % |
| Analysis Time (8 Elements) | 95 minutes |
| Dilution Error | -1 % |
| Direct Viewing Distance From Flame | 0.2 m |
| Exempt Distance From Flame | 7.1 m |
| Maximum Flame Door Open Exposure | 13 minutes |
| Precision | 1 %RSD |
| Sample Consumption | 10 mL |
| Sensitivity Increase | 40 % |
| Viewing Distance Rg1 | 2.3 m |
| Viewing Distance Rg2 | 0.8 m |
| Lamps mfr | four four |
| Background correction mfr | Transverse Zeeman Zeeman |
| Dilution factor mfr | 1 |
| Number of determinations mfr | 53 |
| Reproducibility mfr | 1 |
| Elements measured per sample mfr | 10 elements |
| Analysis time per sample mfr | 2 minutes |
| Lamp positions mfr | 8 positions |
| Default operating mode mfr | Fast Sequential Sequential |
| Exhaust ventilation rate mfr | 6 minute |
| Exhaust fan distance from flame mfr | 3 meters |
| UV intensity emitted mfr | 37.7 |
| UV hazard classification mfr | 3 |
| Burner slot replacement limit, nitrous-oxide/acetylene mfr | 0.47 mm |
| Burner slot replacement limit, air/acetylene mfr | 0.54 mm |
| Acetylene maximum pressure mfr | 105 kPa |
| Acetylene fuel supply pressure range mfr | 65 kPa |
| Acetylene cylinder minimum pressure before replacement mfr | 700 kPa |
| Acetylene maximum consumption rate mfr | 1 hour |
| Solvent minimum specific gravity mfr | 0.75 |
| Acetylene fuel supply pressure mfr | 65...100 kPa |
| Acetylene cylinder replacement pressure mfr | 700 kPa |
| Burner slot replacement threshold mfr | 0.54 mm |
| Minimum solvent specific gravity mfr | 0.75 |
| Schedule mfr | 62 |
| Cleaning mfr | 63 |
| Gas cylinders mfr | 64 |
| Preventing flashbacks mfr | 65 |
| Spare parts mfr | 65 |
| Ultraviolet Hazard mfr | 3 |
| Acetylene cylinder minimum pressure mfr | 700 kPa |
| Acetylene maximum consumption mfr | 1 hour |
| Zeeman magnetic field peak mfr | 0.8 Tesla |
| Zeeman magnet exclusion distance mfr | 300 mm |
| UV intensity mfr | 37.7 |
| Hollow cathode lamp capacity 240 series mfr | Four lamp capacity capacity |
| Hollow cathode lamp capacity 280 series mfr | Eight lamp capacity capacity |
| Burner slot width maximum, N2O/Acetylene Mark 7 mfr | 0.46 mm |
| Burner slot width maximum, Air/Acetylene Mark 7 mfr | 0.54 mm |
| Burner slot maximum width mfr | 0.54 mm |
| Nitrous-oxide/acetylene burner slot maximum width mfr | 0.47 mm |
| Air/acetylene burner slot maximum width mfr | 0.54 mm |
| Mark 7 N2O/acetylene burner slot maximum width mfr | 0.46 mm |
| Mark 7 air/acetylene burner slot maximum width mfr | 0.54 mm |
| Non-Zeeman 240/280FS fuse rating mfr | 4 mm |
| Zeeman circuit breaker rating mfr | 15 breaker |
| Zeeman fuse T3.15 A rating mfr | 3.15 mm |
| Zeeman fuse F315 mA rating mfr | 315 mm |
| Zeeman fuse F1A rating mfr | 1 mm |
| Zeeman fuse T6.3 A rating mfr | 6.3 |
| Zeeman fuse F800 mA rating mfr | 800 mm |
| Standard concentration mfr | 5 mg/L |
Description
Construction and series positioning
- Flame-only Fast Sequential AA: samples are aspirated once and measured sequentially for multiple elements, with transverse Zeeman background correction and an 8-position lamp turret (four lamps in use at a time).
- Within the flame series it is the lower-tier Fast Sequential model; the 280FS AA is the sibling flame instrument, while 240Z/280Z AA are graphite-furnace Zeeman instruments and Duo systems combine one flame and one furnace unit under one computer.
- Default operating mode is Fast Sequential; the SIPS 10 (single pump) and SIPS 20 (dual pump) sample introduction pump systems are separate accessories for SpectrAA FAAS instruments.
Behavior and performance
- Throughput: 10 elements per sample in about 2 minutes; an 8-element analysis is quoted at 95 minutes, and precision is 1 %RSD with 10 mL sample consumption.
- SIPS dilution performance is quoted as -1 % dilution error at a dilution factor of 1 with a 5 mg/L standard, 53 determinations and 1 reproducibility; sensitivity increase is 40 %.
- Flame operation is limited to air/acetylene and nitrous-oxide/acetylene; oxygen or oxygen-enriched air must never be used as oxidant.
Installation, connection and control
- Controlled by Agilent SpectrAA software (October 2025 version 5.6.8 or higher per the manual); communications via USB-GPIB HS/HS+; SPS 4 autosampler and Tube-CAM furnace camera are supported options.
- The instrument must be vented to outside air through a self-contained collector hood, ducting and exhaust fan; exhaust capacity at least 6 m3/min with the fan at least 3 m from the flame.
- Gas connections use left-hand threads for fuel gases and right-hand threads for support gases to prevent cross-connection.
Integration & Operating Notes
The flame door interlock prevents ignition if the door is opened during ignition; if the interlock is lost, contact Agilent immediately. Bypassing it is prohibited. With the flame door open, UV exposure must not exceed 13 minutes over an 8-hour period, and direct viewing at 0.2 m or less is Risk Group 3 (UV intensity 37.7 mW/m2).
Perform leak tests at all joints and seals daily before use with soapy water or a proprietary solution, never with a naked flame. Weekly, check the O-rings inside the burner, spray chamber and bung and the two sealing the gas inlets to the spray chamber; damaged O-rings can cause gas leakage, fire or flashback and must be replaced immediately.
Do not connect the instrument to mains until the operating voltage is verified for the specific outlet. Use only the Agilent-supplied power cord, replacing it only with an equivalent specified in the AA Site Preparation Guide, to ensure electrical safety and proper grounding.
Selection & Integration Q&A
Choose the 240FS AA for sequential multi-element flame analysis (up to 10 elements per sample in about 2 minutes) with transverse Zeeman background correction. The 280FS AA is the higher sibling in the same flame Fast Sequential series; 240Z/280Z AA are graphite-furnace Zeeman instruments for trace levels, and Duo systems combine one flame and one furnace unit under one computer.
Acetylene must be instrument grade, at least 99.5 % purity in acetone, supplied at 65–100 kPa and never above 105 kPa; replace cylinders below 700 kPa. Use left-hand threads for fuel and right-hand threads for support gases, and avoid copper or brass with more than 65 % copper. Only air, nitrous oxide and acetylene may be used as oxidant/fuel; never oxygen.
The exhaust system must extract at least 6 m3/min (200 cfm) with the fan at least 3 m from the flame, venting to outside air via a collector hood and ducting, never within the building. With the flame door open, UV exposure is limited to 13 minutes per 8-hour period; direct viewing at 0.2 m or less is Risk Group 3 (37.7 mW/m2).
Daily leak tests and weekly O-ring checks (burner, spray chamber, bung, gas inlets) are mandatory; the maintenance schedule also covers exhaust, gas hoses, drain tube, burner, spray chamber, liquid trap, compressor filter and coolant. The SIPS 10/20 sample introduction pump systems, SPS 4 autosampler, USB-GPIB HS/HS+ communications and Tube-CAM are separate accessories; only Agilent-supplied or approved spare parts may be used.