Agilent Technologies 240Z GFAAS
Agilent Technologies 240Z GFAAS — Atomic spectroscopy.
The 240Z GFAAS is a transverse-Zeeman graphite furnace atomic absorption spectrometer with a pyrolytic platform Omega tube and GTA 120 atomizer, configured for Pb determination in complex matrices. Its 10-step furnace program spans 95–2500 °C, with ashing to 805 °C and atomization at 2132 °C. It serves trace-metal quantification in food and water matrices, achieving Pb MDL of 2.6 ppb.
Agilent Technologies Graphite Furnace Atomic Absorption Instruments Series — Model Comparison & Selection
| Model | Wavelength range | Sample volume, milliliter | Software | Elements | Lamp current, mA | Inert gas purity, % |
|---|---|---|---|---|---|---|
| 240Z GFAAS Current | 283.3 | 0.02 | 5.5 | Pb | 10 | 99.99 |
| 280Z GFAAS | 228.8 | 0.01 | method-driven, guides us | Cd | 4 | 99.99 |
Full specifications
| Wavelength range | 283.3 nm |
| Acetylene Purity | 99.5 % |
| Elements | Pb |
| Inert Gas Purity | 99.99 % |
| Sample volume | 20 µL |
| Software | 5.5 |
| Instrument mfr | 280 AA |
| Background correction mfr | transverse Zeeman Zeeman |
| Autosampler mfr | 120 |
| Graphite tube mfr | pyrolytic platform Omega tube tube |
| Furnace viewing camera mfr | Tube-CAM CAM |
| MDL in solution mfr | 0.35 μg/kg |
| MDL in sample mfr | 18 μg/kg |
| LOQ in sample mfr | 59 μg/kg |
| Maximum allowable limit (Pb in condiments) mfr | 1000 μg/kg |
| Method Detection Limit (MDL), Pb mfr | 2.6 ppb |
| Method Detection Limit (MDL), Cd mfr | 0.37 ppb |
| Method Detection Limit (MDL), Cr mfr | 1.5 ppb |
| Method Detection Limit (MDL), Ni mfr | 6.0 ppb |
| Certified value, Pb (NIST 1567b Wheat Flour) mfr | 10.4 ppb |
| Measured value, Pb mfr | 10.9 ppb |
| Recovery, Pb mfr | 105 |
| Certified value, Cd (NIST 1567b Wheat Flour) mfr | 25.4 ppb |
| Measured value, Cd mfr | 23.4 ppb |
| Recovery, Cd mfr | 92 |
| Certified value, Cr (DUWF-1 Durum Wheat Flour) mfr | 23.0 ppb |
| Measured value, Cr mfr | 21.5 ppb |
| Recovery, Cr mfr | 93 |
| Certified value, Ni (DUWF-1 Durum Wheat Flour) mfr | 170 ppb |
| Measured value, Ni mfr | 172 ppb |
| Recovery, Ni mfr | 101 |
| Matrix mfr | Water Water |
| Modifier mfr | 4 |
| Instrumentation mfr | 240 AAS |
| Exhaust ventilation rate mfr | 6 m3/min |
| Exhaust fan distance from flame mfr | 3 m |
| Acetylene supply pressure mfr | 65...100 kPa |
| Acetylene cylinder replacement pressure mfr | 700 kPa |
| Acetylene maximum withdrawal rate mfr | 1 hour |
| Air supply pressure mfr | 245...455 kPa |
| Air flow rate mfr | 13.5...20 L/min |
| Nitrous oxide supply pressure mfr | 245...455 kPa |
| Nitrous oxide flow rate mfr | 11...16 L/min |
| Air purge flow rate mfr | 10 L/min |
| Acetylene hose mfr | 1.8 thread |
| Air hose mfr | 1.8 fitting |
| Nitrous oxide hose mfr | 1.8 fitting |
| Lamp capacity 240 series mfr | 4 lamps |
| Lamp capacity 280 series mfr | 8 lamps |
| Zeeman magnetic field mfr | 0.8 peak |
| Minimum distance from magnet plate mfr | 300 mm |
| Solvent specific gravity minimum mfr | 0.75 |
| Default IEEE address 55B mfr | 1 |
| Default IEEE address 240FS/240Z/280FS/280Z mfr | 9 |
| Reconfigured IEEE address for instrument #2 mfr | 8 |
| Acetylene maximum pressure mfr | 105 kPa |
| Zeeman magnetic field peak mfr | 0.8 T |
| Lamps accommodated (240 series) mfr | 4 lamps |
| Lamps accommodated (280 series) mfr | 8 lamps |
| Acetylene supply pressure (allowed range) mfr | 65...100 kPa |
| Air supply pressure (allowed range) mfr | 245...455 kPa |
| N2O supply pressure (allowed range) mfr | 245...455 kPa |
| Air purge supply pressure (allowed range) mfr | 245...455 kPa |
| Air typical flow rate mfr | 13.5...20 L/min |
| N2O typical flow rate mfr | 11...16 L/min |
| Air purge typical flow rate mfr | 10 L/min |
| Gas hose internal diameter mfr | 6.3 mm |
| Recommended acetylene delivery pressure mfr | 75 kPa |
| Recommended air delivery pressure mfr | 350 kPa |
| Recommended N2O delivery pressure mfr | 350 kPa |
| Agilent 240/280 系列原子吸收(包括塞曼) mfr | 5 |
| 安捷伦原子吸收光谱仪 Duo 系统的连接与配置 mfr | 39 |
| Duo 运行设置和 IEEE 电缆的连接 mfr | 39 |
| 为安捷伦原子吸收 Duo 操作配置 IEEE 地址 mfr | 40 |
| 配置 SPS 4 自动进样器 mfr | 41 |
| 确定 SPS 4 COM 端口 mfr | 42 |
| 配置软件以使用手动进样 mfr | 42 |
| 启动 SpectrAA 软件 mfr | 47 |
| SpectrAA 发行说明和软件状态公告 mfr | 44 |
| 安装 Tube-CAM 石墨炉摄像头选件的驱动程序 mfr | 45 |
| 打印机设置 mfr | 45 |
| 故障排除 mfr | 45 |
| 服务安装模块 mfr | 45 |
| 软件概述 mfr | 47 |
| SpectrAA 窗口 mfr | 47 |
| 索引页面 mfr | 47 |
| 工作表窗口 mfr | 48 |
| 报告窗口 mfr | 49 |
| 管理窗口 mfr | 50 |
| 使用 SpectrAA 软件 mfr | 50 |
| 菜单栏 mfr | 50 |
| 工具栏 mfr | 50 |
| 对话框 mfr | 50 |
| 帮助 mfr | 51 |
| 搜索帮助 mfr | 51 |
| 一般信息 mfr | 7 |
| 安捷伦原子吸收文档 mfr | 8 |
| 验证安全状态 mfr | 8 |
| 电气危险 mfr | 9 |
| 热量、蒸气和烟雾 mfr | 10 |
| 压缩气瓶的储存和处理 mfr | 11 |
| 气体软管和接头 mfr | 12 |
| 紫外线辐射 mfr | 12 |
| 警告符号 mfr | 13 |
| 彩色标记 mfr | 14 |
| 火焰操作 mfr | 15 |
| 回火 mfr | 22 |
| 石墨炉和塞曼操作 mfr | 26 |
| Minimum acetylene cylinder pressure mfr | 700 kPa |
| SRM concentration mfr | 25.0 μg/L |
| Characteristic concentration mfr | 0.85 μg/L |
| Instrumental Detection limit mfr | 0.15 μg/L |
| Spike concentration mfr | 25 μg/L |
| Slit Width | 0.5 nm |
| Instrument weight mfr | 50 kg |
| Burner slit width limit N2O/acetylene Mark 7 mfr | 0.46 mm |
| Burner slit width limit air/acetylene Mark 7 mfr | 0.54 mm |
| Burner slit width replacement threshold N2O/acetylene mfr | 0.47 mm |
| Burner slit width replacement threshold air/acetylene mfr | 0.54 mm |
| N2O/air-acetylene burner slit width limit (N2O/acetylene) mfr | 0.47 mm |
| Burner slit width limit (air/acetylene) mfr | 0.54 mm |
| Mark 7 burner slit width (N2O/acetylene) mfr | 0.46 mm |
| Mark 7 burner slit width (air/acetylene) mfr | 0.54 mm |
| Gas hose length mfr | 1.8 m |
| N2O/acetylene burner slit width mfr | 0.46 mm |
| Air/acetylene burner slit width mfr | 0.54 mm |
| Characteristic Mass mfr | 13.9 pg |
| Lamp Current | 10 mA |
| Fuse rating 240 series mfr | 4 mm |
| Fuse rating 280FS mfr | 4 mm |
| Circuit breaker 240Z/280Z mfr | 15 breaker |
| Fuse rating 240Z/280Z mfr | 3.15 mm |
| Fuse (240 series, non-Zeeman) mfr | 4 mm |
| Fuse (280FS, non-Zeeman) mfr | 4 mm |
| Circuit breaker (240Z/280Z Zeeman) mfr | 15 A |
| Fuse (240Z/280Z Zeeman) mfr | 800 mm |
| Acetylene hose fitting mfr | 9 L/H |
| Air hose fitting mfr | 9 UNF |
| N2O hose fitting mfr | 3 UNF |
| USB 转 RS-232 接口 mfr | 43 |
| Minimum distance from magnet plate (pacemakers/magnetic media) mfr | 300 mm |
| Ash Temperature | 598 °C |
| Atomize Temperature | 1435 °C |
| Standard deviation, Pb mfr | 1.4 ppb |
| Standard deviation, Cd mfr | 0.22 ppb |
| Standard deviation, Cr mfr | 2.1 ppb |
| Standard deviation, Ni mfr | 7.5 ppb |
| Standard solution concentration mfr | 50 μg/L |
Description
Construction and series positioning
The 240Z is the Zeeman graphite furnace variant of the 240 series, distinguished from the 280Z by lamp capacity: the 240 series accommodates 4 hollow cathode lamps versus 8 on the 280 series, so the 240Z suits methods targeting a limited element set (here Pb, with Cd/Cr/Ni also demonstrated). Background correction is transverse Zeeman, and the magnet produces up to 0.8 Tesla peak at power-line frequency during the read phase. The atomizer is a GTA 120 graphite tube atomizer fitted with a pyrolytic platform Omega tube; a PSD 120 programmable sample dispenser handles sample introduction, and a Tube-CAM furnace viewing camera is specified.
Furnace behavior and method performance
- The 10-step temperature program runs two drying steps (95 °C, 120 °C), five ashing steps (400, 500, 805, 805, 805 °C), two atomization steps (2132 °C each) and a 2500 °C cleanout.
- For the SRM method, ash temperature is 1000 °C and atomization 2000 °C, showing the program is matrix- and method-specific rather than fixed.
- Pb MDL is 2.6 ppb; Cd 0.37 ppb, Cr 1.5 ppb, Ni 6.0 ppb. In the soy-sauce application, MDL is 0.35 μg/kg in solution and 18 μg/kg in sample, LOQ 59 μg/kg, against a 1000 μg/kg maximum limit for Pb in condiments.
- NIST 1567b Wheat Flour Pb: certified 10.4 ppb, measured 10.9 ppb, SD 1.4 ppb, recovery 105 %.
Installation, gas and control specifics
The instrument requires a grounded 3-wire outlet with correct voltage setting; the 240Z/280Z use a 15 A long-delay circuit breaker plus T3.15 A H 250 V, F315 mA L 250 V, F1A L 250 V, T6.3 A L 250 V and F800 mA L 250 V fuses for 208–240 VAC. Graphite furnace operation requires inert gas (argon or nitrogen) and air; 95 % argon (or nitrogen) + 5 % hydrogen pre-mixed is allowed, but pure hydrogen must never be used. Acetylene, where the flame configuration is used, must be instrument grade ≥99.5 % at 65–100 kPa (recommended 75 kPa, never above 105 kPa). Gas hoses are 1.8 m with 6.3 mm ID; acetylene uses a red 9/16-18UNF L/H fitting, air black 9/16-18UNF, N2O blue 3/4-16UNF. Exhaust must provide at least 6 m³/min (200 cfm) with the fan at least 3 m from the flame. SpectrAA software (Base, PRO, CFR) is installed via USB-GPIB converter; the SPS 4 autosampler is configured over USB with IHelper.EXE.
Integration & Operating Notes
The 240Z/280Z configuration requires a 15 A long-delay circuit breaker plus T3.15 A H 250 V, F315 mA L 250 V, F1A L 250 V, T6.3 A L 250 V and F800 mA L 250 V fuses. Substituting any fuse with a different type or rating compromises fire protection and voids the safety basis of the installation.
The furnace accepts inert gas (argon or nitrogen) plus air, and a pre-mixed 95 % argon (or nitrogen) + 5 % hydrogen blend is permitted. Pure hydrogen must not be used under any circumstance because of explosion hazard; verify the cylinder label and regulator before connecting the furnace gas line.
The transverse Zeeman magnet generates up to 0.8 Tesla peak at power-line frequency during the read phase. Pacemakers and magnetic storage media must remain at least 300 mm from the magnet plate; plan bench layout and operator access so this distance is maintained during analysis.
Laboratory ventilation must be sized to at least 6 m³/min (200 cfm), and the exhaust fan must be located at least 3 m (10 ft) from the flame. This is a pre-installation facility requirement; undersized extraction affects both operator safety and baseline stability.
Following a flashback, shut off gas at the regulator, turn off the instrument, and do not reuse the burner head; contact Agilent service and inspect sample introduction components. Reusing the head risks a repeat event because the slit geometry and internal passages may be distorted.
Selection & Integration Q&A
Choose the 240Z when the method targets a limited element set: the 240 series holds 4 hollow cathode lamps versus 8 on the 280 series. Both use transverse Zeeman correction and the GTA 120; the 280Z pays off only when many elements must be run without lamp changes.
Inert gas (argon or nitrogen) plus air; a pre-mixed 95 % argon (or nitrogen) + 5 % hydrogen blend is allowed, but pure hydrogen is prohibited. Acetylene, if the flame configuration is used, must be ≥99.5 % instrument grade at 65–100 kPa, never above 105 kPa.
Exhaust must provide at least 6 m³/min (200 cfm) with the fan at least 3 m from the flame; the instrument exceeds 50 kg and needs two or more people to lift. Keep magnetic media and pacemakers 300 mm from the Zeeman magnet, which reaches 0.8 Tesla peak.
Weekly O-ring inspection (burner head, spray chamber, plug, inlet) and window cleaning are specified, plus daily gas-supply checks. The PSD 120 dispenser, SPS 4 autosampler, Tube-CAM and UltrAA lamps are separate accessories; the SPS 4 needs USB setup via IHelper.EXE and the PC needs a USB-GPIB converter.