Agilent Technologies 4150 TapeStation
Agilent Technologies 4150 TapeStation — Automated electrophoresis.
The 4150 TapeStation is an automated electrophoresis instrument using ScreenTape consumables for DNA and RNA sample quality control. It processes up to 16 samples per consumable with a 1 µL sample volume, sizing 35–1,000 bp and quantifying 25–500 ng/µL. Positioned below the 4200 model, it lacks 4200-specific liquid-handling tests and is suited for standard NGS and biobank workflows.
Agilent Technologies Tapestation Instruments Series — Model Comparison & Selection
| Model | Sensitivity | Dimensions w d h | Weight, kg | Storage temperature, °C | Analysis time, min | Application |
|---|---|---|---|---|---|---|
| 4150 TapeStation Current | 5 | 254 | 17.8 | -40...70 | 16 | — |
| 4150 TapeStationDx | — | — | — | — | — | automated electrophoresi |
| 4200 TapeStation | 5 | 435 | 25.5 | -40...70 | 16 | — |
| 4200 TapeStationDx System | — | — | — | — | — | automated electrophoresi |
Full specifications
| Sensitivity | 5 ng/µL |
| Analysis time | 16 min |
| Dimensions (w × D × H) | 254 mm |
| Kit Stability | 6 months |
| Quantitative precision mfr | 10 % CV |
| Quantitative accuracy mfr | 20 |
| Quantitative range mfr | 25...500 ng/μL |
| RINe functional range mfr | 25...500 ng/μL |
| Maximum buffer concentration in sample mfr | 200 NaCl |
| Sample volume required mfr | 1 μL |
| Kit size mfr | 112 samples |
| Sizing range mfr | 35...1,000 bp |
| Safety standards mfr | 2 only. |
| ISM Classification mfr | 1 |
| Computer Specifications mfr | Checks the computer specifications and configura |
| Insufficient Mixing mfr | Buffer and sample was not sufficiently vortexed. |
| Sample concentration outside recommended range for application mfr | Linear concentration estimation cannot be achiev |
| Agilent D1000 ScreenTape Assay mfr | 2991 |
| Agilent High Sensitivity D1000 ScreenTape Assay mfr | 2991 |
| Agilent RNA ScreenTape Assay mfr | 2991 |
| Agilent High Sensitivity RNA ScreenTape Assay mfr | 2991 |
| Agilent Genomic DNA ScreenTape Assay mfr | 2991 |
| Agilent D5000 ScreenTape Assay mfr | 2991 |
| Agilent High Sensitivity D5000 ScreenTape Assay mfr | 2991 |
| Agilent Cell-free DNA ScreenTape Assay mfr | 2991 |
| Vision mfr | Camera and LED illumination are checked using th |
| Electrophoresis mfr | The voltage supply, current readout, and electro |
| Robotics mfr | Robotics parts are checked by dedicated movement |
| Pipetting mfr | This check procedure in the Hardware Troubleshoo |
| Cleanup mfr | Resets the instrument to normal operational mode |
| Needle To Pump mfr | Yes Yes |
| Tape Handling mfr | Yes Yes |
| Sample concentration outside recommended range for application. mfr | Linear concentration estimation cannot be achiev |
| Sample prepared too long before analysis mfr | Evaporation or settling may alter the sample to |
| Incorrect Marker peaks picked up mfr | Depending on the sample pattern, wrong peaks can |
| Clear space around instrument mfr | 75 mm |
| Additional top clearance for lid mfr | 25 cm |
| Unpacking equilibration wait mfr | 5 h |
| Second ScreenTape prompt time frame mfr | 15 minutes |
| Needle exchange threshold mfr | 3840 pierces |
| Maximum pierces before inhibition mfr | 4480 pierces |
| Weight | 17.8 kg |
| Power consumption | 50 VA |
| Line voltage mfr | 100...240 |
| Line frequency mfr | 50 |
| Interface mfr | USB USB |
| Incorrect protocol used mfr | Differences exist between Standard and High Sens |
| Samples per consumable mfr | 16 |
| Operating Altitude | 3000 m |
| Sound pressure | 70 decibel |
| Storage temperature | -40...70 °C |
| Operating humidity mfr | 15...80 condensing |
| Software Environment mfr | Checks the compatibility of TapeStation Controll |
| Computer Environment mfr | Evaluation of computer specification and install |
Description
Construction and Series Positioning
The 4150 TapeStation is a benchtop automated electrophoresis system that uses ScreenTape devices and a robotic pipetting mechanism to analyze DNA and RNA samples. Within the Tapestation Instruments series, it is the base model; the 4200 TapeStation adds liquid-handling tests (Tape Handling, Teach Pump-ML, Teach Pump-ML-Z) that are not applicable to the 4150. The 4150 TapeStationDx is a diagnostic variant, while the 4200 TapeStationDx System is a higher-throughput diagnostic configuration. The 4150 is intended for research use where 16 samples per run and standard sensitivity (5 ng/µL) are sufficient.
Assay Behavior and Performance
- Sizing range: 35–1,000 bp; quantitative range: 25–500 ng/µL; RINe functional range: 25–500 ng/µL.
- Quantitative precision: 10 % CV; accuracy: 20 %.
- Analysis time: 16 min per consumable (16 samples).
- Maximum buffer concentration in sample: 200 mM NaCl; higher salt can interfere with electrophoresis.
- Kit size: 112 samples (7 consumables of 16). Kit stability: 6 months.
- RNA assays require heat denaturation at 72 °C for 3 min followed by 2 min on ice; a calibrated block heater or PCR instrument for 200 µL vials is required but not supplied.
Installation, Control and Operation
The instrument is controlled by TapeStation Controller software (revision 5.1 allows up to two instruments per computer). System Diagnostics includes System Verification (~20 min) and Hardware Troubleshooting (~30 min), which use the TapeStation Test Tape 5067-5601, Loading Tips 5067-5598, and Tip Waste Bucket G2992-40046. The software supports PDF/Word reporting and a Security Module edition. Maintenance involves piercing needle exchange (after 4,480 samples; warning at 3,840) and electrode cartridge replacement (after 50,000 samples or one year, whichever comes first).
Integration & Operating Notes
A warning appears after 3,840 samples, and no further analysis is possible beyond 4,480 samples until a needle exchange is done. Plan the exchange during a maintenance window to avoid unexpected downtime; after inserting a new needle, wait approximately two minutes for automated calibration to complete without powering off or terminating software.
The electrode cartridge must be replaced after either 50,000 samples or one year of usage, whichever comes first. A cartridge older than one year triggers a constant 'Electrode Cartridge Change Required' message. The change is successful only if the cartridge is registered, the date of first use is the current day, no samples were analyzed with it, and the electrical functionality test with the TapeStation Test Tape passes; otherwise the instrument remains inoperable on the affected lanes.
RNA and High Sensitivity RNA ScreenTape assays require heat denaturation at 72 °C for 3 min. Many heat blocks and PCR machines display a temperature that can be incorrect by up to 10 °C, so calibrate the block before use. Use an accurately calibrated block heater or PCR instrument fitting 200 µL vials; this is not supplied with the system.
For 96-well sample plates, use the recommended foil seal PN 5067-5154 to prevent evaporation and ensure tight sealing throughout the plate. Inadequate sealing can lead to sample loss and incorrect quantification.
Selection & Integration Q&A
Choose the 4150 for standard DNA/RNA QC with up to 16 samples per run and no need for liquid-handling tests. The 4200 adds Tape Handling and Teach Pump-ML tests (for G2991A/B) and may offer higher throughput. Dx variants are for diagnostic workflows with regulatory requirements. The 4150 is the base research model.
System Verification requires TapeStation Test Tape 5067-5601, a full rack of Loading Tips 5067-5598, and TapeStation Tip Waste Bucket G2992-40046. These are not included in the base delivery and must be ordered separately. The test takes about 20 minutes and requires user interaction for material insertion and status LED confirmation.
Yes, with TapeStation software revision 5.1, up to two TapeStation instruments can be connected to one computer. However, hardware diagnostics can be executed for only one instrument at a time. Plan diagnostic runs sequentially to avoid conflicts.