Agilent Technologies 4150 TapeStation
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 4150 TapeStation

Agilent Technologies 4150 TapeStation — Automated electrophoresis.

The 4150 TapeStation is an automated electrophoresis instrument using ScreenTape consumables for DNA and RNA sample quality control. It processes up to 16 samples per consumable with a 1 µL sample volume, sizing 35–1,000 bp and quantifying 25–500 ng/µL. Positioned below the 4200 model, it lacks 4200-specific liquid-handling tests and is suited for standard NGS and biobank workflows.

Agilent Technologies Tapestation Instruments Series — Model Comparison & Selection

Model SensitivityDimensions w d hWeight, kgStorage temperature, °CAnalysis time, minApplication
4150 TapeStation Current 525417.8-40...7016—
4150 TapeStationDx —————automated electrophoresi
4200 TapeStation 543525.5-40...7016—
4200 TapeStationDx System —————automated electrophoresi

Full specifications

Core Performance & Functional Specifications
Sensitivity 5 ng/µL
Analysis time 16 min
Dimensions (w × D × H) 254 mm
Kit Stability 6 months
Quantitative precision mfr 10 % CV
Quantitative accuracy mfr 20
Quantitative range mfr 25...500 ng/μL
RINe functional range mfr 25...500 ng/μL
Maximum buffer concentration in sample mfr 200 NaCl
Sample volume required mfr 1 μL
Kit size mfr 112 samples
Sizing range mfr 35...1,000 bp
Safety standards mfr 2 only.
ISM Classification mfr 1
Computer Specifications mfr Checks the computer specifications and configura
Insufficient Mixing mfr Buffer and sample was not sufficiently vortexed.
Sample concentration outside recommended range for application mfr Linear concentration estimation cannot be achiev
Agilent D1000 ScreenTape Assay mfr 2991
Agilent High Sensitivity D1000 ScreenTape Assay mfr 2991
Agilent RNA ScreenTape Assay mfr 2991
Agilent High Sensitivity RNA ScreenTape Assay mfr 2991
Agilent Genomic DNA ScreenTape Assay mfr 2991
Agilent D5000 ScreenTape Assay mfr 2991
Agilent High Sensitivity D5000 ScreenTape Assay mfr 2991
Agilent Cell-free DNA ScreenTape Assay mfr 2991
Vision mfr Camera and LED illumination are checked using th
Electrophoresis mfr The voltage supply, current readout, and electro
Robotics mfr Robotics parts are checked by dedicated movement
Pipetting mfr This check procedure in the Hardware Troubleshoo
Cleanup mfr Resets the instrument to normal operational mode
Needle To Pump mfr Yes Yes
Tape Handling mfr Yes Yes
Sample concentration outside recommended range for application. mfr Linear concentration estimation cannot be achiev
Sample prepared too long before analysis mfr Evaporation or settling may alter the sample to
Incorrect Marker peaks picked up mfr Depending on the sample pattern, wrong peaks can
Clear space around instrument mfr 75 mm
Additional top clearance for lid mfr 25 cm
Unpacking equilibration wait mfr 5 h
Second ScreenTape prompt time frame mfr 15 minutes
Needle exchange threshold mfr 3840 pierces
Maximum pierces before inhibition mfr 4480 pierces
Dimensions, Weight & Mechanical Design
Weight 17.8 kg
Electrical, Power & Drive Specifications
Power consumption 50 VA
Line voltage mfr 100...240
Line frequency mfr 50
Interfaces, Connections & Control
Interface mfr USB USB
Incorrect protocol used mfr Differences exist between Standard and High Sens
Materials, Media & Process Compatibility
Samples per consumable mfr 16
Operating Conditions, Environment & Protection
Operating Altitude 3000 m
Sound pressure 70 decibel
Storage temperature -40...70 °C
Operating humidity mfr 15...80 condensing
Software Environment mfr Checks the compatibility of TapeStation Controll
Computer Environment mfr Evaluation of computer specification and install

Description

Construction and Series Positioning

The 4150 TapeStation is a benchtop automated electrophoresis system that uses ScreenTape devices and a robotic pipetting mechanism to analyze DNA and RNA samples. Within the Tapestation Instruments series, it is the base model; the 4200 TapeStation adds liquid-handling tests (Tape Handling, Teach Pump-ML, Teach Pump-ML-Z) that are not applicable to the 4150. The 4150 TapeStationDx is a diagnostic variant, while the 4200 TapeStationDx System is a higher-throughput diagnostic configuration. The 4150 is intended for research use where 16 samples per run and standard sensitivity (5 ng/µL) are sufficient.

Assay Behavior and Performance

Installation, Control and Operation

The instrument is controlled by TapeStation Controller software (revision 5.1 allows up to two instruments per computer). System Diagnostics includes System Verification (~20 min) and Hardware Troubleshooting (~30 min), which use the TapeStation Test Tape 5067-5601, Loading Tips 5067-5598, and Tip Waste Bucket G2992-40046. The software supports PDF/Word reporting and a Security Module edition. Maintenance involves piercing needle exchange (after 4,480 samples; warning at 3,840) and electrode cartridge replacement (after 50,000 samples or one year, whichever comes first).

Integration & Operating Notes

Maintenance
Schedule piercing needle exchange before 4,480 samples

A warning appears after 3,840 samples, and no further analysis is possible beyond 4,480 samples until a needle exchange is done. Plan the exchange during a maintenance window to avoid unexpected downtime; after inserting a new needle, wait approximately two minutes for automated calibration to complete without powering off or terminating software.

User Manual, Troubleshooting / page 179; Maintenance / page 133
Maintenance
Replace electrode cartridge at 50,000 samples or one year

The electrode cartridge must be replaced after either 50,000 samples or one year of usage, whichever comes first. A cartridge older than one year triggers a constant 'Electrode Cartridge Change Required' message. The change is successful only if the cartridge is registered, the date of first use is the current day, no samples were analyzed with it, and the electrical functionality test with the TapeStation Test Tape passes; otherwise the instrument remains inoperable on the affected lanes.

User Manual, Troubleshooting / page 180; Maintenance / page 138
Process Limits
Calibrate heat block for RNA assays to avoid 10 °C error

RNA and High Sensitivity RNA ScreenTape assays require heat denaturation at 72 °C for 3 min. Many heat blocks and PCR machines display a temperature that can be incorrect by up to 10 °C, so calibrate the block before use. Use an accurately calibrated block heater or PCR instrument fitting 200 µL vials; this is not supplied with the system.

User Manual, Good Measurement Practices / page 97
Maintenance
Use correct foil seal PN 5067-5154 for 96-well plates

For 96-well sample plates, use the recommended foil seal PN 5067-5154 to prevent evaporation and ensure tight sealing throughout the plate. Inadequate sealing can lead to sample loss and incorrect quantification.

User Manual, Troubleshooting DNA Applications / page 189

Selection & Integration Q&A

When should I choose the 4150 TapeStation instead of a 4200 TapeStation or a Dx variant?

Choose the 4150 for standard DNA/RNA QC with up to 16 samples per run and no need for liquid-handling tests. The 4200 adds Tape Handling and Teach Pump-ML tests (for G2991A/B) and may offer higher throughput. Dx variants are for diagnostic workflows with regulatory requirements. The 4150 is the base research model.

What consumables and accessories are required for System Verification, and are they included with the instrument?

System Verification requires TapeStation Test Tape 5067-5601, a full rack of Loading Tips 5067-5598, and TapeStation Tip Waste Bucket G2992-40046. These are not included in the base delivery and must be ordered separately. The test takes about 20 minutes and requires user interaction for material insertion and status LED confirmation.

Can I connect two 4150 TapeStations to one computer, and what are the limitations?

Yes, with TapeStation software revision 5.1, up to two TapeStation instruments can be connected to one computer. However, hardware diagnostics can be executed for only one instrument at a time. Plan diagnostic runs sequentially to avoid conflicts.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent 4150 TapeStation System Manual
Manual ·232 pp.
View PDF